Edwards Certitude Delivery System 9630TA23Recalls & reported problems

Edwards Lifesciences LLC · Edwards Certitude Delivery System · Class 3

All models of Edwards Certitude Delivery System (7)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00690103201185), received between Aug 11, 2021 and Mar 26, 2026 — classified as 2 injuries, 3 malfunctions.

The most frequently reported specific device problem is “Burst Container or Vessel” (5 reports).

3 other recalls cover devices cleared under the same FDA submission (P140031); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Burst Container or Vessel5
  • Difficult to Remove2
  • Inflation Problem1
  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions4
  • Insufficient Information1
  • Unspecified Vascular Problem1
  • Vascular Dissection1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 26, 2026Burst Container or Vessel; Inflation Problem

    AS REPORTED BY THE FIELD CLINICAL SPECIALIST (FCS), THE PATIENT WAS UNDERGOING A TRANSCAROTID VALVE-IN-VALVE TRANSCATHETER AORTIC VALVE REPLACEMENT (TAVR) PROCEDURE USING A 23 MM SAPIEN 3 ULTRA RESILIA VALVE. THE VALVE WAS BEING IMPLANTED INSIDE A NON-EDWARDS VALVE. DURING THE PROCEDURE, THERE WAS RESISTANCE WHEN ADVANCING THE CERTITUDE DELIVERY SYSTEM WITH VALVE THROUGH THE 21FR CERTITUDE SHEATH, BUT THE DELIVERY SYSTEM WITH VALVE WAS ABLE TO EXIT THE TIP OF THE SHEATH. AFTER EXITING THE TIP OF THE SHEATH, THERE WAS DIFFICULTY ADVANCING THE DELIVERY SYSTEM THROUGH THE STRUTS OF THE NON-EDWARDS VALVE. IT WAS NOTED THAT THE IMPLANTING TEAM WAS BEHIND THE STRUTS OF THE NON-EDWARDS VALVE. THE DELIVERY SYSTEM CROSSED THE STRUTS AT THE MID VALVE PART OF THE VALVE AND WAS ON THE OUTSIDE OF THE NON-EDWARDS VALVE. THE SHEATH WAS PULLED BACK AND THEN THE SHEATH WAS USED TO PROVIDE MORE SUPPORT WHEN IT WAS NOTICED THAT THE TIP OF THE CERTITUDE SHEATH HAD "BROKE OFF". THE TIP OF THE SHEATH WAS NOTED TO BE STUCK AROUND THE SAPIEN 3 ULTRA RESILIA VALVE. THE VALVE WAS NOT DEPLOYED. THE DELIVERY SYSTEM WITH VALVE AND SHEATH WERE REMOVED AS A SINGLE UNIT. A NEW 21FR CERTITUDE SHEATH AND 23 MM CERT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24703096 ↗
  • MalfunctionMay 5, 2025Burst Container or Vessel; Burst Container or Vessel

    AS REPORTED BY THE EDWARDS FIELD CLINICAL SPECIALIST, DURING A TRANSCATHETER AORTIC VALVE REPLACEMENT PROCEDURE BY TRANSAORTIC APPROACH, THE 23 MM CERTITUDE DELIVERY SYSTEM BALLOON RUPTURED WITH A LONGITUDINAL TEAR AT THE END OF DEPLOYMENT. A 23 MM SAPIEN 3 ULTRA VALVE WAS SUCCESSFULLY IMPLANTED. THE DELIVERY SYSTEM WAS REMOVED WITHOUT INCIDENT. CAUSE WAS UNKNOWN. THE DELIVERY SYSTEM WAS DISCARDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21965181 ↗
  • InjuryJan 17, 2025Burst Container or Vessel; Difficult to Remove; Burst Container or Vessel; Difficult to Remove

    PER REPORT, A VASCULAR SURGEON WAS CALLED TO PATCH A DISSECTION AT THE ACCESS SITE. THE PATIENT REMAINED STABLE.

    Patient problems: Insufficient Information; Vascular Dissection

    MAUDE report 21183170 ↗
  • InjuryMar 16, 2023Burst Container or Vessel; Material Fragmentation; Difficult to Remove; Burst Container or Vessel; Material Fragmentation; Difficult to Remove

    PER PHOTO PROVIDED, IT APPEARED THE BALLOON WAS SEPARATED. PER THE FCS, ALL BALLOON MATERIAL WAS ACCOUNTED FOR UPON REMOVAL. THE SEPARATION OF THE BALLOON WAS BELIEVED TO BE CAUSED BY THE WITHDRAWAL DIFFICULTY.

    Patient problems: No Clinical Signs, Symptoms or Conditions; Unspecified Vascular Problem

    MAUDE report 16557660 ↗
  • MalfunctionAug 11, 2021Burst Container or Vessel; Burst Container or Vessel

    AS REPORTED BY THE FIELD CLINICAL SPECIALIST (FCS), DURING A TRANSAORTIC TAVR PROCEDURE FOR A 23MM SAPIEN 3 ULTRA VALVE, THE CERTITUDE DELIVERY SYSTEM BALLOON BURST DURING INFLATION. THE VALVE APPEARED FULLY DEPLOYED AND THE DELIVERY SYSTEM WAS REMOVED WITHOUT ISSUE. THE SINOTUBULAR JUNCTION (STJ) HAD CALCIUM PRESENT. THE PATIENT LEFT ROOM IN STABLE CONDITION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12311137 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P140031. One submission often covers many models, so check the product and lot details in each record.

  • Z-0835-2019 · Jan 22, 2019 · Terminated
    Edward Lifesciences, Edwards Commander Delivery System: Model Number UDI Number 9600LDS20A (US) 00690103193930 9600LDS20J 00690103194852 9600LDS23A (US) 00690103193947 9600LDS23J 0
  • Z-2435-2018 · Jun 11, 2018 · Terminated
    Edwards SAPIEN 3 Transcatheter Heart Valve (THV), Model: 9600TFX, Size: 26 mm
  • Z-0293-2018 · Jul 21, 2017 · Terminated
    Certitude Delivery System and any associated Multipacks Models - 9600CT20A, 9600CT23A, 9600CT26A, 9600CT29A Product Usage: The Edwards Certitude Delivery System is used for delive

Other Prosthesis, Mitral Valve, Percutaneously Delivered devices

All Prosthesis, Mitral Valve, Percutaneously Delivered devices →