Skip to content

Midas Rex® AF02Recalls & reported problems

Medtronic Powered Surgical Solutions · ATT AF02 LEGEND F2 (B-1) 2.4MM STEM · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Quick summary

This Midas Rex AF02 is a pneumatic surgical drill attachment used during surgery. Of the 161 reports, 158 are malfunctions, and many describe bearings coming loose, attachments that do not lock tools or cut, and some describe overheating. The reports include no deaths and 3 injuries, and reports rose to 34 in the last 12 months from 10 before. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
163
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 163 adverse event reports that list this exact device (DI 00721902481857), received between Mar 16, 2017 and Sep 29, 2026 — classified as 3 injuries, 160 malfunctions.

The most frequently reported specific device problem is “Material Puncture/Hole” (79 reports).

9 other recalls cover devices cleared under the same FDA submission (K020069); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 4 reportsJan 2024: 2 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 10 reportsSep 2024: 6 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 1 reportsJan 2025: 0 reportsFeb 2025: 3 reportsMar 2025: 2 reportsApr 2025: 3 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 3 reportsMay 2026: 5 reportsJun 2026: 21 reportsJul 2026: 3 reportsAug 2026: 1 reportsSep 2026: 1 reportsOct 2026: 0 reports20242025202621
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Puncture/Hole79
  • Detachment of Device or Device Component40
  • No Apparent Adverse Event24
  • Naturally Worn15
  • Overheating of Device14
  • Device Abrasion From Instrument Or Another Object13
  • Difficult to Insert12
  • Physical Resistance/Sticking8

Patient problems

  • No Clinical Signs, Symptoms or Conditions112
  • No Patient Involvement24
  • No Consequences Or Impact To Patient21
  • No Known Impact Or Consequence To Patient6
  • Insufficient Information2
  • Hematoma1
  • Injury1
  • Perforation1

Recent adverse event reports

The 25 most recent of 163 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 29, 2026Naturally Worn

    It was reported that prior to use, the bearings in the attachment were damaged. There was no patient involved in this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50927496 ↗
  • MalfunctionAug 17, 2026Loss of Power

    It was reported that during intraop the attachment was frozen. It was reported that there was no patient impact. Additional information was received stating that detached bearings were found during evaluation

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50269227 ↗
  • MalfunctionJul 5, 2026Difficult to Insert

    IT WAS REPORTED THAT THE ATTACHMENT DOES NOT CUT. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25712301 ↗
  • MalfunctionJul 1, 2026No Apparent Adverse Event

    MEDTRONIC RECEIVED INFORMATION REGARDING A REPAIR REQUEST AND NO ALLEGED PRODUCT DEFICIENCIES WERE REPORTED. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARING WAS FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25679619 ↗
  • MalfunctionJul 1, 2026Difficult to Insert

    IT WAS REPORTED THAT THE ATTACHMENT DOES NOT RETAIN THE TOOL. AT THE TIME OF THE REPORT, THERE WAS NO KNOWN IMPACT TO THE PATIENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: Insufficient Information

    MAUDE report 25675285 ↗
  • MalfunctionJun 26, 2026Naturally Worn

    IT WAS REPORTED THAT ATTACHMENT BEARING WAS DAMAGED PRIOR TO USE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARING DETACHED WAS FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25632084 ↗
  • MalfunctionJun 24, 2026No Apparent Adverse Event

    IT WAS REPORTED THAT THE DEVICE NOT WORKING. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARINGS DETACHED WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25611591 ↗
  • MalfunctionJun 24, 2026Difficult to Insert

    IT WAS REPORTED THAT THE DEVICE CANNOT LOCK TOOLS. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25607127 ↗
  • MalfunctionJun 24, 2026Difficult to Insert

    IT WAS REPORTED THAT THE ATTACHMENT CANNOT LOCK TOOLS. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25604375 ↗
  • MalfunctionJun 16, 2026Overheating of Device

    IT WAS REPORTED THAT THE ATTACHMENT WITH THE REPORT OF OVERHEATING. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25515987 ↗
  • MalfunctionJun 16, 2026Physical Resistance/Sticking

    IT WAS REPORTED THAT THE ATTACHMENT WITH THE REPORT OF LOCK OR ADJUSTMENT RING STICKING. IT WAS REPORTED THERE WAS NO PATIENT INVOLV EMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25514086 ↗
  • MalfunctionJun 16, 2026Naturally Worn

    IT WAS REPORTED THAT THE ATTACHMENT HAD BEARINGS DEFECT. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25513887 ↗
  • MalfunctionJun 15, 2026Difficult to Insert

    IT WAS REPORTED THAT THE ATTACHMENT CANNOT LOCK THE TOOL. THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25498327 ↗
  • MalfunctionJun 15, 2026Naturally Worn

    IT WAS REPORTED THAT THE ATTACHMENT BEARINGS WERE DEFECTIVE. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25498256 ↗
  • MalfunctionJun 12, 2026Naturally Worn

