OASIS 3650-100Recalls & reported problems
Atrium Medical Corporation · OASIS DRAIN, BRU, IN-LINE, ATS COMPATIBLE · Class 2
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 22
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 2
Summary
The FDA's MAUDE database contains 22 adverse event reports that list this exact device (DI 00650862113013), received between Jan 31, 2017 and Dec 14, 2022 — classified as 2 injuries, 20 malfunctions.
The most frequently reported specific device problem is “Device Misassembled During Manufacturing /Shipping” (15 reports).
2 FDA recall records name this device's identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Sep 18, 2023Open, Classified
Instructions for Use (IFU) for the Atrium Ocean, Oasis, and Express chest drains do not provide sufficient precaution instruction for proper set up of catheter(s) and patient tube connections with single collection chamber chest drains
Z-0317-2024 · Root cause: Error in labeling · FDA record ↗
- Sep 5, 2023Open, Classified
Product was re-processed and re-sterilized by a third party using packaging configurations and a sterilization process which are not approved by the manufacture and lack validation.
Z-0486-2024 · Root cause: Labeling mix-ups · FDA record ↗
Most reported problems
Device problems
- Device Misassembled During Manufacturing /Shipping15
- Nonstandard Device15
- Connection Problem2
- Loose or Intermittent Connection2
- Vibration2
- Device Operates Differently Than Expected1
- Device Slipped1
- Fluid/Blood Leak1
Patient problems
- No Clinical Signs, Symptoms or Conditions16
- No Known Impact Or Consequence To Patient4
- No Consequences Or Impact To Patient1
- No Information1
Recent adverse event reports
The 22 most recent of 22 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 14, 2022Nonstandard Device; Device Misassembled During Manufacturing /Shipping; Nonstandard Device; Device Misassembled During Manufacturing /Shipping
THE CUSTOMER REPORTED RECEIVING THE INCORRECT PRODUCT FOR ITEM #5303801 (3650-100) DRAIN OASIS DRY SUCTION BLOOD. THE PRODUCT INSIDE IS ITEM #512404 (3600-100) DRAIN, SINGLE COLLECTION. THE CUSTOMER ALSO REPORTED THAT THE FULL LOT QUANTITY AFFECTED WAS A TOTAL OF 15 UNITS. REFERENCE RELATED MDR: (B)(4).
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15976571 ↗MalfunctionDec 14, 2022Nonstandard Device; Device Misassembled During Manufacturing /Shipping; Nonstandard Device; Device Misassembled During Manufacturing /Shipping
THE CUSTOMER REPORTED RECEIVING THE INCORRECT PRODUCT FOR ITEM #5303801 (3650-100) DRAIN OASIS DRY SUCTION BLOOD. THE PRODUCT INSIDE IS ITEM #512404 (3600-100) DRAIN, SINGLE COLLECTION. THE CUSTOMER ALSO REPORTED THAT THE FULL LOT QUANTITY AFFECTED WAS A TOTAL OF 15 UNITS. REFERENCE RELATED MDR: 3011175548-2022-00411.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15976566 ↗MalfunctionDec 14, 2022Nonstandard Device; Device Misassembled During Manufacturing /Shipping; Nonstandard Device; Device Misassembled During Manufacturing /Shipping
