CARPENTIER-EDWARDS PERIMOUNT MAGNA EASE PERICARDIAL BIOPROSTHESIS - AORTIC 3300TFX (REF 3300TFX23MM)Recalls & reported problems
Edwards Lifesciences LLC · CARPENTIER-EDWARDS PERIMOUNT MAGNA EASE PERICARDIAL AORTIC BIOPROSTHESIS · Class 3 · Implant
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Edwards PERIMOUNT Magna Ease aortic valve has 679 adverse event reports, with 4 deaths and 673 injuries. Common problems include the valve narrowing, rising pressure across the valve and calcium buildup; patients mention shortness of breath and heart failure, and many reports give little detail. No recalls name this device. Reports dropped sharply, with 52 in the last year compared with 198 in the year before.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 671
- Deaths reported
- 4
- Injuries reported
- 665
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 671 adverse event reports that list this exact device (DI 00690103176148), received between Feb 22, 2021 and Sep 14, 2026 — classified as 4 deaths, 665 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (323 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Insufficient Device Problem Information323
- Gradient Increase175
- Device Stenosis168
- Calcified85
- Fluid/Blood Leak59
- Degraded54
- Adverse Event Without Identified Device or Use Problem24
- Perivalvular Leak20
Patient problems
- Insufficient Information457
- Dyspnea200
- Heart Failure/Congestive Heart Failure150
- Fatigue45
- Chest Pain27
- Endocarditis26
- Dizziness17
- No Clinical Signs, Symptoms or Conditions15
Recent adverse event reports
The 25 most recent of 671 reports, in the reporter's own words as published by the FDA.
InjurySep 14, 2026Insufficient Device Problem Information
It was learned through Implant Patient Registry that a patient with a 23mm 3300TFX aortic valve was explanted after an implant duration of 7 years due to unknown reasons. The explanted valve was replaced with a 23mm 11060A valve.Edwards Lifesciences maintains an Implant Patient Registry. This Registry is a patient tracking mechanism for serialized Edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true Post-Market Surveillance Registry. Through the registry, Edwards is notified when these devices are implanted. In addition, patient and/or device status may be reported to the Registry via the implantation data cards. The information is received from various sources (e.g. surgeon, hospital, and patient family members) and is not received in the form of a conventional 'customer complaint'. The information reported may or may not be related to the Edwards device.
Patient problems: Insufficient Information
MAUDE report 50753537 ↗InjuryAug 3, 2026Calcified; Insufficient Device Problem Information
It was learned through Implant Patient Registry that a patient with a 23mm 3300TFX aortic valve was explanted after an implant duration of 9 years, 8 months due to unknown reasons. The explanted valve was replaced with a 23mm 11500A valve.Edwards Lifesciences maintains an Implant Patient Registry. This Registry is a patient tracking mechanism for serialized Edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true Post-Market Surveillance Registry. Through the registry, Edwards is notified when these devices are implanted. In addition, patient and/or device status may be reported to the Registry via the implantation data cards. The information is received from various sources (e.g. surgeon, hospital, and patient family members) and is not received in the form of a conventional 'customer complaint'. The information reported may or may not be related to the Edwards device.
Patient problems: Dyspnea; Insufficient Information
MAUDE report 50114147 ↗MalfunctionJul 28, 2026Defective Device
\[redacted]"--- When replacing the aortic valve, the first tissue valve was not functioning properly. It was discovered on TEE \[transesophageal echocardiogram]" when trying to come off bypass. Decided by MD to replace the valve. Procedure required more time on bypass. Second valve worked properly.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50053696 ↗InjuryJun 9, 2026Insufficient Device Problem Information
IT WAS LEARNED THROUGH IMPLANT PATIENT REGISTRY THAT A PATIENT WITH A 23MM 3300TFX AORTIC VALVE WAS EXPLANTED AFTER AN IMPLANT DURATION OF 9 YEARS, 9 MONTHS DUE TO UNKNOWN REASONS. THE EXPLANTED VALVE WAS REPLACED WITH A 23MM 11500A VALVE. EDWARDS LIFESCIENCES MAINTAINS AN IMPLANT PATIENT REGISTRY. THIS REGISTRY IS A PATIENT TRACKING MECHANISM FOR SERIALIZED EDWARDS IMPLANTABLE DEVICES (BIOPROSTHETIC HEART VALVES AND ANNULOPLASTY RINGS), RATHER THAN A TRUE POST-MARKET SURVEILLANCE REGISTRY. THROUGH THE REGISTRY, EDWARDS IS NOTIFIED WHEN THESE DEVICES ARE IMPLANTED. IN ADDITION, PATIENT AND/OR DEVICE STATUS MAY BE REPORTED TO THE REGISTRY VIA THE IMPLANTATION DATA CARDS. THE INFORMATION IS RECEIVED FROM VARIOUS SOURCES (E.G. SURGEON, HOSPITAL, AND PATIENT FAMILY MEMBERS) AND IS NOT RECEIVED IN THE FORM OF A CONVENTIONAL 'CUSTOMER COMPLAINT'. THE INFORMATION REPORTED MAY OR MAY NOT BE RELATED TO THE EDWARDS DEVICE.
