Cardiovascular Procedure Kit 75179Recalls & reported problems
Terumo Cardiovascular Systems Corporation
All models of Cardiovascular Procedure Kit (227)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00699753500730), received between Aug 15, 2025 and Aug 19, 2025 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Device Misassembled During Manufacturing /Shipping” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Device Misassembled During Manufacturing /Shipping2
- Infusion or Flow Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 19, 2025Device Misassembled During Manufacturing /Shipping; Infusion or Flow Problem; Device Misassembled During Manufacturing /Shipping; Infusion or Flow Problem
PER CLINICAL REVIEW: THE TEAM HAD AN ISSUE WITH THEIR TUBING PACK AND THE CARDIOPLEGIA DELIVERY CIRCUIT WITHIN IT THAT HAD THE WRONG DIAMETER TUBING SIZES. THIS WAS DISCOVERED DURING CARDIOPULMONARY BYPASS (CPB). THE CARDIOPLEGIA CIRCUIT TUBING WAS NOT CHANGED OUT. THERE WAS NO HARM TO THE PATIENT, AND THE SURGERY WAS SUCCESSFULLY COMPLETED. THE RESULT OF THIS INCORRECT ASSEMBLY IS THAT THE INCORRECT VOLUME AND CONCENTRATION OF CARDIOPLEGIA WAS DELIVERED TO THE PATIENT. THIS OCCURRED BECAUSE THE CUSTOMER WAS NOT AWARE OF THE TUBING DIAMETER CHANGE AND DID NOT CHANGE AND DID NOT CHANGE THE FLOW CONSTANT FACTOR ON THE ROLLER PUMP. THE CARDIOPLEGIA CIRCUIT TUBING WAS NOT RETURNED TO THE MANUFACTURER FOR FURTHER INVESTIGATION AND NO PICTURES WERE TAKEN OR SHARED BY THE USER FACILITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22846343 ↗MalfunctionAug 15, 2025Device Misassembled During Manufacturing /Shipping; Infusion or Flow Problem; Device Misassembled During Manufacturing /Shipping; Infusion or Flow Problem
PER CLINICAL REVIEW: THE TEAM EXPERIENCED AN ISSUE WITH THEIR TUBING PACK AND THE CARDIOPLEGIA DELIVERY CIRCUIT WITHIN IT THAT HAD THE WRONG DIAMETER TUBING SIZES. IT WAS DISCOVERED DURING CARDIOPULMONARY BYPASS (CPB). THE CARDIOPLEGIA CIRCUIT TUBING WAS NOT CHANGED OUT. THERE WAS NO HARM TO THE PATIENT, AND THE SURGERY WAS SUCCESSFULLY COMPLETED. THE RESULT OF THIS INCORRECT ASSEMBLY IS THAT THE INCORRECT VOLUME AND CONCENTRATION OF CARDIOPLEGIA WAS DELIVERED TO THE PATIENT. THIS OCCURRED BECAUSE THE CUSTOMER WAS NOT AWARE OF THE TUBING DIAMETER CHANGE AND DID NOT CHANGE THE FLOW CONSTANT FACTOR ON THE ROLLER PUMP. THE CARDIOPLEGIA CIRCUIT TUBING WAS NOT RETURNED TO THE MANUFACTURER FOR FURTHER INVESTIGATION AND NO PICTURES WERE TAKEN OR SHARED BY THE USER FACILITY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22816159 ↗
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026