STAT-Site M B-HB Strips 401010Recalls & reported problems

Stanbio Laboratory, L.P. · The STAT-Site® M β-HB Test is intended for the quantitative determination of the ketone body β-Hydroxybutyrate in serum or plasma using the STAT-Site® M β-HB Analyzer (Catalog #400400). The STAT-Site® M β-HB Test may be used with adults, infants, and children. For laboratory use only. · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Oct 2, 2023Class IIOpen, Classified

    Deterioration in the stability of the Test Strips results in diminished low-range detection capability and inability to meet linearity performance claims. The root cause of the stability deterioration in is identified as the change in material that caused a pH shift in the coating solution resulting in precipitation during coating runs. The device defect can cause false negative results that may lead to a delay in treatment of patients with associated conditions such as diabetic ketoacidosis.

    Z-0490-2024 · Root cause: Component change control · Recall details · FDA record ↗