Patient problems: Insufficient Information
MAUDE report 25447965 ↗InjuryMay 20, 2026Insufficient Device Problem Information
IT WAS LEARNED THROUGH IMPLANT PATIENT REGISTRY THAT A 23MM 3300TFX AORTIC VALVE WAS EXPLANTED AFTER AN IMPLANT DURATION OF 9 YEARS, 9 MONTHS DUE TO UNKNOWN REASON. THE EXPLANTED DEVICE WAS REPLACED WITH A 23MM 11500A AORTIC VALVE. THE PATIENT WAS IN RECOVERY POST PROCEDURE. EDWARDS LIFESCIENCES MAINTAINS AN IMPLANT PATIENT REGISTRY. THIS REGISTRY IS A PATIENT TRACKING MECHANISM FOR SERIALIZED EDWARDS IMPLANTABLE DEVICES (BIOPROSTHETIC HEART VALVES AND ANNULOPLASTY RINGS), RATHER THAN A TRUE POST-MARKET SURVEILLANCE REGISTRY. THROUGH THE REGISTRY, EDWARDS IS NOTIFIED WHEN THESE DEVICES ARE IMPLANTED. IN ADDITION, PATIENT AND/OR DEVICE STATUS MAY BE REPORTED TO THE REGISTRY VIA THE IMPLANTATION DATA CARDS. THE INFORMATION IS RECEIVED FROM VARIOUS SOURCES (E.G. SURGEON, HOSPITAL, AND PATIENT FAMILY MEMBERS) AND IS NOT RECEIVED IN THE FORM OF A CONVENTIONAL "CUSTOMER COMPLAINT". THE INFORMATION REPORTED MAY OR MAY NOT BE RELATED TO THE EDWARDS DEVICE.
Patient problems: Insufficient Information
MAUDE report 25231873 ↗InjuryMay 18, 2026Calcified
IT WAS LEARNED THROUGH IMPLANT PATIENT REGISTRY AND INVESTIGATION THAT A PATIENT WITH A 23MM 3300TFX AORTIC VALVE WAS EXPLANTED AFTER AN IMPLANT DURATION OF 8 YEARS, 6 MONTHS DUE TO CALCIFICATION AND STENOSIS. THE PATIENT PRESENTED WITH SOB AND CHEST PAIN. THE EXPLANTED VALVE WAS REPLACED WITH A 25MM 11500A VALVE. THE PATIENT WAS TAKEN TO THE ICU POST PROCEDURE. EDWARDS LIFESCIENCES MAINTAINS AN IMPLANT PATIENT REGISTRY. THIS REGISTRY IS A PATIENT TRACKING MECHANISM FOR SERIALIZED EDWARDS IMPLANTABLE DEVICES (BIOPROSTHETIC HEART VALVES AND ANNULOPLASTY RINGS), RATHER THAN A TRUE POST-MARKET SURVEILLANCE REGISTRY. THROUGH THE REGISTRY, EDWARDS IS NOTIFIED WHEN THESE DEVICES ARE IMPLANTED. IN ADDITION, PATIENT AND/OR DEVICE STATUS MAY BE REPORTED TO THE REGISTRY VIA THE IMPLANTATION DATA CARDS. THE INFORMATION IS RECEIVED FROM VARIOUS SOURCES (E.G. SURGEON, HOSPITAL, AND PATIENT FAMILY MEMBERS) AND IS NOT RECEIVED IN THE FORM OF A CONVENTIONAL 'CUSTOMER COMPLAINT'. THE INFORMATION REPORTED MAY OR MAY NOT BE RELATED TO THE EDWARDS DEVICE.