THE CUSTOMER REPORTED RECEIVING THE INCORRECT PRODUCT FOR ITEM #5303801 (3650-100) DRAIN OASIS DRY SUCTION BLOOD. THE PRODUCT INSIDE IS ITEM #512404 (3600-100) DRAIN, SINGLE COLLECTION. THE CUSTOMER ALSO REPORTED THAT THE FULL LOT QUANTITY AFFECTED WAS A TOTAL OF 15 UNITS. REFERENCE RELATED MDR: 3011175548-2022-00411.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15976553 ↗MalfunctionDec 14, 2022Nonstandard Device; Device Misassembled During Manufacturing /Shipping; Nonstandard Device; Device Misassembled During Manufacturing /Shipping
THE CUSTOMER REPORTED RECEIVING THE INCORRECT PRODUCT FOR ITEM #5303801 (3650-100) DRAIN OASIS DRY SUCTION BLOOD. THE PRODUCT INSIDE IS ITEM #512404 (3600-100) DRAIN, SINGLE COLLECTION. THE CUSTOMER ALSO REPORTED THAT THE FULL LOT QUANTITY AFFECTED WAS A TOTAL OF (B)(4) UNITS. REFERENCE RELATED MDR: 3011175548-2022-00411.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15976552 ↗MalfunctionDec 14, 2022Nonstandard Device; Device Misassembled During Manufacturing /Shipping; Nonstandard Device; Device Misassembled During Manufacturing /Shipping
THE CUSTOMER REPORTED RECEIVING THE INCORRECT PRODUCT FOR ITEM#: 5303801 (3650-100) DRAIN OASIS DRY SUCTION BLOOD. THE PRODUCT INSIDE IS ITEM#: 512404 (3600-100) DRAIN, SINGLE COLLECTION. THE CUSTOMER ALSO REPORTED THAT THE FULL LOT QUANTITY AFFECTED WAS A TOTAL OF 15 UNITS. REFERENCE RELATED MDR: 3011175548-2022-00411.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15976547 ↗MalfunctionDec 14, 2022Nonstandard Device; Device Misassembled During Manufacturing /Shipping; Nonstandard Device; Device Misassembled During Manufacturing /Shipping
MalfunctionDec 14, 2022Nonstandard Device; Device Misassembled During Manufacturing /Shipping; Nonstandard Device; Device Misassembled During Manufacturing /Shipping
THE CUSTOMER REPORTED RECEIVING THE INCORRECT PRODUCT FOR ITEM#: 5303801 (3650-100) DRAIN OASIS DRY SUCTION BLOOD. THE PRODUCT INSIDE IS ITEM#: 512404 (3600-100) DRAIN, SINGLE COLLECTION. THE CUSTOMER ALSO REPORTED THAT THE FULL LOT QUANTITY AFFECTED WAS A TOTAL OF (B)(4) UNITS. REFERENCE RELATED MDR: 3011175548-2022-00411.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15976540 ↗MalfunctionDec 14, 2022Nonstandard Device; Device Misassembled During Manufacturing /Shipping; Nonstandard Device; Device Misassembled During Manufacturing /Shipping
THE CUSTOMER REPORTED RECEIVING THE INCORRECT PRODUCT FOR ITEM #5303801 (3650-100) DRAIN OASIS DRY SUCTION BLOOD. THE PRODUCT INSIDE IS ITEM #512404 (3600-100) DRAIN, SINGLE COLLECTION. THE CUSTOMER ALSO REPORTED THAT THE FULL LOT QUANTITY AFFECTED WAS A TOTAL OF 15 UNITS. REFERENCE RELATED MDR: 3011175548-2022-00411.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15976536 ↗MalfunctionDec 14, 2022Nonstandard Device; Device Misassembled During Manufacturing /Shipping; Nonstandard Device; Device Misassembled During Manufacturing /Shipping
