EDWARDS CERTITUDE DELIVERY SYSTEM 9600SDS26Recalls & reported problems
Edwards Lifesciences LLC · EDWARDS CERTITUDE DELIVERY SYSTEM · Class 3 · Implant
All models of Edwards Certitude Delivery System (7)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00690103194159), received between Mar 18, 2021 and Apr 16, 2026 — classified as 4 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Burst Container or Vessel” (6 reports).
3 other recalls cover devices cleared under the same FDA submission (P140031); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Burst Container or Vessel6
- Difficult to Remove3
- Adverse Event Without Identified Device or Use Problem1
- Difficult to Insert1
- Material Rupture1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- Hemorrhage/Blood Loss/Bleeding1
- Unspecified Vascular Problem1
- Vascular Dissection1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 16, 2026Burst Container or Vessel
AS REPORTED BY THE FIELD CLINICAL SPECIALIST (FCS), DURING A LEFT-SIDE TRANS CAROTID TAVR (TRANSCATHETER AORTIC VALVE REPLACEMENT) PROCEDURE WITH A 26MM SAPIEN 3 VALVE IN THE NATIVE AORTIC POSITION, THE BALLOON TO THE 26MM CERTITUDE DELIVERY SYSTEM (DS) BURST WHILE BEING DEPLOYED. THE DS WAS NOT ABLE TO BE REMOVED THROUGH THE 18FR CERTITUDE SHEATH; THE DS AND SHEATH HAD TO BE REMOVED TOGETHER AS A UNIT. A SECOND AND NEW 18FR CERTITUDE SHEATH WAS THEN INSERTED INTO THE CAROTID ARTERY TO FINISH CLOSING THE ARTERIOTOMY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24900613 ↗InjuryNov 10, 2025Difficult to Insert; Difficult to Insert
AS REPORTED BY THE FIELD CLINICAL SPECIALIST (FCS), DURING A TRANSAORTIC TRANSCATHETER AORTIC VALVE REPLACEMENT (TAVR) PROCEDURE WITH A 26 MM SAPIEN 3 ULTRA RESILIA VALVE, THE 26 MM CERTITUDE DELIVERY SYSTEM AND 26 MM SAPIEN 3 ULTRA RESILIA VALVE WERE INSERTED INTO THE 18FR CERTITUDE SHEATH, BUT THE SYSTEM WOULD NOT ADVANCE THROUGH THE LOADER. THE DELIVERY SYSTEM AND VALVE WERE REMOVED. A SECOND 26 MM CERTITUDE DELIVERY SYSTEM AND 26 MM SAPIEN 3 ULTRA RESILIA VALVE WERE PREPARED. PRIOR TO INSERTING THE SECOND DELIVERY SYSTEM AND VALVE INTO THE CERTITUDE SHEATH, AN ATTEMPT WAS MADE TO ADVANCE THE DELIVERY SYSTEM AND VALVE THROUGH THE LOADER. THE DELIVERY SYSTEM AND VALVE WERE UNABLE TO ADVANCE THROUGH THE LOADER. NO ATTEMPT WAS MADE TO INSERT THE DELIVERY SYSTEM AND VALVE INTO THE CERTITUDE SHEATH. A DECISION WAS MADE TO USE A 26 MM COMMANDER DELIVERY SYSTEM. AT THIS POINT, THE PATIENT REQUIRED BLOOD TRANSFUSIONS DUE TO BLEEDING AROUND THE SHEATH SITE. IT WAS INDICATED THAT THE DIFFICULTY USING THE SAPIEN 3 ULTRA RESILIA VALVE WITH THE CERTITUDE DELIVERY SYSTEM CREATED A DELAY IN PROCEEDING WITH VALVE IMPLANTATION, WHICH ALLOWED FOR EXTRA BLEEDING TIME. A THIRD 26 MM SAPIEN 3 ULTRA