Patient problems: Dyspnea
MAUDE report 25209010 ↗InjuryApr 29, 2026Insufficient Device Problem Information
IT WAS LEARNED THROUGH IMPLANT PATIENT REGISTRY THAT A PATIENT WITH A 23MM 3300TFX AORTIC VALVE WAS EXPLANTED AFTER AN IMPLANT DURATION OF 7 YEARS, 5 MONTHS DUE TO UNKNOWN REASONS. THE EXPLANTED VALVE WAS REPLACED WITH A 23MM 11500A VALVE. EDWARDS LIFESCIENCES MAINTAINS AN IMPLANT PATIENT REGISTRY. THIS REGISTRY IS A PATIENT TRACKING MECHANISM FOR SERIALIZED EDWARDS IMPLANTABLE DEVICES (BIOPROSTHETIC HEART VALVES AND ANNULOPLASTY RINGS), RATHER THAN A TRUE POST-MARKET SURVEILLANCE REGISTRY. THROUGH THE REGISTRY, EDWARDS IS NOTIFIED WHEN THESE DEVICES ARE IMPLANTED. IN ADDITION, PATIENT AND/OR DEVICE STATUS MAY BE REPORTED TO THE REGISTRY VIA THE IMPLANTATION DATA CARDS. THE INFORMATION IS RECEIVED FROM VARIOUS SOURCES (E.G. SURGEON, HOSPITAL, AND PATIENT FAMILY MEMBERS) AND IS NOT RECEIVED IN THE FORM OF A CONVENTIONAL 'CUSTOMER COMPLAINT'. THE INFORMATION REPORTED MAY OR MAY NOT BE RELATED TO THE EDWARDS DEVICE.
Patient problems: Insufficient Information
MAUDE report 25033436 ↗InjuryApr 26, 2026Degraded
IT WAS REPORTED THAT A PATIENT WITH A 23MM 3300TFX AORTIC VALVE UNDERWENT A VALVE-IN-VALVE PROCEDURE AFTER AN IMPLANT DURATION OF (B)(6) DUE TO DEGENERATION AND STENOSIS. THE PATIENT PRESENTED WITH SHORTNESS OF BREATH AND FATIGUE. THE TAVR PROCEDURE WAS SUCCESSFULLY PERFORMED WITH A 23MM 9755RSL VALVE. THE PATIENT WAS STABLE AND IN RECOVERY POST-PROCEDURE. ON POD #7 PPM WAS PLACED DUE TO HEART BLOCK AND THE PATIENT DISCHARGED ON POD #8.
Patient problems: Dyspnea; Heart Block
MAUDE report 24993634 ↗InjuryApr 10, 2026Insufficient Device Problem Information; Device Stenosis
IT WAS LEARNED THROUGH IMPLANT PATIENT REGISTRY AND INVESTIGATION THAT A PATIENT WITH A 23MM 3300TFX AORTIC VALVE WAS EXPLANTED AFTER AN IMPLANT DURATION OF 7 YEARS, 11 MONTHS DUE TO HIGH GRADIENTS. THE PATIENT PRESENTED WITH SHORTNESS OF BREATH. THE EXPLANTED VALVE WAS REPLACED WITH A 23MM 11500A VALVE. THE PATIENT WAS IN RECOVERY. THE PATIENT WAS IN STABLE CONDITION POST PROCEDURE AND WAS DISCHARGED ON POD#5.
Patient problems: Dyspnea; Insufficient Information
MAUDE report 24841629 ↗InjuryApr 8, 2026Calcified
IT WAS LEARNED THROUGH IMPLANT PATIENT REGISTRY AND INVESTIGATION THAT A PATIENT WITH A 23MM 3300TFX AORTIC VALVE WAS EXPLANTED AFTER AN IMPLANT DURATION OF 8 YEARS, 4 MONTHS DUE TO CALCIFICATION WITH STENOSIS. THE EXPLANTED VALVE WAS REPLACED WITH A 29MM 11500A VALVE. PER MEDICAL RECORDS, THE PATIENT PRESENTED WITH NO CARDIAC SYMPTOMS BUT HAD INCREASING GRADIENTS. HE UNDERWENT REDO AVR, THE 3300TFX WAS EXPLANTED AND ONE OF THE LEAFLETS WAS CALCIFIED, PRESUMED THIS CONTRIBUTED TO THE STENOSIS. AFTER DEBRIDEMENT OF THE ANNULUS, A 29MM 11500A VALVE WAS IMPLANTED WITH PLEDGETED SUTURES, SEATED, AND SECURED WITH COR-KNOTS. THE PATIENT REMAINED STABLE, RETURNED TO THE