iCAST 85414Recalls & reported problems
Atrium Medical Corporation · iCAST COVERED STENT, 7MMX38MMX120CM · Class 2 · Implant
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 10
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00650862854145), received between May 31, 2016 and Oct 4, 2022 — classified as 8 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Activation, Positioning or Separation Problem” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K050814); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Activation, Positioning or Separation Problem1
- Break1
- Burst Container or Vessel1
- Device Dislodged or Dislocated1
- Difficult or Delayed Positioning1
- Difficult to Remove1
- Expiration Date Error1
- Inflation Problem1
Patient problems
- No Consequences Or Impact To Patient3
- No Known Impact Or Consequence To Patient3
- Aneurysm1
- No Clinical Signs, Symptoms or Conditions1
- No Information1
- Occlusion1
Recent adverse event reports
The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.
InjuryOct 4, 2022Difficult to Remove; Material Separation; Separation Failure; Difficult to Remove; Separation Failure
DURING A PHYSICIAN-MODIFIED ENDOVASCULAR GRAFT (PMEG) PROCEDURE CASE AND AFTER DEPLOYING THE ICAST STENT IN THE CELIAC ARTERY THE CATHETER SEPARATED COMPLETELY APPROXIMATELY 4CM FROM THE PROXIMAL MARKER ON THE BALLOON LEAVING THE BALLOON PORTION AND PART OF THE CATHETER REMAINING IN THE CELIAC ARTERY. THE DOCTOR THEN USED A SNARE AND WAS ABLE TO RETRIEVE THE REMAINING CATHETER AND BALLOON SYSTEM. THE PATIENT IS DOING FINE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15539958 ↗InjuryFeb 24, 2020Expiration Date Error
REPORT RECEIVED STATED THAT A PHYSICIAN IMPLANTED AN ICAST STENT IN A PATIENT THAT THE STENTS EXPIRATION DATE HAD EXPIRED. THE STENT WAS IMPLANTED IN THE SUBCLAVIAN AND THE PATIENT IS REPORTED TO BE IN STABLE CONDITION.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9744028 ↗InjuryAug 16, 2019Obstruction of Flow
MalfunctionMay 14, 2019Migration; Migration
TECH NOTICED STENT WAS LOOSE AND NOT ON THE BALLOON, SO THEY DID NOT PUT IT IN THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8609573 ↗MalfunctionFeb 7, 2019Activation, Positioning or Separation Problem
InjuryNov 8, 2017Device Dislodged or Dislocated
IT WAS REPORTED DURING A PHYSICIAN MODIFIED ABDOMINAL AORTIC ANEURYSM (AAA) GRAFT CASE STENT WAS PLACED IN THE RIGHT RENAL ARTERY. ON BALLOON INFLATION THE PHYSICIAN NOTICED THAT THE STENT WAS NOT PRESENT ON THE BALLOON. ANOTHER STENT WAS USED AND SUCCESSFULLY DEPLOYED INTO INTENDED TARGET SITE. THE MISSING STENT WAS FOUND DISLODGED IN THE SHEATH AND WAS REMOVED. PATIENT HAD NO NEGATIVE EFFECTS OF THIS EVENT.
Patient problems: Aneurysm
MAUDE report 7010477 ↗InjurySep 12, 2017Break
IT WAS REPORTED THAT THE PHYSICIAN DELIVERED THE STENT TO THE TARGET AND DETERMINED IT WAS TOO LONG. HE REMOVED THE STENT SUCCESSFULLY AND COMMENTED THAT THE HUB WAS DAMAGED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6859310 ↗InjuryOct 10, 2016Difficult or Delayed Positioning
THE STENT COULD NOT BE IMPLANTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6013520 ↗InjurySep 23, 2016Mechanical Jam
IT WAS REPORTED THAT THE PHYSICIAN WAS ATTEMPTING TO STENT AN ILIAC ARTERY VIA A BRACHIAL APPROACH THROUGH A 7 FR ARROW SHEATH. AS THE DOCTOR WAS RETRACTING THE BALLOON THROUGH THE SHEATH, THE BALLOON GOT STUCK ON THE TAPERED END OF THE SHEATH.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 5973627 ↗InjuryMay 31, 2016Burst Container or Vessel; Inflation Problem; Pressure Problem
THE PHYSICIAN WAS USING THE STENT IN A KISSING STENT PROCEDURE TO TREAT BILATERAL COMMON ILIAC DISEASE. AFTER CROSSING EACH LESION THE STENT WAS POSITION AND THE SHEATH WITHDRAWN. THE LEFT STENT DID NOT INFLATE. THE BALLOON WOULD NOT HOLD PRESSURE. BLOOD WAS VISIBLE IN THE INSUFFLATOR INDICATING THAT THE BALLOON HAD BURST.
Patient problems: No Information
MAUDE report 5688893 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K050814. One submission often covers many models, so check the product and lot details in each record.
- Z-1077-2022 · Mar 3, 2022 · Open, ClassifiediCast Covered Stent System Product Code/Part Number: 85400 85401 85402 85403 85404 85405 85406 85407 85408 85409 85410 85411 85412 85413 85414 85415 85416 8541
- Z-2761-2017 · May 25, 2017 · TerminatedAtrium iCAST Balloon Expandable Covered Stent; 7mm x 22mm x 120cm; Model/Part Number 85455 (Marketed in U.S. only) Atrium Advanta V12 Balloon Expandable Covered Stent; 7mm x 22mm
Other Prosthesis, tracheal, expandable devices
- iCAST 85453Atrium Medical Corporation · Prosthesis, tracheal, expandable43 reports
- WALLSTENT™ BILIARY H965430500Boston Scientific Corporation · Prosthesis, tracheal, expandable13 reports
- iCAST 85419Atrium Medical Corporation · Prosthesis, tracheal, expandable13 reports
- iCAST 85445Atrium Medical Corporation · Prosthesis, tracheal, expandable12 reports
- iCAST 85416Atrium Medical Corporation · Prosthesis, tracheal, expandable12 reports
- iCAST 85455Atrium Medical Corporation · Prosthesis, tracheal, expandable11 reports
- WALLSTENT® Endoprosthesis H965404130Boston Scientific Corporation · Prosthesis, tracheal, expandable11 reports
- iCAST 85424Atrium Medical Corporation · Prosthesis, tracheal, expandable9 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page