Cardiovascular Procedure Kit 75106-02Recalls & reported problems
Terumo Cardiovascular Systems Corporation
All models of Cardiovascular Procedure Kit (227)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00699753504677), received between Jan 25, 2019 and Jan 25, 2019 — classified as 1 injury.
The most frequently reported specific device problem is “Disconnection” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Disconnection1
Patient problems
- Blood Loss1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJan 25, 2019Disconnection; Disconnection
IT WAS REPORTED THAT DURING THE CARDIOPULMONARY BYPASS (CPB) PROCEDURE, THE TUBING CONNECTOR FROM THE TERUMO CARDIOVASCULAR PROCEDURE KIT AND THE TUBING FROM ANOTHER MANUFACTURER'S KIT DISCONNECTED, RESULTING IN APPROXIMATELY 2000-2200 ML BLOOD LOSS. THIS WAS A USER MADE CONNECTION THAT HAD NOT BEEN TIE BANDED. THERE WAS A DELAY OF APPROXIMATELY 3 MINUTES WHILE THE CONNECTOR AND TUBING WERE RECONNECTED. THE USER INDICATED THERE WAS NO BLOOD TRANSFUSION REQUIRED. THE SURGERY WAS SUCCESSFULLY COMPLETED. IT WAS ALSO REPORTED THAT THERE NO SERIOUS ADVERSE PATIENT IMPACT OR POST-SURGICAL COMPLICATIONS.
Patient problems: Blood Loss; No Consequences Or Impact To Patient
MAUDE report 8278861 ↗
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page