SURGICAL ICU, AND WAS DISCHARGED ON POD #6. HOSPITAL PATHOLOGY REPORT: GROSS EXAM, PROSTHETIC AORTIC VALVE WITH ATTACHED LEAFLETS, ONE LEAFLET HAS AN AREA OF WHITE EXUDATE. ALSO RECEIVED ARE MULTIPLE WHITE FRAGMENT OF SOFT TISSUE WITH CALCIFICATIONS. FINAL DIAGNOSIS: AORTIC VALVE, NEGATIVE FOR ACTIVE INFLAMMATION. EDWARDS LIFESCIENCES MAINTAINS AN IMPLANT PATIENT REGISTRY. THIS REGISTRY IS A PATIENT TRACKING MECHANISM FOR SERIALIZED EDWARDS IMPLANTABLE DEVICES (BIOPROSTHETIC HEART VALVES AND ANNULOPLASTY RINGS), RATHER THA
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24824602 ↗InjuryApr 8, 2026Degraded
IT WAS LEARNED THROUGH IMPLANT PATIENT REGISTRY AND INVESTIGATION THAT A PATIENT WITH A 23MM 3300TFX AORTIC VALVE WAS EXPLANTED AFTER AN IMPLANT DURATION OF 6 YEARS, 6 MONTHS DUE TO DEGENERATION, SEVERE STENOSIS AND REGURGITATION. THE EXPLANTED VALVE WAS REPLACED WITH A 23MM 11060A AORTIC VALVED CONDUIT. PER MEDICAL RECORDS, THE PATIENT PRESENTED WITH HEART FAILURE NYHA-3 AND UNDERWENT REDO AVR WITH BENTALL PROCEDURE. THE AORTIC VALVE WAS REMOVED AND REPLACED WITH A 23MM AVC WITH MATTRESS SUTURES AND SECURED WITH COR-KNOTS. POST BYPASS TEE SHOWED WELL-SEATED AORTIC VALVE WITH NO PVL. THE PATIENT WAS TRANSFERRED TO THE ICU IN STABLE CONDITION AND WAS DISCHARGED ON POD #17. HOSPITAL PATHOLOGY REPORT: GROSS EXAM ONLY, AORTIC VALVE WITH MILD DEGENERATIVE CHANGES. EDWARDS LIFESCIENCES MAINTAINS AN IMPLANT PATIENT REGISTRY. THIS REGISTRY IS A PATIENT TRACKING MECHANISM FOR SERIALIZED EDWARDS IMPLANTABLE DEVICES (BIOPROSTHETIC HEART VALVES AND ANNULOPLASTY RINGS), RATHER THAN A TRUE POST-MARKET SURVEILLANCE REGISTRY. THROUGH THE REGISTRY, EDWARDS IS NOTIFIED WHEN THESE DEVICES ARE IMPLANTED. IN ADDITION, PATIENT AND/OR DEVICE STATUS MAY BE REPORTED TO THE REGISTRY VIA THE IMPLANTATION DAT
Patient problems: Heart Failure/Congestive Heart Failure
MAUDE report 24824585 ↗InjuryMar 31, 2026Perivalvular Leak; Insufficient Device Problem Information
IT WAS REPORTED VIA THE IMPLANT PATIENT REGISTRY THAT A PATIENT WITH A 23MM 3300TFX AORTIC VALVE WAS EXPLANTED AFTER AN IMPLANT DURATION OF 7 DAYS DUE TO UNKNOWN REASON. THE EXPLANTED VALVE WAS REPLACED WITH A 21MM 3300TFX VALVE. EDWARDS LIFESCIENCES MAINTAINS AN IMPLANT PATIENT REGISTRY. THIS REGISTRY IS A PATIENT TRACKING MECHANISM FOR SERIALIZED EDWARDS IMPLANTABLE DEVICES (BIOPROSTHETIC HEART VALVES AND ANNULOPLASTY RINGS), RATHER THAN A TRUE POST-MARKET SURVEILLANCE REGISTRY. THROUGH THE REGISTRY, EDWARDS IS NOTIFIED WHEN THESE DEVICES ARE IMPLANTED. IN ADDITION, PATIENT AND/OR DEVICE STATUS MAY BE REPORTED TO THE REGISTRY VIA THE IMPLANTATION DATA CARDS. THE INFORMATION IS RECEIVED FROM VARIOUS SOURCES (E.G. SURGEON, HOSPITAL, AND PATIENT FAMILY MEMBERS) AND IS NOT RECEIVED IN THE FORM OF A CONVENTIONAL "CUSTOMER COMPLAINT". THE INFORMATION REPORTED MAY OR MAY NOT BE RELATED TO THE EDWARDS DEVICE.
