CD HORIZON® Spinal System 8116439Recalls & reported problems
Medtronic Sofamor Danek, Inc. · CROSSLINK 8116439 39 45MM X10 MULTI · Class 2 · Implant
All models of CD HORIZON Spinal System (342)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00721902819667), received between Mar 7, 2018 and Dec 3, 2025 — classified as 1 injury, 4 malfunctions.
The most frequently reported specific device problem is “Break” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break5
Patient problems
- No Known Impact Or Consequence To Patient3
- Inadequate Osseointegration1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryDec 3, 2025Break; Break
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING SPINAL THERAPY. IT WAS REPORTED THAT DURING A POSTERIOR SPINAL FUSION PROCEDURE, THE T20 TORX DRIVER WAS STRIPPED. THE TIP OF THE DRIVER WAS FRACTURED AND CROSSLINK FRACTURED WHERE IT ATTACHES TO ROD. THE FRACTURED IMPLANTS OCCURRED PRIOR TO THE REVISION SURGERY ON 11/13. IT WAS SUSPECTED BY THE PHYSICIAN THAT THE IMPLANTS (SCREW AND CROSSLINK) FRACTURED DUE TO STRESS ON HARDWARE FROM NON-FUSION OF THE PATIENT. THE CROSSLINK AND SCREW WERE EXPLANTED FROM THE PATIENT. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Patient problems: Inadequate Osseointegration
MAUDE report 23701738 ↗MalfunctionSep 19, 2022Break; Break
INFORMATION WAS RECEIVED FROM A HEALTHCARE PROFESSIONAL VIA MANUFACTURER REPRESENTATIVE REGARDING A SET SCREW USED IN A SPINAL THERAPY FOR SCOLIOSIS. IT WAS REPORTED THAT THE SURGEON HAD USED A NUT AT THE END OF THE ASSEMBLY TO PROTECT THE SCREW HEAD. WHEN TIGHTENING, THE SCREW THREAD BROKE AND METAL FILINGS APPEARED. THIS WAS AN ISOLATED INCIDENT. TO COMPENSATE FOR THE INCIDENT, A SECOND DEVICE OF THE SAME REFERENCE AND BATCH WAS USED. THERE WERE NO CLINICAL CONSEQUENCES FOR THE PATIENT. THERE WERE NO PATIENT COMPLICATIONS. NO FURTHER COMPLICATIONS WERE REPORTED/ANTICIPATED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15444109 ↗MalfunctionJul 24, 2019Break; Break
IT WAS REPORTED THAT THE PATIENT UNDERWENT OPEN LUMBAR FUSION T10-S2 DUE TO DEGENERATIVE DISC DISEASE. INTRA-OPERATIVELY, THE SET SCREW BROKE OFF AT THE THREAD LEVEL NOT AT THE SMOOTH BREAK OFF PORTION OF SET SCREW. METAL CUTTING BURR WAS USED TO CUT THE SET SCREW AND REMOVED CROSSLINK WITH VICEGRIPS. THERE WERE NO PATIENT COMPLICATIONS AS A RESULT OF ALLEGED EVENT. NO FRAGMENT OF THE PRODUCT IS REMAINING IN THE PATIENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8820261 ↗MalfunctionMar 31, 2019Break; Break
IT WAS REPORTED THAT ON UNKNOWN DATE, INTRA-OP, WHEN THE SCREW WAS SCREWED, THE SCREW PLUG BROKE AT THE NON-BREAKING POINT. THE PRODUCT CAME IN CONTACT WITH THE PATIENT. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 8468024 ↗MalfunctionMar 7, 2018Break; Break
IT WAS REPORTED THAT PATIENT WAS PRESENTED WITH THORACIC SPINAL CANAL STENOSIS AND UNDERWENT LAMINA OF THORACIC VERTEBRA CUTTING DECOMPRESSION ,CANALIS SPINALIS EXPANDED MOLDING, VERTEBRAL PEDICLE INTERNAL FIXATION OPERATION. INTRA-OP, CROSS LINK PLUG SCREW ABNORMALLY FRACTURED. NO PATIENT COMPLICATIONS WERE REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7319728 ↗
Other APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY devices
- CD HORIZON® SOLERA® VOYAGER™ Spinal System 6550017Medtronic Sofamor Danek, Inc. · Appliance, fixation, spinal intervertebral body131 reports
- EVEREST® Spinal System 2901-10001VB Spine LLC · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY119 reports
- ZEVO™ Anterior Cervical Plate System 7710015Medtronic Sofamor Danek, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall1 report
- Mariner RDX MX1-000040Seaspine Orthopedics Corporation · Appliance, fixation, spinal intervertebral body1 recall0 reports
- Mariner RDX MX1-000030Seaspine Orthopedics Corporation · Appliance, fixation, spinal intervertebral body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430112Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- Elegance® Anterior Cervical Plate 53430224Clariance · Appliance, Fixation, Spinal Intervertebral Body1 recall0 reports
- XLP 7565050Nuvasive, Inc. · APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BODY1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page