SWAN-GANZ PACEPORT 991F8Recalls & reported problems
Edwards Lifesciences LLC · SWAN-GANZ THERMODILUTION A-V PACEPORT CATHETER · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00690103151961), received between Sep 12, 2022 and Sep 12, 2022 — classified as 1 injury.
The most frequently reported specific device problem is “Material Separation” (1 report).
4 other recalls cover devices cleared under the same FDA submission (K160084); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Material Separation1
Patient problems
- Cardiac Tamponade1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjurySep 12, 2022Material Separation; Material Separation
IT WAS REPORTED VIA MEDWATCH REPORT 0039 0111 2022 0018 THAT A PATIENT BECAME HYPOTENSIVE WITH VENTRICULAR ARRHYTHMIA UPON PLACEMENT OF A MODEL 991F8 SWAN GANZ PACING CATHETER. TRANSESOPHAGEAL ECHOCARDIOGRAM REVEALED A LARGE PERICARDIAL EFFUSION WITH TAMPONADE. A PERICARDIAL WINDOW WAS PERFORMED AND INJURY TO THE RIGHT VENTRICLE WAS FOUND, WHICH REQUIRED PRIMARY REPAIR WITH PLACEMENT OF A PERICARDIAL DRAIN. WHEN THE CATHETER WAS REMOVED, THE BALLOON WAS FOUND SHEARED OFF. THE DEVICE WAS SAVED FOR RETURN.
Patient problems: Cardiac Tamponade
MAUDE report 15402163 ↗
FDA clearance and classification
Cleared through 510(k) K160084 on May 3, 2016 (decision: Substantially Equivalent). Submitted by Edwards Lifesciences, LLC as “Swan Ganz Catheters”. The FDA review took 109 days. FDA record ↗
The FDA regulates product code KRA (catheter, continuous flush) under 21 CFR 870.1210 as a Class II device. 21 CFR 870.1210 ↗
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Recalls under the same FDA submission
These recalls involve devices cleared under K160084. One submission often covers many models, so check the product and lot details in each record.
- Z-0763-2019 · Dec 21, 2018 · TerminatedEdwards Lifesciences Swan Ganz Thermodilution Catheter, REF 131F7P (5 pack of 131F7) Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians
- Z-0762-2019 · Dec 21, 2018 · TerminatedEdwards Lifesciences Swan Ganz Thermodilution Catheter, REF131F 7J Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly determ
- Z-0764-2019 · Dec 21, 2018 · TerminatedEdwards Lifesciences Swan Ganz Thermodilution Catheter, REF 131VF7P Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly deter
- Z-0765-2019 · Dec 21, 2018 · TerminatedEdwards Lifesciences Swan Ganz Thermodilution Catheter, REF 151F7 Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly determi
Common questions about the SWAN-GANZ PACEPORT 991F8
Has the SWAN-GANZ PACEPORT 991F8 been recalled?
No FDA recall record in our data names this exact device identifier (DI 00690103151961), based on the FDA recall file dated Nov 27, 2026. However, 4 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the SWAN-GANZ PACEPORT 991F8?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Sep 12, 2022. That report lists device problem “Material Separation” (1); patient problem “Cardiac Tamponade” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the SWAN-GANZ PACEPORT 991F8?
None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the SWAN-GANZ PACEPORT 991F8 FDA approved?
It was cleared through 510(k) K160084 on May 3, 2016. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Is the SWAN-GANZ PACEPORT 991F8 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.
Who makes the SWAN-GANZ PACEPORT 991F8, and is it still sold?
The FDA's device database lists Edwards Lifesciences LLC as the company responsible for it (brand SWAN-GANZ PACEPORT, model 991F8). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Catheter, continuous flush devices
- PROWLER SELECT 606-S255XMedos International Sàrl · Catheter, continuous flush2 deaths973 reports
- PhenomTM 27 Catheter FG15150-0615-1SMicro Therapeutics, Inc. · Catheter, continuous flush2 deaths139 reports
- MarathonTM 105-5056Micro Therapeutics, Inc. · Catheter, continuous flush124 reports
- Renegade™ HI-FLO™ Kit M001183020Boston Scientific Corporation · Catheter, continuous flush109 reports
- MarksmanTM FA-55150-1030Micro Therapeutics, Inc. · Catheter, continuous flush109 reports
- EkoSonic® 6779-002Boston Scientific Corporation · Catheter, Continuous Flush1 recall4 deaths55 reports
- Chameleon CH08-40-75USMozarc Medical US LLC · Catheter, continuous flush2 recalls4 reports
- RebarTM 105-5081-153Micro Therapeutics, Inc. · Catheter, continuous flush2 deaths102 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page