Patient problems: Hemodynamic instability; Insufficient Information
MAUDE report 24750076 ↗InjuryMar 31, 2026Insufficient Device Problem Information; Device Stenosis
IT WAS LEARNED THROUGH IMPLANT PATIENT REGISTRY THAT A 23MM 3300TFX MAGNA EASE VALVE IN THE AORTIC POSITION WAS EXPLANTED AND REPLACED WITH A 23MM 11500A INSPIRIS RESILIA AORTIC VALVE AFTER AN IMPLANT DURATION OF 7 YEARS, 28 DAYS. REASON FOR REINTERVENTION IS UNKNOWN. EDWARDS LIFESCIENCES MAINTAINS AN IMPLANT PATIENT REGISTRY. THIS REGISTRY IS A PATIENT TRACKING MECHANISM FOR SERIALIZED EDWARDS IMPLANTABLE DEVICES (BIOPROSTHETIC HEART VALVES AND ANNULOPLASTY RINGS), RATHER THAN A TRUE POST-MARKET SURVEILLANCE REGISTRY. THROUGH THE REGISTRY, EDWARDS IS NOTIFIED WHEN THESE DEVICES ARE IMPLANTED. IN ADDITION, PATIENT AND/OR DEVICE STATUS MAY BE REPORTED TO THE REGISTRY VIA THE IMPLANTATION DATA CARDS. THE INFORMATION IS RECEIVED FROM VARIOUS SOURCES (E.G. SURGEON, HOSPITAL, AND PATIENT FAMILY MEMBERS) AND IS NOT RECEIVED IN THE FORM OF A CONVENTIONAL "CUSTOMER COMPLAINT". THE INFORMATION REPORTED MAY OR MAY NOT BE RELATED TO THE EDWARDS DEVICE.
Patient problems: Chest Pain; Dizziness; Fatigue; Insufficient Information
MAUDE report 24749604 ↗InjuryMar 30, 2026Device Stenosis
IT WAS REPORTED AND LEARNED THROUGH IMPLANT PATIENT REGISTRY THAT A PATIENT WITH A 27MM 3300TFX AORTIC VALVE WAS EXPLANTED AFTER AN IMPLANT DURATION OF 8 YEARS, 10 MONTHS DUE TO STENOSIS. PATIENT PRESENTED WITH FATIGUE. THE EXPLANTED VALVE WAS REPLACED WITH A 21MM 11500A INSPIRIS RESILIA AORTIC VALVE.
Patient problems: Fatigue
MAUDE report 24728121 ↗InjuryMar 20, 2026Calcified; Degraded; Insufficient Device Problem Information
IT WAS LEARNED THROUGH IMPLANT PATIENT REGISTRY AND INVESTIGATION THAT A 23MM 3300TFX MAGNA EASE VALVE IN AORTIC POSITION WAS EXPLANTED AND REPLACED WITH A 23MM 11500A INSPIRIS RESILIA AORTIC VALVE AFTER AN IMPLANT DURATION OF 9 YEARS, 2 MONTHS DUE TO DEGENERATION AND CALCIFICATION. THE PATIENT PRESENTED WITH HEART FAILURE AND SHORTNESS OF BREATH. PATIENT WAS RECOVERED POST-OPERATIVELY.
Patient problems: Dyspnea; Heart Failure/Congestive Heart Failure; Insufficient Information
MAUDE report 24656170 ↗InjuryMar 20, 2026Insufficient Device Problem Information; Material Split, Cut or Torn
IT WAS LEARNED THROUGH IMPLANT PATIENT REGISTRY AND INVESTIGATION THAT A 23MM 3300TFX MAGNA EASE AORTIC VALVE IN AORTIC POSITION WAS EXPLANTED AFTER AN IMPLANT DURATION OF 6 YEARS, 8 MONTHS DUE TO LEAFLET PROLAPSE RESULTING IN REGURGITATION. THE PATIENT PRESENTED WITH HEART FAILURE AND SHORTNESS OF BREATH. THE REPLACEMENT VALVE IS A 23MM 11500A INSPIRIS RESILIA AORTIC VALVE.
Patient problems: Dyspnea; Heart Failure/Congestive Heart Failure; Insufficient Information
MAUDE report 24656130 ↗InjuryMar 20, 2026Insufficient Device Problem Information
IT WAS LEARNED THROUGH IMPLANT PATIENT REGISTRY THAT A 23MM 3300TFX AORTIC VALVE WAS EXPLANTED AFTER AN IMPLANT DURATION OF 8 YEARS AND 5 MONTHS DUE TO UNKNOWN REASON. THE VALVE WAS REPLACED WITH A 11500A 25MM VALVE.
Patient problems: Insufficient Information
MAUDE report 24654543 ↗InjuryMar 20, 2026Insufficient Device Problem Information
IT WAS REPORTED THAT A 23MM 3300TFX VALVE IN AORTIC POSITION UNDERWENT AN ATTEMPTED VIV PROCEDURE AFTER AN IMPLANT DURATION OF SEVEN (7) YEARS, NINE (9) MONTHS DUE TO UNKNOWN REASONS. DURING THE PROCEDURE, AFTER A FEW ATTEMPTS WERE MADE WITH THE COMMANDER SYSTEM OF THE SAPIEN VALVE, WITH DIFFICULTY KEEPING THE WIRE IN THE VENTRICLE, AND CONVERSATIONS ABOUT WHAT WOULD BE BEST AND SAFEST FOR THE PATIENT, THEY DECIDED TO ABORT THE CASE AND DISCUSS OPTIONS WITH THE PATIENT. THE AORTIC REPLACEMENT, ANGULATION, AND DIFFICULTY CROSSING A WIRE THROUGH THE HEAVILY CALCIFIED ANNULUS WERE ALL REASONS TO ABORT AND CONSIDER ALTERNATIVE OPTIONS. THE IMPLANTING PHYSICIAN ALSO STATED THAT THE SURGICAL VALVE WAS TOO SMALL FOR THE ANATOMY, CAUSING AN EARLY FAILURE.
