Skip to content

Atomlab 500 Dose Calibrator 086-330Recalls & reported problems

Biodex Medical Systems, Inc. · Atomlab 500 Dose Calibrator · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Recalls naming this device

  1. Aug 16, 2021Terminated

    When deleting a previously entered custom isotope, the software deletes the isotope, but not the associated dial setting. Possible other stored custom isotopes may then have incorrect dial settings and may lead to an incorrect dose measurement and/or misadministration

    Z-2580-2021 · Root cause: Software design · Recall details · FDA record ↗

Other Calibrator, Dose, Radionuclide devices

All Calibrator, Dose, Radionuclide devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page