Atellica® VTR Syringes 11561345Recalls & reported problems
Westmed, Inc. · 3mL Vented Luer Lock Syringe, (~50IU)16.7IU/mL Balanced Heparin, Vented Tip Cap (Private Label) · Class 1
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00709078013386), received between Aug 14, 2024 and Dec 18, 2025 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak3
- Component Missing1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Insufficient Information1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 18, 2025Component Missing; Component Missing
IT WAS REPORTED, BLOOD LEAKED WHILE EXPRESSING AIR, AND SPLASHED OUT FROM THE TOP OF THE CAP ONTO THE OPERATOR'S FACE. IT WAS ADDITIONALLY REPORTED, THE OPERATOR WAS WEARING GLOVES AND GOGGLES AND FOLLOWED THE HOSPITAL GUIDELINES FOR BLOOD EXPOSURE. WHILE INVESTIGATING THE EVENT, IT WAS FOUND THAT THE AFFECTED SYRINGE WAS MISSING THE WHITE FILTER OF THE TIP CAP. THERE WAS NO REPORT OF A DELAY IN THERAPY, HURT, HARM OR INJURY AS A RESULT OF THIS REPORTED EVENT.
Patient problems: Insufficient Information
MAUDE report 23844120 ↗MalfunctionAug 14, 2024Fluid/Blood Leak; Fluid/Blood Leak
SUNMED HOLDINGS LLC. RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT PATIENTS. THIS IS THE THIRD OF THREE REPORTS. REFER TO 2028807-2024-00024 FOR THE FIRST REPORT. REFER TO 2028807-2024-00025 FOR THE SECOND REPORT. IT WAS REPORTED, DURING AIR EXPULSION FOLLOWING BLOOD GAS AND WITH THE SYRINGE CAP IN PLACE, BLOOD LEAKED THROUGH THE CAP. IT WAS ADDITIONALLY NOTED, THE CAP REMAINED ON THE SYRINGE TIP. THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19981293 ↗MalfunctionAug 14, 2024Fluid/Blood Leak; Fluid/Blood Leak
SUNMED HOLDINGS LLC. RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT PATIENTS. THIS IS THE FIRST OF THREE REPORTS. REFER TO 2028807-2024-00025 FOR THE SECOND REPORT. REFER TO 2028807-2024-00026 FOR THE THIRD REPORT. IT WAS REPORTED, DURING AIR EXPULSION FOLLOWING BLOOD GAS AND WITH THE SYRINGE CAP IN PLACE, BLOOD LEAKED THROUGH THE CAP. IT WAS ADDITIONALLY NOTED, THE CAP REMAINED ON THE SYRINGE TIP. THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19981292 ↗MalfunctionAug 14, 2024Fluid/Blood Leak; Fluid/Blood Leak
SUNMED HOLDINGS LLC. RECEIVED A SINGLE REPORT THAT REFERENCED THREE DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THREE DIFFERENT PATIENTS. THIS IS THE SECOND OF THREE REPORTS. REFER TO 2028807-2024-00024 FOR THE FIRST REPORT. REFER TO 2028807-2024-00026 FOR THE THIRD REPORT. IT WAS REPORTED, DURING AIR EXPULSION FOLLOWING BLOOD GAS AND WITH THE SYRINGE CAP IN PLACE, BLOOD LEAKED THROUGH THE CAP. IT WAS ADDITIONALLY NOTED, THE CAP REMAINED ON THE SYRINGE TIP. THERE WAS NO REPORTED INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19981273 ↗
Other Arterial Blood Sampling Kit devices
- MEDLINE ART995Medline Industries, Inc. · Arterial blood sampling kit4 recalls0 reports
- MEDLINE ART1070AMedline Industries, Inc. · Arterial blood sampling kit3 recalls0 reports
- MEDLINE ART1180Medline Industries, Inc. · Arterial blood sampling kit2 recalls0 reports
- MEDLINE ART775BMedline Industries, Inc. · Arterial blood sampling kit2 recalls0 reports
- MEDLINE ART890AMedline Industries, Inc. · Arterial blood sampling kit2 recalls0 reports
- Centurion ART155ACenturion Medical Products Corporation · Arterial blood sampling kit2 recalls0 reports
- MEDLINE ART1160Medline Industries, Inc. · Arterial blood sampling kit2 recalls0 reports
- OTHER P42453Medline Industries, Inc. · Arterial blood sampling kit2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page