EDWARDS CERTITUDE DELIVERY SYSTEM 9600SDS23Recalls & reported problems
Edwards Lifesciences LLC · EDWARDS CERTITUDE DELIVERY SYSTEM · Class 3 · Implant
All models of Edwards Certitude Delivery System (7)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00690103194142), received between Feb 24, 2023 and Jan 30, 2025 — classified as 2 injuries, 1 malfunction.
The most frequently reported specific device problem is “Burst Container or Vessel” (2 reports).
3 other recalls cover devices cleared under the same FDA submission (P140031); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Burst Container or Vessel2
- Difficult to Remove1
- Material Rupture1
Patient problems
- Insufficient Information1
- Low Blood Pressure/ Hypotension1
- Vascular Dissection1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryJan 30, 2025Burst Container or Vessel; Difficult to Remove; Burst Container or Vessel; Difficult to Remove
AS REPORTED BY THE EDWARDS FIELD CLINICAL SPECIALIST, DURING A TRANSCATHETER AORTIC VALVE REPLACEMENT PROCEDURE BY TRANSCAROTID APPROACH, A NATIVE BASILICA OF THE LEFT LEAFLET WAS PERFORMED WITH EFFACED SINUSES. THE RIGHT CORONARY WAS ALSO WIRED BECAUSE OF POTENTIAL RISK OF CORONARY OCCLUSION WITH AN EFFACED SINUS. BASILICA, CAROTID ACCESS, CORONARY ACCESS, CERTITUDE SHEATH PLACEMENT AND VALVE DEPLOYMENT WERE ALL ACHIEVED WITH SUCCESS. POST VALVE DEPLOYMENT ON TRANSESOPHAGEAL ECHO, THERE WAS TRIVIAL LEAK BELOW THE LEFT CUSP. ON ANGIOGRAM, THE LEAK WAS MUCH WORSE, AND THE DECISION WAS MADE TO POST DILATE WITH 1 ADDITIONAL CC OF VOLUME. WHEN THIS WAS PERFORMED, THE CERTITUDE BALLOON RUPTURED CIRCUMFERENTIALLY. UPON TRYING TO REMOVE THE BALLOON THROUGH THE CERTITUDE SHEATH, THE FLANGES GOT CAUGHT ON AND WOULD NOT ALLOW THE DELIVERY SYSTEM TO BE REMOVED. THE SURGEON REMOVED THE SHEATH WITH THE BALLOON FLANGES FLARED WHICH DISSECTED THE LEFT CAROTID ARTERY. THE PROCEDURE TURNED INTO A VASCULAR REPAIR WITH A GRAFT ADDED TO THE LEFT CAROTID ARTERY. THE PATIENT LEFT THE TABLE ALIVE AND STABLE.
Patient problems: Vascular Dissection
MAUDE report 21275887 ↗MalfunctionAug 1, 2023Material Rupture; Material Rupture
AS REPORTED BY AN EDWARDS LIFESCIENCES AFFILIATE, REGARDING IMPLANT OF A 23MM SAPIEN 3 VALVE IN THE AORTIC POSITION VIA CAROTID APPROACH. THE CERTITUDE DELIVERY SYSTEM BALLOON RUPTURED DURING DEPLOYMENT. THE IMPLANTERS BELIEVE A PIECE OF CALCIUM ON THE STJ CAUSED THE THE BALLOON TO RUPTURE. NO HARM TO THE PATIENT WAS NOTED AT THIS.
