Skip to content

Cardiovascular Procedure Kit 76374-01Recalls & reported problems

Terumo Cardiovascular Systems Corporation

All models of Cardiovascular Procedure Kit (227)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Feb 20, 2024Open, Classified

    Pressure relief valve included in certain lots of cardiovascular procedure kits was reportedly opening at approximately 200-300mmHg versus 400mmHg, potential myocardial tissue damage potentially requiring surgical or medical/pharmacological intervention

    Z-1401-2024 · Root cause: Process control · Recall details · FDA record ↗

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page