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 23513681 ↗MalfunctionMar 17, 2025Burst Container or Vessel; Difficult to Remove; Burst Container or Vessel; Difficult to Remove
AS REPORTED BY A FIELD CLINICAL SPECIALIST, A PATIENT UNDERWENT A TRANSCAROTID TAVR PROCEDURE WITH A 26MM SAPIEN 3 VALVE IN AORTIC POSITION. DURING THE PROCEDURE, THE BALLOON RUPTURED DURING DEPLOYMENT. THE VALVE WAS SUCCESSFULLY DEPLOYED. THE SHEATH AND CERTITUDE DELIVERY SYSTEM WERE REMOVED. PATIENT IS CURRENTLY IN RECOVERY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21617829 ↗MalfunctionSep 18, 2024Burst Container or Vessel; Difficult to Remove; Burst Container or Vessel; Difficult to Remove
AS REPORTED BY A FIELD CLINICAL SPECIALIST (FCS), DURING A TAVR PROCEDURE WITH A 26MM SAPIEN 3 VALVE IN THE AORTIC POSITION, THE 26MM CERTITUDE BALLOON RUPTURED DURING DEPLOYMENT. THE FCS ADVISED THAT THE RUPTURED WAS VERY DIFFICULT TO REMOVE. THE BALLOON AND SHEATH HAD TO BE REMOVED TOGETHER AND A NEW DELIVERY BALLOON AND SHEATH HAD TO BE INSERTED AND USED TO POST DILATE THE VALVE. THERE WERE NO ISSUES AT THE TIME OF THIS NOTE. UPON FOLLOW UP, THE FCS ADVISED THAT THE BALLOON WAS TORN. THE EVENT WAS AT THE LAST 1 OR 2CCS OF BALLOON INFLATION. AN ATTEMPT WAS MADE TO REMOVE THE DELIVERY SYSTEM THROUGH THE SHEATH, BUT THE BALLOON WINGS WOULD NOT ALLOW. THE ENTIRE SYSTEM, DELIVERY AND SHEATH WERE REMOVED TOGETHER. NO INSTRUMENTS WERE USED TO REMOVE THE BALLOON COVER. NO EXTRA VOLUME ADDED TO THE BALLOON PRIOR TO THE INFLATION. THE SPEED OF THE INFLATION TECHNIQUE WAS SLOW. THERE WAS NO LEAK IN THE DELIVERY SYSTEM NOTICED. BLOOD WAS SEEN IN THE SYRINGE WHEN NEGATIVE PRESSURE WAS PULLED IN THE INFLATION DEVICE. THERE WAS NO DIFFICULTY WITH THE VALVE ALIGNMENT. THERE WAS DIFFICULTY WITHDRAWING THE DELIVERY SYSTEM. THE WITHDRAWAL DIFFICULTY WAS EXPERIENCED AFTER THE BALLOON RUPTURED. THE
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20252013 ↗MalfunctionJul 11, 2024Burst Container or Vessel; Burst Container or Vessel
AS REPORTED BY AN EDWARDS LIFESCIENCES FIELD CLINICAL SPECIALIST, A CERTITUDE DELIVERY SYSTEM BALLOON BURST DURING INFLATION/DEPLOYMENT OF A 26MM SAPIEN 3 ULTRA RESILIA VALVE IN THE AORTIC POSITION BY CAROTID APPROACH. BALLOON INFLATION TECHNIQUE WAS FAST, AND AN ADDITIONAL 2CC OF VOLUME WAS ADDED TO THE BALLOON PRIOR TO INFLATION. SINOTUBULAR JUNCTION CALCIUM IS SUSPECTED AS THE CAUSE OF THE BALLOON BURST. THE VALVE WAS ASSESSED BY PHYSICIANS AND DETERMINED TO BE FULLY EXPANDED. THE COMMANDER DELIVERY SYSTEM WAS WITHDRAWN WITHOUT ISSUE. PATIENT TOLERATED THE PROCEDURE WELL, AND NO PATIENT COMPLICATIONS WERE NOTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19726681 ↗InjuryNov 22, 2023Burst Container or Vessel; Difficult to Remove; Burst Container or Vessel; Difficult to Remove