Patient problems: Insufficient Information
MAUDE report 24653635 ↗InjuryMar 14, 2026Calcified
IT WAS LEARNED THROUGH IMPLANT PATIENT REGISTRY AND INVESTIGATION THAT A PATIENT WITH A 23MM 3300TFX AORTIC VALVE WAS EXPLANTED AFTER AN IMPLANT DURATION OF 8 YEARS, 3 MONTHS DUE TO CALCIFICATION. THE PATIENT PRESENTED WITH SHORTNESS OF BREATH AND VERTIGO WITH STANDING. THE EXPLANTED VALVE WAS REPLACED WITH A 25MM 11500A VALVE. THE PATIENT WAS IN STABLE CONDITION POST PROCEDURE AND WAS DISCHARGE ON POD#7. EDWARDS LIFESCIENCES MAINTAINS AN IMPLANT PATIENT REGISTRY. THIS REGISTRY IS A PATIENT TRACKING MECHANISM FOR SERIALIZED EDWARDS IMPLANTABLE DEVICES (BIOPROSTHETIC HEART VALVES AND ANNULOPLASTY RINGS), RATHER THAN A TRUE POST-MARKET SURVEILLANCE REGISTRY. THROUGH THE REGISTRY, EDWARDS IS NOTIFIED WHEN THESE DEVICES ARE IMPLANTED. IN ADDITION, PATIENT AND/OR DEVICE STATUS MAY BE REPORTED TO THE REGISTRY VIA THE IMPLANTATION DATA CARDS. THE INFORMATION IS RECEIVED FROM VARIOUS SOURCES (E.G. SURGEON, HOSPITAL, AND PATIENT FAMILY MEMBERS) AND IS NOT RECEIVED IN THE FORM OF A CONVENTIONAL CUSTOMER COMPLAINT. THE INFORMATION REPORTED MAY OR MAY NOT BE RELATED TO THE EDWARDS DEVICE.
Patient problems: Dyspnea; Vertigo
MAUDE report 24599683 ↗InjuryFeb 27, 2026Insufficient Device Problem Information
IT WAS LEARNED VIA THE PATIENT REGISTRY THAT A PATIENT WITH A 23MM 3300TFX PERICARDIAL VALVE IN THE PULMONARY POSITION UNDERWENT REDO PULMONARY VALVE REPLACEMENT AFTER AN IMPLANT DURATION OF SEVEN (7) YEARS DUE TO UNKNOWN REASONS. THE EXPLANTED VALVE WAS REPLACED WITH A 27MM 11500A INSPIRIS VALVE. EDWARDS LIFESCIENCES MAINTAINS AN IMPLANT PATIENT REGISTRY. THIS REGISTRY IS A PATIENT TRACKING MECHANISM FOR SERIALIZED EDWARDS IMPLANTABLE DEVICES (BIOPROSTHETIC HEART VALVES AND ANNULOPLASTY RINGS), RATHER THAN A TRUE POST-MARKET SURVEILLANCE REGISTRY. THROUGH THE REGISTRY, EDWARDS IS NOTIFIED WHEN THESE DEVICES ARE IMPLANTED. IN ADDITION, PATIENT AND/OR DEVICE STATUS MAY BE REPORTED TO THE REGISTRY VIA THE IMPLANTATION DATA CARDS. THE INFORMATION IS RECEIVED FROM VARIOUS SOURCES (E.G. SURGEON, HOSPITAL, AND PATIENT FAMILY MEMBERS) AND IS NOT RECEIVED IN THE FORM OF A CONVENTIONAL 'CUSTOMER COMPLAINT'. THE INFORMATION REPORTED MAY OR MAY NOT BE RELATED TO THE EDWARDS DEVICE.
