CD HORIZON® Spinal System 8115558Recalls & reported problems
Medtronic Sofamor Danek, Inc. · CROSSLINK 8115558 58TO80MM X10 MULTI · Class 2 · Implant
All models of CD HORIZON Spinal System (342)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00721902729782), received between May 8, 2017 and Apr 19, 2022 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Break” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break4
Patient problems
- No Known Impact Or Consequence To Patient3
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 19, 2022Break; Break
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING POSTERIOR LUMBAR SPINAL FUSION. IT WAS REPORTED THAT DURING FINAL TIGHTENING OF CROSSLINK ONE OUTER SET SCREW ON ROD SHEARED OFF BELOW THE NORMAL SHEAR POINT, LEAVING A SMALL CONICAL PIECE OF THE LEADING EDGE OF THE CROSSLINK SET SCREW AND NO PURCHASE TO THE ROD. AT THIS POINT THERE WAS NO ABILITY TO REMOVE THE IMPROPERLY TIGHTENED SET SCREW SO THE RODS HAD TO BE REMOVED TO REMOVE THE CROSSLINK FROM THE PATIENT. NEW RODS WERE PLACED WITHOUT A CROSSLINK. THERE WAS NO PATIENT SYMPTOM REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14140888 ↗MalfunctionMar 22, 2020Break; Break
IT WAS REPORTED THAT THE PATIENT PRESENTED WITH LUMBAR DEGENERATION; AND UNDERWENT POSTERIOR LUMBAR DECOMPRESSION AND FUSION. INTRA-OP, THE ALLEGED IMPLANT BROKE. THE BROKEN IMPLANT WAS THEN COMPLETELY EXPLANTED WITH NO BROKEN FRAGMENTS REMAINING INSIDE THE PATIENT. THE BROKEN PRODUCT WAS REPLACED WITH A NEW ONE, AND THE SURGERY WAS COMPLETED. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9863799 ↗MalfunctionNov 16, 2018Break; Break
IT WAS REPORTED THAT THE PATIENT WAS DIAGNOSED WITH LUMBAR DEGENERATION; AND UNDERWENT LUMBAR INTERNAL FIXATION. INTRA-OP, THE CROSSLINK PLATE BROKE. THE BROKEN PLATE WAS COMPLETELY EXPLANTED FROM THE PATIENT. NO FRAGMENTS OF THE BROKEN IMPLANT REMAINED INSIDE THE PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8076977 ↗MalfunctionMay 8, 2017Break; Break
IT WAS REPORTED THAT DURING SURGERY, THE SET SCREW OF THE CROSSLINK BROKE IN THE MIDDLE. THEN THE CROSSLINK WAS CHANGED WITH ANOTHER ONE. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THE EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6550908 ↗
Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- XLP 7565050Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430344Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- ATLANTIS® Anterior Cervical Plate System 876-170Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- XLP 7565060Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
- Elegance® Anterior Cervical Plate 53430113Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page