AS REPORTED BY A FIELD CLINICAL SPECIALIST, A PATIENT UNDERWENT A TRANSCAROTID PROCEDURE WITH A 26MM SAPIEN 3 VALVE INSIDE A PRE-EXISTING 27MM FREESTYLE VALVE IN AORTIC POSITION. THIS WAS A PLANNED LEFT CAROTID ACCESS CASE USING A 26MM CERTITUDE DELIVERY SYSTEM AND SHEATH SYSTEM. THE SHEATH, VALVE, AND DELIVERY SYSTEM WERE PREPPED PER IFU. THE 26MM SAPIEN 3 VALVE WAS DEPLOYED WITH 24CC PER PHYSICIANS REQUEST. DURING BALLOON INFLATION, THE BALLOON HAD RUPTURED. THE VALVE WAS DEPLOYED AT 60/40 POSITION. THE SURGEON HAD DIFFICULTY PULLING THE BALLOON BACK INTO THE SHEATH. AS A RESULT, THE CAROTID ARTERY WAS TORN AND NEEDED SURGICAL REPAIR. THE PATIENT REMAINED STABLE DURING THE PROCEDURE. AS PER FOLLOW-UP WITH THE FCS, THE VALVE WAS DEPLOYED AT 60/40 INTENTIONALLY DUE TO CALCIUM. THE VALVE WAS POST-DILATED BECAUSE THE VALVE WAS UNDERDEPLOYED NOT DUE TO PVL. THE BALLOON WAS PARTIALLY INFLATED WHEN IT BURST. THERE WAS 1CC EXTRA VOLUME ADDED TO THE BALLOON PRIOR TO THE INFLATION. THE INFLATION TECHNIQUE USED WAS SLOW AND 24CC OF VOLUME WAS USED FOR INFLATION. THE PERCEIVED ROOT CAUSE OF THE BALLOON BURST IS DUE TO THE CALCIUM. THE DEVICE WAS GETTING CAUGHT AT THE DISTAL TIP DURING WITHDR
Patient problems: Vascular Dissection
MAUDE report 18196731 ↗InjuryJul 12, 2022Burst Container or Vessel; Burst Container or Vessel
A CERTITUDE DELIVERY SYSTEM BALLOON RUPTURED DURING THE DEPLOYMENT OF A SAPIEN 3 VALVE IN THE AORTIC POSITION. AS REPORTED BY AN EDWARDS FIELD CLINICAL SPECIALIST, DURING A TRANSCAROTID TAVR PROCEDURE, ACCESS TO THE CAROTID AND VESSEL DILATION WERE COMPLETED WITHOUT ISSUE. POST DEPLOYMENT OF A 26MM SAPIEN 3 VALVE, BLOOD WAS OBSERVED IN THE INFLATION DEVICE WHEN PULLED NEGATIVE, INDICATING AN APPARENT BALLOON RUPTURE. DIFFICULTIES WERE ENCOUNTERED DURING THE WITHDRAWAL OF THE DELIVERY SYSTEM AND THE DELIVERY SYSTEM WAS NOT ABLE TO BE PULLED BACK INTO THE CERTITUDE SHEATH. AN ATTEMPT WAS MADE TO REMOVE THE DELIVERY SYSTEM AND SHEATH TOGETHER. THE SHEATH WAS REMOVED, HOWEVER THE DISTAL BALLOON SHOULDER DETACHED FROM DELIVERY SYSTEM AND WAS WEDGED AT THE CAROTID OSTIUM. A CUT-DOWN WAS DONE TO EXPOSE THE VESSEL AS CLOSE TO THE OSTIUM AS POSSIBLE AND ATTEMPTS WERE MADE TO GRAB DISTAL SHOULDER/NOSE CONE/ AND REMOVE THE BALLOON FROM THE OSTIUM. THE DISTAL BALLOON WAS GRABBED AND REMOVED. THE CAROTID WAS LIGATED AND REPAIRED. THE PATIENT WAS TRANSFERRED TO POST-OP IN STABLE CONDITION. THE VALVE REMAINS IMPLANTED IN THE PATIENT.