Patient problems: Insufficient Information
MAUDE report 24472494 ↗InjuryFeb 19, 2026Insufficient Device Problem Information
IT WAS LEARNED THROUGH IMPLANT PATIENT REGISTRY THAT A PATIENT WITH A 23MM 3300TFX AORTIC VALVE WAS EXPLANTED AFTER AN IMPLANT DURATION OF 6 YEARS, 10 MONTHS DUE TO UNKNOWN REASONS. THE EXPLANTED VALVE WAS REPLACED WITH A 23MM 11500A VALVE. EDWARDS LIFESCIENCES MAINTAINS AN IMPLANT PATIENT REGISTRY. THIS REGISTRY IS A PATIENT TRACKING MECHANISM FOR SERIALIZED EDWARDS IMPLANTABLE DEVICES (BIOPROSTHETIC HEART VALVES AND ANNULOPLASTY RINGS), RATHER THAN A TRUE POST-MARKET SURVEILLANCE REGISTRY. THROUGH THE REGISTRY, EDWARDS IS NOTIFIED WHEN THESE DEVICES ARE IMPLANTED. IN ADDITION, PATIENT AND/OR DEVICE STATUS MAY BE REPORTED TO THE REGISTRY VIA THE IMPLANTATION DATA CARDS. THE INFORMATION IS RECEIVED FROM VARIOUS SOURCES (E.G. SURGEON, HOSPITAL, AND PATIENT FAMILY MEMBERS) AND IS NOT RECEIVED IN THE FORM OF A CONVENTIONAL 'CUSTOMER COMPLAINT'. THE INFORMATION REPORTED MAY OR MAY NOT BE RELATED TO THE EDWARDS DEVICE.
Patient problems: Insufficient Information
MAUDE report 24393072 ↗InjuryFeb 19, 2026Insufficient Device Problem Information; Device Stenosis
IT WAS LEARNED THROUGH IMPLANT PATIENT REGISTRY THAT A 23MM 3300TFX AORTIC VALVE WAS EXPLANTED AFTER AN IMPLANT DURATION OF 8 YEARS, 1 MONTH DUE TO UNKNOWN REASON. THE EXPLANTED DEVICE WAS REPLACED WITH A 23MM 11500A AORTIC VALVE. EDWARDS LIFESCIENCES MAINTAINS AN IMPLANT PATIENT REGISTRY. THIS REGISTRY IS A PATIENT TRACKING MECHANISM FOR SERIALIZED EDWARDS IMPLANTABLE DEVICES (BIOPROSTHETIC HEART VALVES AND ANNULOPLASTY RINGS), RATHER THAN A TRUE POST-MARKET SURVEILLANCE REGISTRY. THROUGH THE REGISTRY, EDWARDS IS NOTIFIED WHEN THESE DEVICES ARE IMPLANTED. IN ADDITION, PATIENT AND/OR DEVICE STATUS MAY BE REPORTED TO THE REGISTRY VIA THE IMPLANTATION DATA CARDS. THE INFORMATION IS RECEIVED FROM VARIOUS SOURCES (E.G. SURGEON, HOSPITAL, AND PATIENT FAMILY MEMBERS) AND IS NOT RECEIVED IN THE FORM OF A CONVENTIONAL "CUSTOMER COMPLAINT". THE INFORMATION REPORTED MAY OR MAY NOT BE RELATED TO THE EDWARDS DEVICE.
Patient problems: Heart Failure/Congestive Heart Failure; Insufficient Information
MAUDE report 24388372 ↗InjuryFeb 13, 2026Insufficient Device Problem Information
IT WAS LEARNED VIA THE PATIENT REGISTRY THAT A PATIENT WITH A 23MM 3300TFX PERICARDIAL AORTIC VALVE UNDERWENT REDO AORTIC VALVE REPLACEMENT AFTER AN IMPLANT DURATION OF FIVE (5) YEARS, 11 MONTHS DUE TO UNKNOWN REASONS. THE EXPLANTED VALVE WAS REPLACED WITH A 27MM 11500A INSPIRIS VALVE. EDWARDS LIFESCIENCES MAINTAINS AN IMPLANT PATIENT REGISTRY. THIS REGISTRY IS A PATIENT TRACKING MECHANISM FOR SERIALIZED EDWARDS IMPLANTABLE DEVICES (BIOPROSTHETIC HEART VALVES AND ANNULOPLASTY RINGS), RATHER THAN A TRUE POST-MARKET SURVEILLANCE REGISTRY. THROUGH THE REGISTRY, EDWARDS IS NOTIFIED WHEN THESE DEVICES ARE IMPLANTED. IN ADDITION, PATIENT AND/OR DEVICE STATUS MAY BE REPORTED TO THE REGISTRY VIA THE IMPLANTATION DATA CARDS. THE INFORMATION IS RECEIVED FROM VARIOUS SOURCES (E.G. SURGEON, HOSPITAL, AND PATIENT FAMILY MEMBERS) AND IS NOT RECEIVED IN THE FORM OF A CONVENTIONAL 'CUSTOMER COMPLAINT'. THE INFORMATION REPORTED MAY OR MAY NOT BE RELATED TO THE EDWARDS DEVICE.
