Skip to content

CELLEX CLXCC-IRecalls & reported problems

Therakos LLC · CellEX Procedural Kit RoHS 3 Correct Connect (OUS) · Class 3

All models of CELLEX (4)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00705030000029), received between Oct 30, 2025 and May 6, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Coagulation in Device or Device Ingredient” (2 reports).

8 other recalls cover devices cleared under the same FDA submission (P860003); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 2 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Coagulation in Device or Device Ingredient2
  • Fluid/Blood Leak2

Patient problems

  • No Clinical Signs, Symptoms or Conditions4

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 6, 2026Coagulation in Device or Device Ingredient

    THE CUSTOMER CONTACTED THERAKOS TO REPORT CLOTS OBSERVED IN THE RETURN LINE WITH THEIR CELLEX PHOTOPHERESIS KIT ("KIT") DURING AN EXTRACORPOREAL PHOTOPHERESIS (ECP) TREATMENT. THE CUSTOMER ALSO REPORTED ALARM #17: RETURN PRESSURE ALARMS AND OBSERVED BLOOD CLOTTING IN THE RETURN LINE WHEN APPROXIMATELY 355 ML OF WHOLE BLOOD HAD BEEN PROCESSED. THE ECP TREATMENT WAS ABORTED, AND RESIDUAL BLOOD WAS NOT RETURNED TO THE PATIENT. NO PRODUCT OR PHOTOGRAPHS WILL BE RETURNED FOR EVALUATION. THE PATIENT HAS BEEN REPORTED TO HAVE BEEN STABLE. NO PATIENT HARM HAS BEEN REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25103703 ↗
  • MalfunctionMay 5, 2026Fluid/Blood Leak

    THE CUSTOMER CONTACTED THERAKOS TO REPORT THEY EXPERIENCED A PUMP TUBING ORGANIZER (PTO) LEAK WITH THEIR CELLEX PHOTOPHERESIS KIT ("KIT") OCCURRING DURING THE BLOOD PRIME PHASES OF AN EXTRACORPOREAL PHOTOPHERESIS (ECP) TREATMENT. THE CUSTOMER REPORTED THAT THE LEAK APPEARED TO BE COMING FROM THE RETURN RESERVOIR AND THAT THE PATIENT HAD NOT YET BEEN CONNECTED. THE CUSTOMER ALSO REPORTED THAT ALARM #57: RETURN PUMP (#3) ERROR ALSO OCCURRED. THE KIT HAS BEEN DISCARDED, BUT THE CUSTOMER PROVIDED PHOTOGRAPHS FOR EVALUATION. NO PATIENT HARM HAS BEEN REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25086205 ↗
  • MalfunctionApr 30, 2026Fluid/Blood Leak

    THE CUSTOMER CONTACTED THERAKOS TO REPORT THAT THEY HAD TO ABORT AN EXTRACORPOREAL PHOTOPHERESIS (ECP) TREATMENT WITH THEIR CELLEX PHOTOPHERESIS KIT ("KIT") AFTER A TWO-MINUTE PHOTOACTIVATION AND UNSPECIFIED AIR DETECTED ALARMS. THE CUSTOMER THEN REALIZED THERE WAS A BLOOD LEAK FROM THE PHOTOACTIVATION MODULE. THE CUSTOMER HAS PROVIDED PHOTOGRAPHS FOR EVALUATION. THERE WAS NO REPORT OF PATIENT HARM ASSOCIATED WITH THIS EVENT, AND THE PATIENT HAS BEEN REPORTED TO HAVE BEEN STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25046674 ↗
  • MalfunctionOct 30, 2025Coagulation in Device or Device Ingredient; Coagulation in Device or Device Ingredient

    THE CUSTOMER CONTACTED THERAKOS TO REPORT BLOOD CLOTTING IN THE RETURN LINE WITH THEIR CELLEX PHOTOPHERESIS KIT ("KIT") DURING AN EXTRACORPOREAL PHOTOPHERESIS (ECP) TREATMENT. THE CUSTOMER REPORTED THEY OBSERVED CLOTS IN THE KIT AFTER MORE THAN 500 ML OF WHOLE BLOOD HAD BEEN PROCESSED. THE CLOTS WERE OBSERVED IN THE RETURN LINE, CENTRIFUGE BOWL, RETURN BAG, AND FILTER. THE ECP TREATMENT WAS ABORTED AND RESIDUAL BLOOD WITHIN THE KIT WAS NOT RETURNED TO THE PATIENT. THE CUSTOMER REPORTED THE PATIENT WAS IN STABLE CONDITION. THE CUSTOMER STATED THEY INJECTED IV CALCIUM INTO THE COLLECT LINE OF THE KIT. THE CUSTOMER STATED THAT THE LIKELY CAUSE OF THE CLOTTING WAS DUE TO THE INJECTION OF IV CALCIUM.THE CUSTOMER WILL RETURN PHOTOGRAPHS AND THE SMART CARD FOR EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23428457 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P860003. One submission often covers many models, so check the product and lot details in each record.

  • Z-1822-2026 · May 16, 2025 · Open, Classified
    Brand Name: Cellex Photopheresis System Product Name: CELLEX Photopheresis Kit Model/Catalog Number: CLXUSA Software Version: Not Applicable Product Description: THERAKOS Photopher
  • Z-0567-2022 · Dec 10, 2021 · Open, Classified
    THERAKOS CELLEX Photopheresis Procedural Kit, CLXUSA
  • Z-0212-2022 · Oct 4, 2021 · Open, Classified
    THERAKOS CELLEX Photopheresis Procedural Kit PRODUCT CODE: CLXUSA
  • Z-1322-2016 · Dec 21, 2015 · Terminated
    The THERAKOS CELLEX Photopheresis System is a photoimmune therapy. Catalog CLXUSA. Manufacturing lot D342 UDI: 20705030200003 The system designed to separate the patient's blood i
  • Z-0595-2012 · Sep 1, 2011 · Terminated
    Therakos Cellex Photopheresis System CellexUSA & Cellex. The photopheresis system is indicated for use in the ultraviolet-A (UVA) irradiation in the presence of the photoactive dr
  • Z-2067-2011 · Feb 1, 2011 · Terminated
    Therakos UVAR XTS Procedural Kits Is indicated for use in the ultraviolet-A (UVA) irradiation.
  • Z-2839-2011 · Dec 2, 2010 · Terminated
    Therakos Cellex Procedural Kits. Designed to interface with the Cellex Photopheresis System to perform cell separation and photoactivation in a single, closed and sterile circuit.
  • Z-1504-06 · Jul 18, 2006 · Terminated
    Therakos Photopheresis Procedural Kit for use with the UVAR XTS Instrument. (catalog number XT125)

Other SYSTEM, PHOTOPHERESIS, EXTRACORPOREAL devices

All SYSTEM, PHOTOPHERESIS, EXTRACORPOREAL devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page