NA AX083020003Recalls & reported problems

Steris Mexico, S. de R.L. de C.V. · The Atlas Transfer Carriage, Painted Carbon Steel 26-inch, x 37.5-inch, x 66-inch is designed for use with sterilization systems. · Class 2

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
23
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 23 adverse event reports that list this exact device (DI 00724995150389), received between Sep 9, 2021 and Sep 30, 2026 — classified as 1 injury, 22 malfunctions.

The most frequently reported specific device problem is “Device Tipped Over” (13 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 2 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 3 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 2 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Tipped Over13
  • Unintended Movement4
  • Device Fell3
  • Insufficient Device Problem Information1
  • Mechanical Problem1
  • Use of Device Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions16
  • Pain3
  • Insufficient Information2
  • Ecchymosis1
  • Limb Pain1

Recent adverse event reports

The 23 most recent of 23 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 30, 2026Device Fell

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50951205 ↗
  • InjurySep 10, 2026Device Fell

    The user facility reported that while an employee was using their Evolution Transfer Carriage, the carriage became unstable and fell to the floor. The employee was reportedly bruised and sore. It is unknown if medical treatment was sought or administered.

    Patient problems: Ecchymosis

    MAUDE report 50713949 ↗
  • MalfunctionJul 16, 2026Use of Device Problem

    Patient problems: Limb Pain

    MAUDE report 25843714 ↗
  • MalfunctionMay 27, 2026Unintended Movement

    THE USER FACILITY REPORTED THAT WHILE AN EMPLOYEE WAS UNLOADING THEIR STERILIZER, THEIR EVOLUTION TRANSFER CARRIAGE BECAME UNSTABLE AND FELL, DAMAGING THE LOADING CART. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25302607 ↗
  • MalfunctionMay 14, 2026Unintended Movement

    THE USER FACILITY REPORTED THAT WHILE AN EMPLOYEE WAS UNLOADING THEIR STERILIZER, THEIR EVOLUTION TRANSFER CARRIAGE BECAME UNSTABLE AND FELL, DAMAGING THE INSTRUMENT RACK. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25173150 ↗
  • MalfunctionMay 1, 2026Unintended Movement

    THE USER FACILITY REPORTED THAT WHILE AN EMPLOYEE WAS UNLOADING THEIR STERILIZER, THEIR EVOLUTION TRANSFER CARRIAGE BECAME UNSTABLE AND THE LOAD FELL, DAMAGING THE TRANSFER CARRIAGE. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25056377 ↗
  • MalfunctionMar 25, 2026Device Fell

    THE USER FACILITY REPORTED THAT AN EMPLOYEE OBTAINED A SCRATCH ON THEIR HAND AFTER THEIR EVOLUTION TRANSFER CARRIAGE FELL. THE EMPLOYEE DID NOT SEEK MEDICAL TREATMENT AND CONTINUED WORKING.

    Patient problems: Pain

    MAUDE report 24688746 ↗
  • MalfunctionJan 28, 2026Mechanical Problem

    THE USER FACILITY REPORTED THAT THE FRONT WHEELS TO THEIR EVOLUTION TRANSFER CARRIAGE DID NOT LATCH PROPERLY TO THE STERILIZER'S DOCKING STATION SUBSEQUENTLY CAUSING THE CARRIAGE TO FALL. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24203864 ↗
  • MalfunctionDec 18, 2025Unintended Movement; Unintended Movement

    THE USER FACILITY REPORTED THAT THEIR EVOLUTION TRANSFER CARRIAGE FELL TO THE FLOOR WHILE AN EMPLOYEE WAS UNLOADING THE STERILIZER. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23843486 ↗
  • MalfunctionSep 12, 2025Device Tipped Over; Device Tipped Over

    THE USER FACILITY REPORTED THAT WHILE AN EMPLOYEE WAS UNLOADING THEIR STERILIZER, THEIR EVOLUTION TRANSFER CARRIAGE BECAME UNSTABLE AND THE INSTRUMENT RACK FELL TO THE FLOOR. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23039419 ↗
  • MalfunctionSep 4, 2025Device Tipped Over; Device Tipped Over

    THE USER FACILITY REPORTED THEIR EVOLUTION TRANSFER CARRIAGE "FELL DOWN" WHILE TRANSPORTING A LOAD. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22974397 ↗
  • MalfunctionMar 15, 2024Device Tipped Over; Device Tipped Over

    THE USER FACILITY REPORTED THAT WHILE AN EMPLOYEE WAS UNLOADING THEIR STERILIZER, THE EVOLUTION TRANSFER CARRIAGE BECAME UNSTABLE AND THE SHELF WITH STERILE HOT INSTRUMENTS ON IT FELL ONTO THE FLOOR. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18914860 ↗
  • MalfunctionFeb 9, 2024Device Tipped Over; Device Tipped Over

    THE USER FACILITY REPORTED THAT THEIR EVOLUTION TRANSFER CARRIAGE FELL FORWARD WHILE AN EMPLOYEE WAS UNLOADING THEIR STERILIZER RESULTING IN SHOULDER PAIN. THE EMPLOYEE SOUGHT MEDICAL TREATMENT AND RECEIVED AN X-RAY. THE EMPLOYEE HAS RETURNED TO WORK.

