HEMOSPHERE HEMPSC100Recalls & reported problems

Edwards Lifesciences LLC · Pressure Cable · Class 2

All models of HemoSphere (6)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00690103199321), received between Feb 9, 2024 and Apr 3, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Incorrect Measurement” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect Measurement2
  • Incorrect, Inadequate or Imprecise Result or Readings2

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 3, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT DURING USE THE HEMOSPHERE CABLE PROVIDED INACCURATE VALUES. THE SYSTOLIC BLOOD PRESSURE WAS READING 15 TO 20 POINTS DIFFERENCE FROM THE HEMOSPHERE TO THE ANESTHESIA MONITOR. ONCE THE SUSPECT CABLE WAS EXCHANGED FOR A NEW ONE THE READINGS DISPLAYED AS EXPECTED. THERE WAS NO TREATMENT ADMINISTERED FOR THE INACCURATE VALUES AND NO ALLEGATION OF IS NO PATIENT INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21764522 ↗
  • MalfunctionJul 25, 2024Incorrect Measurement; Incorrect Measurement

    AS REPORTED, WITH THIS HEMOSPHERE PRESSURE CABLE, RANDOM AND UNSTABLE VALUES WERE PROVIDED (MEDWATCH # (B)(4) ). THE SAME CABLE HAD PRESENTED THE SAME ISSUE WITH ANOTHER PATIENT AND DIFFERENT NURSE (MEDWATCH # (B)(4) ). NO FURTHER INFORMATION COULD BE OBTAINED. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS AVAILABLE FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19826310 ↗
  • MalfunctionJul 25, 2024Incorrect Measurement; Incorrect Measurement

    AS REPORTED, WITH THIS HEMOSPHERE PRESSURE CABLE, RANDOM AND UNSTABLE VALUES WERE PROVIDED (MEDWATCH #50624). THE SAME CABLE HAD PRESENTED THE SAME ISSUE WITH ANOTHER PATIENT AND DIFFERENT NURSE (MEDWATCH #50621). NO FURTHER INFORMATION COULD BE OBTAINED. THERE WAS NO ALLEGATION OF PATIENT INJURY. THE DEVICE WAS AVAILABLE FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19826240 ↗
  • MalfunctionFeb 9, 2024Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    IT WAS REPORTED THAT THERE WERE INACCURATE BLOOD PRESSURE READINGS DURING PATIENT USE. THE PSC100 PRESSURE CABLE WAS CONNECTED TO THE HEMOSPHERE MONITOR. THE READING ON THE HEMOSPHERE MONITOR THAT WAS SENT TO THE PHILLIPS MONITOR DID NOT MATCH. THE NURSE NOTICED THE DIFFERENCE IMMEDIATELY. THERE WERE NO ERROR MESSAGES DISPLAYED. THE DISPLAYED NUMBERS AND WHAT THEY WERE EXPECTED TO BE ARE UNKNOWN. A NEW CABLE WAS EXCHANGED FOR THE SUSPECT CABLE AND IT WAS CONNECTED AND IT RESOLVED THE ISSUE. THERE IS NO PATIENT INJURY. THERE WAS NO INAPPROPRIATE PATIENT TREATMENT ADMINISTERED. THE PATIENT DEMOGRAPHICS WERE REQUESTED AND WERE NOT PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18673892 ↗

Other Adjunctive Predictive Cardiovascular Indicator devices

All Adjunctive Predictive Cardiovascular Indicator devices →