Patient problems: Insufficient Information
MAUDE report 24338697 ↗InjuryFeb 12, 2026Degraded
IT WAS LEARNED THROUGH IMPLANT PATIENT REGISTRY AND INVESTIGATION THAT A PATIENT WITH A 23MM 3300TFX AORTIC VALVE WAS EXPLANTED AFTER AN IMPLANT DURATION OF 8 YEARS, 9 MONTHS DUE TO DEGENERATION. THE PATIENT PRESENTED WITH SOB. THE EXPLANTED VALVE WAS REPLACED WITH A 25MM 11500A VALVE. THE PATIENT WAS IN STABLE CONDITION POST PROCEDURE AND DISCHARGED POD #11. EDWARDS LIFESCIENCES MAINTAINS AN IMPLANT PATIENT REGISTRY. THIS REGISTRY IS A PATIENT TRACKING MECHANISM FOR SERIALIZED EDWARDS IMPLANTABLE DEVICES (BIOPROSTHETIC HEART VALVES AND ANNULOPLASTY RINGS), RATHER THAN A TRUE POST-MARKET SURVEILLANCE REGISTRY. THROUGH THE REGISTRY, EDWARDS IS NOTIFIED WHEN THESE DEVICES ARE IMPLANTED. IN ADDITION, PATIENT AND/OR DEVICE STATUS MAY BE REPORTED TO THE REGISTRY VIA THE IMPLANTATION DATA CARDS. THE INFORMATION IS RECEIVED FROM VARIOUS SOURCES (E.G. SURGEON, HOSPITAL, AND PATIENT FAMILY MEMBERS) AND IS NOT RECEIVED IN THE FORM OF A CONVENTIONAL 'CUSTOMER COMPLAINT'. THE INFORMATION REPORTED MAY OR MAY NOT BE RELATED TO THE EDWARDS DEVICE.
Patient problems: Dyspnea
MAUDE report 24328978 ↗InjuryFeb 6, 2026Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
IT WAS LEARNED THROUGH IMPLANT PATIENT REGISTRY AND INVESTIGATION THAT A PATIENT WITH A 23MM 3300TFX AORTIC VALVE WAS EXPLANTED AFTER AN IMPLANT DURATION OF (B)(6) DUE TO ENDOCARDITIS. THE EXPLANTED VALVE WAS REPLACED WITH A 23MM 11500A VALVE. PATIENT DISCHARGE ON POD#8. PER PATHOLOGY REPORT, AORTIC VALVE AND GRAFT WITH ENDOCARDITIS. LEAFLETS WITH FRIABLE VEGETATIONS.
Patient problems: Endocarditis; Insufficient Information
MAUDE report 24284757 ↗
Other replacement Heart-valve devices
- CARPENTIER-EDWARDS PERIMOUNT MAGNA EASE PERICARDIAL BIOPROSTHESIS - AORTIC 3300TFXEdwards Lifesciences LLC · replacement Heart-valve1 death656 reports
- CARPENTIER-EDWARDS PERIMOUNT MAGNA EASE PERICARDIAL BIOPROSTHESIS - AORTIC 3300TFXEdwards Lifesciences LLC · replacement Heart-valve5 deaths533 reports
- CARPENTIER-EDWARDS PERIMOUNT MAGNA MITRAL EASE PERICARDIAL BIOPROSTHESIS 7300TFXEdwards Lifesciences LLC · replacement Heart-valve1 death425 reports
- CARPENTIER-EDWARDS PERIMOUNT MAGNA MITRAL EASE PERICARDIAL BIOPROSTHESIS 7300TFXEdwards Lifesciences LLC · replacement Heart-valve1 death391 reports
- CARPENTIER-EDWARDS PERIMOUNT MAGNA MITRAL EASE PERICARDIAL BIOPROSTHESIS 7300TFXEdwards Lifesciences LLC · replacement Heart-valve1 death290 reports
- CARPENTIER-EDWARDS PERIMOUNT MAGNA EASE PERICARDIAL BIOPROSTHESIS - AORTIC 3300TFXEdwards Lifesciences LLC · replacement Heart-valve260 reports
- CARPENTIER-EDWARDS PERIMOUNT MAGNA MITRAL EASE PERICARDIAL BIOPROSTHESIS 7300TFXEdwards Lifesciences LLC · replacement Heart-valve2 deaths229 reports
- CARPENTIER-EDWARDS PERIMOUNT MAGNA MITRAL EASE PERICARDIAL BIOPROSTHESIS 7300TFXEdwards Lifesciences LLC · replacement Heart-valve2 deaths142 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page