Skip to content

SWAN-GANZ PACEPORT 780F75MRecalls & reported problems

Edwards Lifesciences LLC · SWAN-GANZ PACEPORT OXIMETRY TD CATHETER · Class 2

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00690103197341), received between Oct 11, 2022 and Oct 11, 2022 — classified as 1 malfunction.

The most frequently reported specific device problem is “Material Rupture” (1 report).

4 other recalls cover devices cleared under the same FDA submission (K160084); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Most reported problems

Device problems

  • Material Rupture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 11, 2022Material Rupture; Material Rupture

    IT WAS REPORTED THAT, BEFORE USE, THE BALLOON ON THE SWAN-GANZ CATHETER WAS RUPTURED. THE DEVICE WAS BEING USED WITH A HEMOSPHERE MONITOR. THERE WAS NO PATIENT INJURY. THE PRODUCT WAS RETURNED FOR EVALUATION AND THE CENTRAL AREA OF THE BALLOON LATEX WAS MISSING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15577660 ↗

FDA clearance and classification

Cleared through 510(k) K160084 on May 3, 2016 (decision: Substantially Equivalent). Submitted by Edwards Lifesciences, LLC as “Swan Ganz Catheters”. The FDA review took 109 days. FDA record ↗

The FDA regulates product code KRA (catheter, continuous flush) under 21 CFR 870.1210 as a Class II device. 21 CFR 870.1210 ↗

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Recalls under the same FDA submission

These recalls involve devices cleared under K160084. One submission often covers many models, so check the product and lot details in each record.

  • Z-0765-2019 · Dec 21, 2018 · Terminated
    Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF 151F7 Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly determi
  • Z-0764-2019 · Dec 21, 2018 · Terminated
    Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF 131VF7P Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly deter
  • Z-0762-2019 · Dec 21, 2018 · Terminated
    Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF131F 7J Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians to rapidly determ
  • Z-0763-2019 · Dec 21, 2018 · Terminated
    Edwards Lifesciences Swan Ganz Thermodilution Catheter, REF 131F7P (5 pack of 131F7) Product Usage: The Swan-Ganz Thermodilution Catheter provides a diagnostic tool for physicians

Common questions about the SWAN-GANZ PACEPORT 780F75M

Has the SWAN-GANZ PACEPORT 780F75M been recalled?

No FDA recall record in our data names this exact device identifier (DI 00690103197341), based on the FDA recall file dated Nov 27, 2026. However, 4 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.

What problems have been reported with the SWAN-GANZ PACEPORT 780F75M?

The FDA's MAUDE database lists 1 adverse event report for this device, received on Oct 11, 2022. That report lists device problem “Material Rupture” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the SWAN-GANZ PACEPORT 780F75M?

None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the SWAN-GANZ PACEPORT 780F75M FDA approved?

It was cleared through 510(k) K160084 on May 3, 2016. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.

Is the SWAN-GANZ PACEPORT 780F75M MRI safe?

According to the device's FDA GUDID record, its labeling lists it as MR Unsafe: it should not be taken into an MRI scanner room.

Who makes the SWAN-GANZ PACEPORT 780F75M, and is it still sold?

The FDA's device database lists Edwards Lifesciences LLC as the company responsible for it (brand SWAN-GANZ PACEPORT, model 780F75M). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other Catheter, continuous flush devices

All Catheter, continuous flush devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page