CER-1Recalls & reported problems
Medivators Inc. · CER-1 · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00677964021695), received between Sep 7, 2018 and Sep 7, 2018 — classified as 1 injury.
The most frequently reported specific device problem is “Off-Label Use” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Off-Label Use1
- Use of Incorrect Control/Treatment Settings1
Patient problems
- Chemical Exposure1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjurySep 7, 2018Use of Incorrect Control/Treatment Settings; Off-Label Use; Use of Incorrect Control/Treatment Settings; Off-Label Use
IT WAS REPORTED THAT A FACILITY WAS REPROCESSING ENDOSCOPES WITH THEIR CER-1 AUTOMATIC ENDOSCOPE REPROCESSOR (AER) WITH THE INCORRECT AMOUNT OF RINSES PER THE HIGH-LEVEL DISINFECTANT LABELING AND AER INSTRUCTIONS-FOR-USE. THE FACILITY WAS USING CIDEX OPA HLD WHICH REQUIRES 3 RINSES BUT THEIR AER WAS SET UP WITH 2 RINSES. THERE IS POTENTIAL THAT ENDOSCOPES MAY HAVE CONTAINED HLD RESIDUAL AFTER REPROCESSING WHICH COULD CAUSE CHEMICAL COLITIS OR ADVERSE REACTIONS TO PATIENTS.
Patient problems: Chemical Exposure; No Known Impact Or Consequence To Patient
MAUDE report 7857564 ↗
Other Washer, Cleaner, Automated, Endoscope devices
- CER-2 OptimaMedivators Inc. · Washer, Cleaner, Automated, Endoscope2 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page