Patient problems: Insufficient Information
MAUDE report 17436701 ↗InjuryFeb 24, 2023Burst Container or Vessel; Burst Container or Vessel
AS REPORTED BY A FIELD CLINICAL SPECIALIST (FCS), DURING A CAROTID TAVR PROCEDURE WITH A 23MM SAPIEN 3 VALVE, DURING FULL DEPLOYMENT OF THE VALVE, THE 23 MM CERTITUDE DELIVERY DEVICE BALLOON RUPTURED. THE PATIENT HAS A HEAVILY CALCIFIED SINOTUBULAR JUNCTION (STJ) AND ASCENDING AORTA. THE BALLOON WAS FULLY INFLATED WHEN IT BURST. THE PATIENT REMAINED HYPOTENSIVE POST DEPLOYMENT. CPR WAS INITIATED AND EPI WAS GIVEN AND THE PATIENT FULLY RECOVERED. THE VALVE WAS POST DILATED WITH A 22MM ATLAS GOLD BALLOON. THE VALVE WAS POST DILATED BECAUSE THE PHYSICIAN WANTED TO MAKE SURE THE VALVE WAS FULLY EXPANDED SINCE THE DELIVERY BALLOON RUPTURED DURING THE END OF DEPLOYMENT. THE PHYSICIAN HAD TO USE SOME FORCE TO PULL THE DELIVERY OUT OF THE SHEATH BUT THERE WAS NO DAMAGE TO THE SHEATH AND THE BALLOON WAS INTACT. THE PATIENT WAS STABLE AT THE END OF THE CASE AND THERE WAS MILD PARAVALVULAR LEAK AND NO CENTRAL LEAK OBSERVED. PER THE FCS, THE PATIENT HAD A SEVERELY CALCIFIED STJ. THERE WAS NO EXTRA VOLUME ADDED TO THE BALLOON PRIOR TO INFLATION. THE PATIENT'S PRESSURE STARTED TO DROP PRIOR TO DEPLOYMENT, ONCE THE SAPIEN 3 VALVE CROSSED THE NATIVE VALVE. THE PERCEIVED ROOT CAUSE OF THE HYPOTENSI
Patient problems: Low Blood Pressure/ Hypotension
MAUDE report 16436972 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P140031. One submission often covers many models, so check the product and lot details in each record.
- Z-0835-2019 · Jan 22, 2019 · TerminatedEdward Lifesciences, Edwards Commander Delivery System: Model Number UDI Number 9600LDS20A (US) 00690103193930 9600LDS20J 00690103194852 9600LDS23A (US) 00690103193947 9600LDS
- Z-2435-2018 · Jun 11, 2018 · TerminatedEdwards SAPIEN 3 Transcatheter Heart Valve (THV), Model: 9600TFX, Size: 26 mm
- Z-0293-2018 · Jul 21, 2017 · TerminatedCertitude Delivery System and any associated Multipacks Models - 9600CT20A, 9600CT23A, 9600CT26A, 9600CT29A Product Usage: The Edwards Certitude Delivery System is used for delive
Other Prosthesis, Mitral Valve, Percutaneously Delivered devices
- EDWARDS SAPIEN 3 TRANSCATHETER HEART VALVE (THV) 9600TFXEdwards Lifesciences LLC · Prosthesis, Mitral Valve, Percutaneously Delivered59 deaths1,124 reports
- Edwards SAPIEN 3 Ultra Transcatheter Heart Valve 9750TFX26Edwards Lifesciences LLC · Prosthesis, Mitral Valve, Percutaneously Delivered64 deaths1,070 reports
- EDWARDS SAPIEN 3 TRANSCATHETER HEART VALVE (THV) 9600TFXEdwards Lifesciences LLC · Prosthesis, Mitral Valve, Percutaneously Delivered14 deaths975 reports
- Edwards SAPIEN 3 Ultra Transcatheter Heart Valve 9750TFX23Edwards Lifesciences LLC · Prosthesis, Mitral Valve, Percutaneously Delivered42 deaths936 reports
- EDWARDS SAPIEN 3 TRANSCATHETER HEART VALVE (THV) 9600TFXEdwards Lifesciences LLC · Prosthesis, Mitral Valve, Percutaneously Delivered1 recall22 deaths733 reports
- SAPIEN 3 Ultra System 9750CM26Edwards Lifesciences LLC · Prosthesis, Mitral Valve, Percutaneously Delivered26 deaths531 reports
- SAPIEN 3 Ultra RESILIA 9755RSL29Edwards Lifesciences LLC · Prosthesis, Mitral Valve, Percutaneously Delivered35 deaths321 reports
- SAPIEN 3 Ultra RESILIA 9755RSL26Edwards Lifesciences LLC · Prosthesis, Mitral Valve, Percutaneously Delivered18 deaths314 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page