TRUWAVE, VAMP PXAVMPRecalls & reported problems

Edwards Lifesciences LLC · PRESSURE MONITORING SET · Class 2

All models of TRUWAVE, VAMP (19)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00690103011494), received between Oct 25, 2024 and Mar 13, 2025 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Disconnection” (5 reports).

1 FDA recall record name this device's identifier, 1 still open.

11 other recalls cover devices cleared under the same FDA submission (K142749, K173586, K896819); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 5 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520265
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Oct 4, 2023Class IIOpen, Classified

    Their is the potential for flushing difficulties during setup of the pressure monitoring kit with pressure transducer.

    Z-0414-2024 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Disconnection5
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Hemorrhage/Blood Loss/Bleeding1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 13, 2025Disconnection; Disconnection

    IT WAS REPORTED THAT THERE WAS A FAILURE WITH A PRESSURE MONITORING KIT. THE TUBING ON THE DEVICE BROKE AND BECAME DISCONNECTED ON THE PATIENT SIDE OF THE FLOW RESTRICTOR. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT HARM OR INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21595172 ↗
  • MalfunctionMar 13, 2025Disconnection; Disconnection

    IT WAS REPORTED THAT THERE WAS A FAILURE WITH A PRESSURE MONITORING KIT. THE TUBING ON THE DEVICE BROKE AND BECAME DISCONNECTED ON THE PATIENT SIDE OF THE FLOW RESTRICTOR. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT HARM OR INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21595164 ↗
  • MalfunctionMar 13, 2025Disconnection; Disconnection

    IT WAS REPORTED THAT THERE WAS A FAILURE WITH A PRESSURE MONITORING KIT. THE TUBING ON THE DEVICE BROKE AND BECAME DISCONNECTED ON THE PATIENT SIDE OF THE FLOW RESTRICTOR. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT HARM OR INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21595090 ↗
  • MalfunctionMar 13, 2025Disconnection; Disconnection

    IT WAS REPORTED THAT THERE WAS A FAILURE WITH A PRESSURE MONITORING KIT. THE TUBING ON THE DEVICE BROKE AND BECAME DISCONNECTED ON THE PATIENT SIDE OF THE FLOW RESTRICTOR. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT HARM OR INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21595070 ↗
  • MalfunctionMar 13, 2025Disconnection; Disconnection

    IT WAS REPORTED THAT THERE WAS A FAILURE WITH A PRESSURE MONITORING KIT. THE TUBING ON THE DEVICE BROKE AND BECAME DISCONNECTED ON THE PATIENT SIDE OF THE FLOW RESTRICTOR. THIS OCCURRED DURING PATIENT USE. THERE WAS NO PATIENT HARM OR INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21595050 ↗
  • MalfunctionNov 5, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED DURING USE THAT THE PXAVMP VAMP AND TUBING SEPARATED CAUSING BACKFLOW OF BLOOD OUT OF THE PATIENT WHEN THE PRIMARY NURSE WAS LEVELING THE TRANSDUCER OF THE ARTERIAL LINE. WITH TIMELY INTERVENTION THE LINE WAS PRESERVED AND PATIENT BLOOD LOSS WAS MINIMAL. THERE WAS NO SUBSTANTIAL PRESSURE APPLIED TO THE CONNECTION IN THE PROCESS OF LEVELING THE TRANSDUCER AND MINIMAL MANIPULATION OF THE LINE DURING MOVEMENT. ANOTHER IDENTICAL DEVICE WAS OBTAINED AND USED ON THE PATIENT AND NO HARM OCCURRED FROM THIS EVENT. SUBJECT WAS A 41 YEAR OLD MALE.

    Patient problems: Hemorrhage/Blood Loss/Bleeding

    MAUDE report 20614904 ↗
  • MalfunctionOct 25, 2024

    PRIMARY NURSE WAS LEVELING THE TRANSDUCER OF THE ARTERIAL LINE, THE VAMP AND TUBING SEPARATED CAUSING BACK FLOW OF BLOOD OUT OF THE PATIENT. WITH TIMELY INTERVENTION, THE LINE WAS PRESERVED, AND PATIENT BLOOD LOSS WAS MINIMAL. I WAS AT THE BEDSIDE ASSISTING WITH REPOSITIONING AT THE TIME, THERE WAS NO SUBSTANTIAL PRESSURE APPLIED TO THE CONNECTION IN THE PROCESS OF LEVELING THE TRANSDUCER AND MINIMAL MANIPULATION OF THE LINE DURING MOVEMENT. ANOTHER IDENTICAL DEVICE WAS OBTAINED AND USED ON PATIENT AND NO HARM OCCURRED FROM THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20532153 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K142749, K173586, K896819. One submission often covers many models, so check the product and lot details in each record.

  • Z-0413-2024 · Oct 4, 2023 · Open, Classified
    TruWave Pressure monitoring set, REF: PX212: cc/12 in (30cm), PX260: (3 cc/60in (150cm), PX272: 3 cc/72in (180cm), PX284: 3 cc/84 in (210 cm), PX2X2: x2 (3cc), PX3X3284C: X3 (3cc),
  • Z-1964-2021 · May 26, 2021 · Terminated
    Edwards TruWave (30 cc)/VAMP Jr Kit, REF: VMP306PX, Pressure Monitoring Set, Rx Only, CE, UDI: (01)57460691950370
  • Z-1966-2021 · May 26, 2021 · Terminated
    Edwards TruWave (3 cc)/VAMP Jr, REF: VMP406PX, Pressure Monitoring Set, Rx Only, CE UDI: (01)07460691958746
  • Z-1963-2021 · May 26, 2021 · Terminated
    Edwards VAMP Jr. 6 in (15 cm) REF: VMP306, Blood Management System, Rx Only, Single Use, CE, Sterile EO UDI: (01)57460691944782
  • Z-1961-2021 · May 26, 2021 · Terminated
    Edwards TruWave x 2 (3 cc)/VAMP Jr/54 in (136) Kit, REF: T443952B, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 57460691930174 ***N
  • Z-1959-2021 · May 26, 2021 · Terminated
    Edwards TruWave (3 cc)/VAMP Jr 48 in (120 cm) Kit, REF: PXVJ356, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 57460691954507
  • Z-1965-2021 · May 26, 2021 · Terminated
    Edwards TruWave 3 cc/VAMP Jr Kit, REF: VMP426PX, Pressure Monitoring Set, Rx Only, CE UDI: (01)57460691950547
  • Z-1962-2021 · May 26, 2021 · Terminated
    Edwards REF: VJ0990, Blood Management System, Rx Only, Single Use, CE, Sterile EO UDI: (01)57460691951520
  • Z-1960-2021 · May 26, 2021 · Terminated
    Edwards TruWave (30 cc)/VAMP Jr/52 in (132 cm) Kit, REF: PXVJ0711, Pressure Monitoring Set, RX Only, CE, Sterile EO, Sterilized using ethylene Oxide, UDI: (01) 07460691954489
  • Z-2776-2015 · Jul 31, 2015 · Terminated
    Edwards Pressure Monitoring Kit with TruWave disposable pressure transducer. A sterile, single-use kit that monitors pressures when attached to pressure monitoring catheters. For
  • Z-1146-2013 · Jan 30, 2013 · Terminated
    Pressure Monitoring sets with VAMP Plus closed blood sampling system model # T001691A, PXVP0566, PXVP2260, PXVP0826, PXVP0827, PXVP2284AT3, PXVP23X3AT3, T398807D, T100605B, T001762

Other Catheter, continuous flush devices

All Catheter, continuous flush devices →