Patient problems: Unspecified Vascular Problem
MAUDE report 14987160 ↗InjuryMar 29, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
AFTER REVIEWING THE MEDICAL RECORDS PROVIDED, IT IS NOTED THAT THE PATIENT HAD A VIV (26MM S3 IN A SURGICAL VALVE) IN THE TRICUSPID POSITION VIA TRANSJUGULAR AND RIGHT FEMORAL VEIN APPROACH. A 26MM SAPIEN 3 VALVE WAS IMPLANTED USING A CERTITUDE DELIVERY SYSTEM THROUGH A 18FR CERTITUDE SHEATH. POST PROCEDURE, THE TEAM REALIZED THE PICC LINE WAS NOT FUNCTIONING. FLUOROSCOPY REVEALED THAT IT WAS CURLED IN THE NECK. THEY DID NOT ATTEMPT TO REPOSITION IT. INSTEAD, A 5F SHEATH WAS PLACED IN THE RIGHT FEMORAL VEIN WITH ULTRASOUND GUIDANCE. THIS WAS TO BE USED UNTIL THE PICC LINE COULD BE CORRECTED. FOLLOW UP CONFIRMED THIS WAS THE SAME SIDE AS THE EDWARDS DEVICES, AND THE ISSUES WITH THE PICC LINE RELATED TO AN INTERACTION WITH THE EDWARDS DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13934749 ↗MalfunctionMar 18, 2021Material Rupture; Material Rupture
AS REPORTED BY AN EDWARDS FIELD CLINICAL SPECIALIST, DURING DEPLOYMENT OF A 26MM SAPIEN 3 VALVE IN THE AORTIC POSITION, THE CERTITUDE DELIVERY SYSTEM BALLOON BURST. THE DELIVERY SYSTEM WAS ABLE TO BE REMOVED WITH THE ESHEATH WITHOUT ISSUE AND THERE WERE NO PATIENT INJURIES. THE PATIENT HAD MILD PVL ON ANGIO AND TRACE PVL ON TEE. A POST DILATION WAS NOT PERFORMED. PATIENT LEFT THE ROOM IN STABLE CONDITION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11516230 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P140031. One submission often covers many models, so check the product and lot details in each record.
- Z-0835-2019 · Jan 22, 2019 · TerminatedEdward Lifesciences, Edwards Commander Delivery System: Model Number UDI Number 9600LDS20A (US) 00690103193930 9600LDS20J 00690103194852 9600LDS23A (US) 00690103193947 9600LDS
- Z-2435-2018 · Jun 11, 2018 · TerminatedEdwards SAPIEN 3 Transcatheter Heart Valve (THV), Model: 9600TFX, Size: 26 mm
- Z-0293-2018 · Jul 21, 2017 · TerminatedCertitude Delivery System and any associated Multipacks Models - 9600CT20A, 9600CT23A, 9600CT26A, 9600CT29A Product Usage: The Edwards Certitude Delivery System is used for delive
Other Prosthesis, Mitral Valve, Percutaneously Delivered devices
- EDWARDS SAPIEN 3 TRANSCATHETER HEART VALVE (THV) 9600TFXEdwards Lifesciences LLC · Prosthesis, Mitral Valve, Percutaneously Delivered59 deaths1,124 reports
- Edwards SAPIEN 3 Ultra Transcatheter Heart Valve 9750TFX26Edwards Lifesciences LLC · Prosthesis, Mitral Valve, Percutaneously Delivered64 deaths1,070 reports
- EDWARDS SAPIEN 3 TRANSCATHETER HEART VALVE (THV) 9600TFXEdwards Lifesciences LLC · Prosthesis, Mitral Valve, Percutaneously Delivered14 deaths975 reports
- Edwards SAPIEN 3 Ultra Transcatheter Heart Valve 9750TFX23Edwards Lifesciences LLC · Prosthesis, Mitral Valve, Percutaneously Delivered42 deaths936 reports
- EDWARDS SAPIEN 3 TRANSCATHETER HEART VALVE (THV) 9600TFXEdwards Lifesciences LLC · Prosthesis, Mitral Valve, Percutaneously Delivered1 recall22 deaths733 reports
- SAPIEN 3 Ultra System 9750CM26Edwards Lifesciences LLC · Prosthesis, Mitral Valve, Percutaneously Delivered26 deaths531 reports
- SAPIEN 3 Ultra RESILIA 9755RSL29Edwards Lifesciences LLC · Prosthesis, Mitral Valve, Percutaneously Delivered35 deaths321 reports
- SAPIEN 3 Ultra RESILIA 9755RSL26Edwards Lifesciences LLC · Prosthesis, Mitral Valve, Percutaneously Delivered18 deaths314 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page