Cardiovascular Procedure Kit 75533Recalls & reported problems

Terumo Cardiovascular Systems Corporation

All models of Cardiovascular Procedure Kit (227)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00699753498990), received between Jun 21, 2018 and Jun 21, 2018 — classified as 1 malfunction.

The most frequently reported specific device problem is “Connection Problem” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Feb 20, 2024Open, Classified

    Pressure relief valve included in certain lots of cardiovascular procedure kits was reportedly opening at approximately 200-300mmHg versus 400mmHg, potential myocardial tissue damage potentially requiring surgical or medical/pharmacological intervention

    Z-1401-2024 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Connection Problem1
  • Crack1
  • Fluid/Blood Leak1

Patient problems

  • Blood Loss1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 21, 2018Crack; Fluid/Blood Leak; Connection Problem; Crack; Fluid/Blood Leak; Connection Problem

    THE CUSTOMER REPORTED THAT DURING CARDIOPULMONARY BYPASS (CPB) PROCEDURE, THE LARGE LUER ON END OF LINE (17B, L2) WAS A CRACKED AND LEAKING. IT RESULTED IN APPROXIMATELY 5 CC BLOOD LOSS. THE DEVICE WAS CHANGED OUT. THE SURGICAL PROCEDURE WAS COMPLETED SUCCESSFULLY BUT THERE WAS A 5 MINUTES DELAY OCCURRED.

    Patient problems: Blood Loss; No Known Impact Or Consequence To Patient

    MAUDE report 7624700 ↗