CLEARSIGHT CSCSRecalls & reported problems

Edwards Lifesciences LLC · CLEARSIGHT FINGER CUFF SMALL MULTI PACK · Class 2

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00690103193343), received between May 12, 2025 and May 12, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K140312, K160552, K182245); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 12, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    AS REPORTED DURING USE THE SMALL CLEARSIGHT CUFF WAS PROVIDING INACCURATE VALUES. THE CLEARSIGHT CUFF HAD A DAMPED WAVEFORM AND WAS READING 60S OVER 30S WITH MEAN ARTERIAL PRESSURES OF 40 AND THE NONINVASIVE BLOOD PRESSURE ON THE SAME SIDE WAS READING 110 OVER 80S WITH MAP GREATER THAN 65. THE CLINICIAN REPORTED THAT HE HAD REMOVED THE CLEARSIGHT CUFF TO ADJUST IT BECAUSE THERE SEEMED TO BE A GAP BETWEEN IT AND THE FINGER IT WAS ON. THE PATIENT RECEIVED MEDICATIONS TO INCREASE THE BLOOD PRESSURE BASED ON THE CLEARSIGHT CUFF READINGS. THE CLINICAL SALES REPRESENTATIVE REP STOPPED MONITORING REMOVED THE SMALL CLEARSIGHT CUFF AND REMEASURED THE PATIENT. THE SIZER INDICATED THE PATIENT WAS A MEDIUM CUFF INSTEAD OF A SMALL. WHEN THE CORRECT SIZE CUFF WAS PLACED ON THE PATIENT THE BLOOD PRESSURE VALUES INCREASED TO 100 OVER 60S. THERE WAS NO PATIENT INJURY. THE DEVICE IS AVAILABLE FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22003598 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K140312, K160552, K182245. One submission often covers many models, so check the product and lot details in each record.

  • Z-1193-2019 · Mar 21, 2019 · Terminated
    EV1000 Clinical Platform, Models: EV1000A, EV1000NI, EV1000CS Product Usage: EV1000A Monitor is used to optimize fluid management through the use of minimally invasive monitoring
  • Z-1732-2015 · May 18, 2015 · Terminated
    ClearSight, Model No. EVHRS Heart Reference Sensor. The EV1000 Clinical Platform Nl and ClearSight finger cuffs noninvasively measures blood pressure and associated hemodynamic pa

Other PLETHYSMOGRAPH, IMPEDANCE devices

All PLETHYSMOGRAPH, IMPEDANCE devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.