EDWARDS CERTITUDE INTRODUCER SHEATH SET 9600IS18Recalls & reported problems
Edwards Lifesciences LLC · EDWARDS CERTITUDE INTRODUCER SHEATH SET · Class 3
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00690103194173), received between May 6, 2021 and May 9, 2023 — classified as 2 injuries.
The most frequently reported specific device problem is “Difficult to Insert” (2 reports).
3 other recalls cover devices cleared under the same FDA submission (P140031); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Difficult to Insert2
- Difficult to Advance1
- Patient-Device Incompatibility1
Patient problems
- Perforation of Vessels1
- Vascular Dissection1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMay 9, 2023Difficult to Insert; Patient-Device Incompatibility; Difficult to Advance; Difficult to Insert; Patient-Device Incompatibility; Difficult to Advance
AS REPORTED BY AN EDWARDS LIFESCIENCES FIELD CLINICAL SPECIALIST, DIFFICULTY INSERTING THE SHEATH INTO THE PATIENT AND A DISSECTION IN THE CAROTID ARTERY OCCURRED. A PLANNED VASCULAR CUTDOWN OF LEFT CAROTID ARTERY WAS DONE FOR ACCESS. A 18FR CERTITUDE SHEATH EXPERIENCED MORE RESISTANCE THAN USUAL DURING INSERTION. A 29 MM SAPIEN 3 VALVE WAS SUCCESSFULLY DEPLOYED IN THE AORTIC ANNULUS. AS THE SHEATH WAS REMOVED AT THE END OF THE PROCEDURE, A DISSECTION WAS FOUND IN THE CAROTID ARTERY. THE CAROTID ARTERY WAS SURGICALLY REPAIRED WITH A GRAFT. ANGIOGRAPHY AFTER THE SURGICAL REPAIR SHOWED A DISSECTION PROXIMAL TO THE GRAFT AND A CLAMP TISSUE INJURY DISTAL TO THE GRAFT. A COVERED STENT WAS THEN PLACED TO COVER BOTH PROXIMAL AND DISTAL DISEASED AREAS OF CAROTID AND GRAFT SEGMENT. FINAL ANGIOGRAPHY REVEALED BRISK CAROTID FLOW. THE PATIENT IS RECOVERING WELL AND WAS DISCHARGED ON POST OPERATIVE DAY 4. THE FIELD CLINICAL SPECIALIST REPORTED THAT THE PERCEIVED ROOT CAUSE OF THE INSERTION DIFFICULTY WAS CAROTID ARTERY LUMEN SIZE AND CAROTID ARTERY DISEASE.
Patient problems: Vascular Dissection
MAUDE report 16896194 ↗InjuryMay 6, 2021Difficult to Insert; Difficult to Insert
AS FOUND REPORTED BY THE FIELD CLINICAL SPECIALIST, POST VALVE DEPLOYMENT, BLEEDING WAS NOTED AT THE CAROTID INSERTION SITE. HEMOSTASIS ACHIEVED BUT ONCE THE CERTITUDE SHEATH WAS REMOVED, THE BLEEDING PERSISTED. MULTIPLE SUTURES WERE NEEDED TO ACHIEVE HEMOSTASIS. THE PHYSICIAN BELIEVED THE TRANSITION OF DILATOR/SHEATH GOT CAUGHT ON SOME CALCIFICATION AND MAY HAVE TORN THE CAROTID ARTERY. THERE WAS SOME DIFFICULTY INSERTING THE SHEATH, HOWEVER THE PHYSICIANS FELT CALCIFICATION INTERACTING WITH THE ESHEATH ON INSERTION MAY BE THE CAUSE OF THE INJURY.
Patient problems: Perforation of Vessels
MAUDE report 11788475 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P140031. One submission often covers many models, so check the product and lot details in each record.
- Z-0835-2019 · Jan 22, 2019 · TerminatedEdward Lifesciences, Edwards Commander Delivery System: Model Number UDI Number 9600LDS20A (US) 00690103193930 9600LDS20J 00690103194852 9600LDS23A (US) 00690103193947 9600LDS
- Z-2435-2018 · Jun 11, 2018 · TerminatedEdwards SAPIEN 3 Transcatheter Heart Valve (THV), Model: 9600TFX, Size: 26 mm
- Z-0293-2018 · Jul 21, 2017 · TerminatedCertitude Delivery System and any associated Multipacks Models - 9600CT20A, 9600CT23A, 9600CT26A, 9600CT29A Product Usage: The Edwards Certitude Delivery System is used for delive
Other Prosthesis, Mitral Valve, Percutaneously Delivered devices
- EDWARDS SAPIEN 3 TRANSCATHETER HEART VALVE (THV) 9600TFXEdwards Lifesciences LLC · Prosthesis, Mitral Valve, Percutaneously Delivered59 deaths1,124 reports
- Edwards SAPIEN 3 Ultra Transcatheter Heart Valve 9750TFX26Edwards Lifesciences LLC · Prosthesis, Mitral Valve, Percutaneously Delivered64 deaths1,070 reports
- EDWARDS SAPIEN 3 TRANSCATHETER HEART VALVE (THV) 9600TFXEdwards Lifesciences LLC · Prosthesis, Mitral Valve, Percutaneously Delivered14 deaths975 reports
- Edwards SAPIEN 3 Ultra Transcatheter Heart Valve 9750TFX23Edwards Lifesciences LLC · Prosthesis, Mitral Valve, Percutaneously Delivered42 deaths936 reports
- EDWARDS SAPIEN 3 TRANSCATHETER HEART VALVE (THV) 9600TFXEdwards Lifesciences LLC · Prosthesis, Mitral Valve, Percutaneously Delivered1 recall22 deaths733 reports
- SAPIEN 3 Ultra System 9750CM26Edwards Lifesciences LLC · Prosthesis, Mitral Valve, Percutaneously Delivered26 deaths531 reports
- SAPIEN 3 Ultra RESILIA 9755RSL29Edwards Lifesciences LLC · Prosthesis, Mitral Valve, Percutaneously Delivered35 deaths321 reports
- SAPIEN 3 Ultra RESILIA 9755RSL26Edwards Lifesciences LLC · Prosthesis, Mitral Valve, Percutaneously Delivered18 deaths314 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page