Skip to content

NA 1150001Recalls & reported problems

Medtronic Xomed, Inc. · PROSTHESIS 1150001 UNIV TITAN CAM HEAD · Class 2 · Implant

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00681490909907), received between Jan 24, 2020 and Jul 1, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

1 other recall cover devices cleared under the same FDA submission (K002897); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 1, 2024Break; Break

    IT WAS REPORTED THAT DURING LEFT TYMPANOPLASTY INSIDE THE EAR CANAL, WHILE THE DOCTOR WAS PLACING THE PROSTHESIS, IT FRACTURED. THE DOCTOR DECIDED TO REMOVE THE ROUND HEAD, LEAVING THE PROSTHESIS SHOE INSIDE THE PATIENT'S EAR. THE PROCEDURE WAS EXTENDED UPTO 2 HOURS. THE PROCEDURE WAS NOT COMPLETED. THE DOCTOR LEFT THE SHOE INSIDE THE EAR BUT REMOVED THE REST OF THE PROSTHESIS. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19642018 ↗
  • MalfunctionJul 1, 2024Break; Break

    IT WAS REPORTED THAT DURING LEFT TYMPANOPLASTY INSIDE THE EAR CANAL, WHILE THE DOCTOR WAS PLACING THE PROSTHESIS, IT FRACTURED. THE DOCTOR DECIDED TO REMOVE THE ROUND HEAD, LEAVING THE PROSTHESIS SHOE INSIDE THE PATIENT'S EAR. THE PROCEDURE WAS EXTENDED UPTO 2 HOURS. THE PROCEDURE WAS NOT COMPLETED. THE DOCTOR LEFT THE SHOE INSIDE THE EAR BUT REMOVED THE REST OF THE PROSTHESIS. THERE WAS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19642017 ↗
  • MalfunctionJan 24, 2020Break; Break

    A HEALTHCARE PROVIDER (HCP) REPORTED THAT DURING A PROCEDURE FOR A CHOLESTEATOMA RESECTION, ARTIFICIAL OSSICULAR IMPLANTATION AND TYMPANOPLASTY, THE ARTIFICIAL OSSICULAR IMPLANT WAS BROKEN DURING THE TYMPANIC PORTION OF THE SURGERY. THE SURGEON REMOVED THE BROKEN DEVICE AND ¿SWITCHED TO THE AUTOLOGOUS OSSICLE FOR HEARING RECONSTRUCTION AND COMPLETED THE OPERATION.¿ THE PATIENT WAS ALIVE WITH NO INJURY. ON FOLLOW UP, IT WAS CONFIRMED THERE WAS NO PROCEDURE DELAY. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9625835 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K002897. One submission often covers many models, so check the product and lot details in each record.

  • Z-2284-2013 · Aug 23, 2013 · Terminated
    Universal Titanium Prosthesis, Round Head, with Flex H/A titanium shoe. The ear prosthesis is implantable and indicated for hearing loss.

Other PROSTHESIS, PARTIAL OSSICULAR REPLACEMENT devices

All PROSTHESIS, PARTIAL OSSICULAR REPLACEMENT devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page