    Patient problems: Pain

    MAUDE report 18676806 ↗
  • MalfunctionDec 8, 2023Device Tipped Over; Device Tipped Over

    THE USER FACILITY REPORTED THAT THEIR 66" EVOLUTION TRANSFER CARRIAGE FELL TO THE FLOOR WHILE AN EMPLOYEE WAS UNLOADING THE STERILIZER. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18291456 ↗
  • MalfunctionNov 17, 2023Device Tipped Over; Device Tipped Over

    THE USER FACILITY REPORTED THAT THEIR 66" EVOLUTION TRANSFER CARRIAGE FELL TO THE GROUND DURING UNLOADING FROM A STERILIZER. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18159616 ↗
  • MalfunctionAug 10, 2023Insufficient Device Problem Information; Insufficient Device Problem Information

    THE USER FACILITY REPORTED THAT AN EMPLOYEE OBTAINED AN INJURY WHILE UNLOADING THEIR EVOLUTION TRANSFER CARRIAGE FROM THE STERILIZER. NO MEDICAL TREATMENT WAS SOUGHT OR ADMINISTERED. THE USER FACILITY DID NOT DISCLOSE SPECIFIC DETAILS PERTAINING TO HOW THE EMPLOYEE OBTAINED THE INJURY.

    Patient problems: Insufficient Information

    MAUDE report 17503044 ↗
  • MalfunctionFeb 16, 2023Device Tipped Over; Device Tipped Over

    THE USER FACILITY REPORTED THAT WHILE AN EMPLOYEE WAS OPERATING THEIR EVOLUTION TRANSFER CARRIAGE, IT BECAME UNSTABLE AND FELL TO THE FLOOR RESULTING IN AN INJURY.

    Patient problems: Insufficient Information

    MAUDE report 16385001 ↗
  • MalfunctionSep 16, 2022Device Tipped Over; Device Tipped Over

    THE USER FACILITY REPORTED THAT WHILE AN EMPLOYEE WAS UNLOADING THEIR EVOLUTION TRANSFER CARRIAGE, IT FELL TO THE FLOOR CONTACTING THEIR FOOT. THE EMPLOYEE RECEIVED X-RAYS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15439605 ↗
  • MalfunctionAug 10, 2022Device Tipped Over; Device Tipped Over

    THE USER FACILITY REPORTED THAT THEIR LOADING CART FELL TO THE FLOOR WHILE AN EMPLOYEE WAS GLIDING IT ONTO THEIR 66" TRANSFER CARRIAGE. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15199123 ↗
  • MalfunctionAug 10, 2022Device Tipped Over; Device Tipped Over

    THE USER FACILITY REPORTED THAT WHEN AN EMPLOYEE WAS REMOVING THE LOADING CART FROM INSIDE THE STERILIZER, THEIR 66" TRANSFER CARRIAGE DID NOT REMAIN ENGAGED IN THE DOCKING STATION. THE TRANSFER CARRIAGE SUBSEQUENTLY FELL TO THE GROUND CAUSING THE OPERATOR TO OBTAIN AN INJURY TO THEIR FOOT. NO MEDICAL TREATMENT WAS SOUGHT OR ADMINISTERED.

    Patient problems: Pain

    MAUDE report 15198948 ↗
  • MalfunctionApr 28, 2022Device Tipped Over; Device Tipped Over

    THE USER FACILITY REPORTED THAT THEIR EVOLUTION TRANSFER CARRIAGE FELL TO THE FLOOR WHILE AN EMPLOYEE WAS UNLOADING THEIR STERILIZER. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14237903 ↗
  • MalfunctionMar 9, 2022Device Tipped Over; Device Tipped Over

    THE USER FACILITY REPORTED THAT THEIR EVOLUTION TRANSFER CARRIAGE FELL TO THE FLOOR. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13715224 ↗
  • MalfunctionSep 9, 2021Device Tipped Over; Device Tipped Over

    THE USER FACILITY REPORTED THAT THEIR EVOLUTION TRANSFER CARRIAGE FELL TO THE FLOOR WHILE AN EMPLOYEE WAS UNLOADING THEIR STERILIZER. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12451493 ↗

Other Sterilizer, steam devices