THE CUSTOMER REPORTED RECEIVING THE INCORRECT PRODUCT FOR ITEM #5303801 (3650-100) DRAIN OASIS DRY SUCTION BLOOD. THE PRODUCT INSIDE IS ITEM #512404 (3600-100) DRAIN, SINGLE COLLECTION. THE CUSTOMER ALSO REPORTED THAT THE FULL LOT QUANTITY AFFECTED WAS A TOTAL OF (B)(4) UNITS. REFERENCE RELATED MDR: 3011175548-2022-00411.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15976531 ↗MalfunctionDec 14, 2022Nonstandard Device; Device Misassembled During Manufacturing /Shipping; Nonstandard Device; Device Misassembled During Manufacturing /Shipping
THE CUSTOMER REPORTED RECEIVING THE INCORRECT PRODUCT FOR ITEM #5303801 (3650-100) DRAIN OASIS DRY SUCTION BLOOD. THE PRODUCT INSIDE IS ITEM #512404 (3600-100) DRAIN, SINGLE COLLECTION. THE CUSTOMER ALSO REPORTED THAT THE FULL LOT QUANTITY AFFECTED WAS A TOTAL OF 15 UNITS. REFERENCE RELATED MDR: 3011175548-2022-00411.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15976381 ↗MalfunctionDec 14, 2022Nonstandard Device; Device Misassembled During Manufacturing /Shipping; Nonstandard Device; Device Misassembled During Manufacturing /Shipping
THE CUSTOMER REPORTED RECEIVING THE INCORRECT PRODUCT FOR ITEM #5303801 (3650-100) DRAIN OASIS DRY SUCTION BLOOD. THE PRODUCT INSIDE IS ITEM #512404 (3600-100) DRAIN, SINGLE COLLECTION. THE CUSTOMER ALSO REPORTED THAT THE FULL LOT QUANTITY AFFECTED WAS A TOTAL OF (B)(4) UNITS. REFERENCE RELATED MDR: 3011175548-2022-00411.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15976379 ↗MalfunctionDec 14, 2022Nonstandard Device; Device Misassembled During Manufacturing /Shipping; Nonstandard Device; Device Misassembled During Manufacturing /Shipping
THE CUSTOMER REPORTED RECEIVING THE INCORRECT PRODUCT FOR ITEM # 5303801 (3650-100) DRAIN OASIS DRY SUCTION BLOOD. THE PRODUCT INSIDE IS ITEM #512404 (3600-100) DRAIN, SINGLE COLLECTION. THE CUSTOMER ALSO REPORTED THAT THE FULL LOT QUANTITY AFFECTED WAS A TOTAL OF (B)(4) UNITS. REFERENCE RELATED MDR: 3011175548-2022-00411.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15976378 ↗MalfunctionDec 14, 2022Nonstandard Device; Device Misassembled During Manufacturing /Shipping; Nonstandard Device; Device Misassembled During Manufacturing /Shipping
THE CUSTOMER REPORTED RECEIVING THE INCORRECT PRODUCT FOR ITEM #5303801 (3650-100) DRAIN OASIS DRY SUCTION BLOOD. THE PRODUCT INSIDE IS ITEM #512404 (3600-100) DRAIN, SINGLE COLLECTION. THE CUSTOMER ALSO REPORTED THAT THE FULL LOT QUANTITY AFFECTED WAS A TOTAL OF (B)(4) UNITS. REFERENCE RELATED MDR: 3011175548-2022-00411.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15976376 ↗MalfunctionDec 14, 2022Nonstandard Device; Device Misassembled During Manufacturing /Shipping; Nonstandard Device; Device Misassembled During Manufacturing /Shipping
THE CUSTOMER REPORTED RECEIVING THE INCORRECT PRODUCT FOR ITEM#: 5303801 (3650-100) DRAIN OASIS DRY SUCTION BLOOD. THE PRODUCT INSIDE IS ITEM#: 512404 (3600-100) DRAIN, SINGLE COLLECTION. THE CUSTOMER ALSO REPORTED THAT THE FULL LOT QUANTITY AFFECTED WAS A TOTAL OF 15 UNITS. REFERENCE RELATED MDR: 3011175548-2022-00411.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15976370 ↗MalfunctionDec 13, 2022Nonstandard Device; Device Misassembled During Manufacturing /Shipping; Nonstandard Device; Device Misassembled During Manufacturing /Shipping
