Skip to content

Cardiovascular Procedure Kit 76415Recalls & reported problems

Terumo Cardiovascular Systems Corporation

All models of Cardiovascular Procedure Kit (227)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00699753507227), received between Sep 27, 2018 and Sep 27, 2018 — classified as 1 injury.

The most frequently reported specific device problem is “Connection Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Connection Problem1
  • Disconnection1

Patient problems

  • Blood Loss1
  • No Consequences Or Impact To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjurySep 27, 2018Disconnection; Connection Problem; Disconnection; Connection Problem

    IT WAS REPORTED THAT DURING A CARDIOPULMONARY BYPASS (CPB) PROCEDURE, THE 3/8 INCHES TUBING ATTACHED TO THE OUTLET PORT ON THE FX25 BEING USED GOT DISCONNECTED. THE CUSTOMER MADE THE CONNECTION BY PLACING THE TUBING OVER THE 2ND BARB OF THE OXYGENATOR OUTLET PORT. WHILE ON BYPASS THE LINE DISCONNECTED RESULTING IN APPROXIMATELY A 500 ML BLOOD LOSS. THE PRODUCT WAS NOT CHANGED OUT SINCE THE PERFUSIONIST WAS ABLE TO SAFELY RECONNECT THE LINE. THERE WAS NO TRANSFUSION REQUIRED AND THE SURGERY WAS COMPLETED SUCCESSFULLY. THE CUSTOMER CONFIRMED THAT PATIENT WAS OKAY.

    Patient problems: Blood Loss; No Consequences Or Impact To Patient

    MAUDE report 7915756 ↗

Sources and updates