    IT WAS REPORTED THAT THE DURING PRE-OP THE ATTACHMENT HAD BEARING ISSUE. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25481566 ↗
  • MalfunctionJun 11, 2026Loss of Power

    IT WAS REPORTED THAT THE ATTACHMENT WITH THE REPORT OF LOSS OF POWER. IT WAS REPORTED THAT THERE WAS NO PATIENT IMPACT.ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25466783 ↗
  • MalfunctionJun 9, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT THE ATTACHMENT WITH THE REPORT OF DETACHMENT OF COMPONENT. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25444988 ↗
  • MalfunctionJun 9, 2026Overheating of Device; Naturally Worn

    IT WAS REPORTED THAT THE BEARING WAS WORN AND OVERHEATING PRIOR TO USE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT BEARING DETACHED WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25438886 ↗
  • MalfunctionJun 8, 2026Material Twisted/Bent

    IT WAS REPORTED THAT THE DURING PREOP FOOT PROTECTION WAS ASKEW. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25423821 ↗
  • MalfunctionJun 4, 2026Physical Resistance/Sticking

    IT WAS REPORTED THAT THE ATTACHMENT WAS LOCK FAILURE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25395199 ↗
  • MalfunctionJun 4, 2026Difficult to Insert

    IT WAS REPORTED THAT THE ATTACHMENT WAS UNABLE TO LOCK PROPERLY WITH THE TOOL ONTO THE HANDPIECE. IT WAS REPORTED THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25389100 ↗
  • MalfunctionJun 1, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT PARTS WERE FELL OUT AFTER A USE. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25353582 ↗
  • MalfunctionJun 1, 2026Difficult to Insert

    IT WAS REPORTED THAT THE DURING INTRAOP IN ATTACHMENT THE BUR COULD NOT BE INSERTED. AT THE TIME OF THE REPORT, THERE WAS NO KNOWN IMPACT TO THE PATIENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT THERE WERE BROKEN BEARINGS INSIDE THE DEVICE.

    Patient problems: Insufficient Information

    MAUDE report 25353364 ↗
  • MalfunctionJun 1, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT THE ATTACHMENT WITH THE REPORT OF BEARING DETACHMENT. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25350827 ↗
  • MalfunctionJun 1, 2026Detachment of Device or Device Component

    IT WAS REPORTED THAT THE DEVICE OF PARTS FALL OFF. IT WAS REPORTED THAT THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS RECEIVED STATING THAT DETACHED BEARINGS WERE FOUND DURING EVALUATION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25348435 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K020069. One submission often covers many models, so check the product and lot details in each record.

  • Z-1602-2016 · Apr 8, 2016 · Terminated
    Medtronic Midas Rex Dissecting Tool Legend 10 cm 4mm BA DIAM. Intended for use in neurosurgery.
  • Z-0321-2016 · Oct 21, 2015 · Terminated
    Tool Legend 16cm 2.4 mm wire pass for Midas Rex Legend Dissecting Tool. Used by surgeons to drill holes.
  • Z-0546-2010 · Nov 30, 2009 · Terminated
    Medtronic Midas Rex Legend Footed Attachment, AF01. Intended for use with all Legend and Legend EHS motors for general surgery including maxillofacial , craniofacial and sternotom
  • Z-2156-2008 · May 30, 2008 · Terminated
    Medtronic Midas Rex Legend tapered dissecting tool, Model F2/8TA23 Medtronic Powered Surgical Solutions, Forth Worth, TX Intended for use in Spinal, Cranial, Orthopedic, General S
  • Z-1574-05 · Sep 1, 2005 · Terminated
    Medtronic Midas Rex Classic, GS and Legend dissecting tools manufactured by Medtronic Inc. Powered Surgical Solutions, 4620 North Beach Street, Fort Worth, tX 76137 USA, Rx only, S
  • Z-1067-05 · Jun 29, 2005 · Terminated
    Medtronic MIDAS REX Legend High Speed Pneumatic System, Gold Touch Motor. Manufactured by Medtronic, Inc. Powered Surgical Solutions, 4620 North Beach Street, Fort Worth, TX 76137
  • Z-0278-05 · Nov 11, 2004 · Terminated
    Medtronic Midas Rex Legend High Speed Pneumatic System
  • Z-0213-05 · Nov 3, 2004 · Terminated
    Medtronic Midas Rex Legend Lubricant/ Diffuser Cartridge Catalog number PA100-A, manufactured by Medtronic Powered Surgical Solutions, Fort Worth, Texas
  • Z-0419-03 · Nov 11, 2002 · Terminated
    Legend Lubricant System (Legend Gold, Gold Touch and Platinum motors and Legend Lubricant/Diffuser cartridge)

Other MOTOR, DRILL, PNEUMATIC devices

All MOTOR, DRILL, PNEUMATIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page