THE CUSTOMER REPORTED RECEIVING THE INCORRECT PRODUCT FOR ITEM #5303801 (3650-100) DRAIN OASIS DRY SUCTION BLOOD. THE PRODUCT INSIDE IS ITEM #512404 (3600-100) DRAIN, SINGLE COLLECTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15968497 ↗MalfunctionSep 24, 2021Use of Device Problem; Improper or Incorrect Procedure or Method; Use of Device Problem
THE REPORT RECEIVED STATED THAT THE RED SUCTION BELLOWS OF THE OASIS DRAIN INDICATED THAT SUCTION WAS APPLIED AND FUNCTIONING. BUT THE REGULATOR WAS SET UP INCORRECTLY AND TOO LOW BY MISTAKE (-4= 30MMHG). FACILITY INQUIRED HOW COULD THE SUCTION STILL FUNCTION WITH THE LOW SETTING. THE INSTRUCTIONS FOR USE (IFU) STATES THAT THE SUCTION SOURCE SHOULD BE SET TO -80MMHG OR HIGHER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12530601 ↗MalfunctionJul 3, 2019No Flow
InjuryJul 11, 2017Device Operates Differently Than Expected
DURING PROCEDURE THE OASIS BRU DRAIN WAS BEING USED TO CAPTURE THE PATIENTS BLOOD IN ORDER TO BE RE-INFUSED WITH THEIR OWN BLOOD. A PART OF THE DRAIN MALFUNCTIONED AND THE BLOOD THAT WAS COLLECTED WAS UNABLE TO BE RE-INFUSED TO THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6703667 ↗MalfunctionJun 11, 2017Device Slipped; Vibration; Connection Problem
REPORT RECEIVED STATED THAT THE DRAINS DRY SUCTION REGULATOR SEEMED TO BE LOOSE AND NOT PROPERLY CONNECTED, AND THAT PART OF THE DRAIN WAS VIBRATING.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6632879 ↗MalfunctionJun 11, 2017Loose or Intermittent Connection; Connection Problem
REPORT RECEIVED STATED THAT THE DRAINS DRY SUCTION REGULATOR SEEMED TO BE LOOSE AND NOT PROPERLY CONNECTED, AND THAT PART OF THE DRAIN WAS VIBRATING.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6632877 ↗MalfunctionJun 11, 2017Loose or Intermittent Connection; Vibration
REPORT RECEIVED STATED THAT THE DRAINS DRY SUCTION REGULATOR SEEMED TO BE LOOSE AND NOT PROPERLY CONNECTED, AND THAT PART OF THE DRAIN WAS VIBRATING.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6632875 ↗InjuryJan 31, 2017Fluid/Blood Leak
REPORT RECEIVED STATED THAT DURING OPERATION OF A DRAIN IT WAS NOTICED THAT BLOOD WAS LEAKING ONTO THE FLOOR AND ON FRONT OF DRAIN.
Patient problems: No Information
MAUDE report 6289887 ↗
Other APPARATUS, AUTOTRANSFUSION devices
- EXPRESS 4050-100NAtrium Medical Corporation · APPARATUS, AUTOTRANSFUSION2 recalls2 reports
- NA ATL2001Medtronic, Inc. · APPARATUS, AUTOTRANSFUSION1 death79 reports
- autoLog IQ™ ATLGIQMedtronic, Inc. · APPARATUS, AUTOTRANSFUSION58 reports
- NA BTC93Medtronic, Inc. · APPARATUS, AUTOTRANSFUSION1 recall2 reports
- OCEAN 2050-000Atrium Medical Corporation · Apparatus, Autotransfusion1 recall2 reports
- NA BTC98Medtronic, Inc. · APPARATUS, AUTOTRANSFUSION1 recall0 reports
- HAEMONETICS CELL SAVER ELITE SYSTEM CSE-P-125Haemonetics Corporation · Apparatus, Autotransfusion47 reports
- NA ATLS24Medtronic, Inc. · APPARATUS, AUTOTRANSFUSION18 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.