NA Quick Connect (REF QPC3005E)Recalls & reported problems
Steris Corporation · The Quick Connect for Pentax EG-3670urk Ultrasound Video Endoscopes are used in the SYSTEM 1E and, or SYSTEM 1 endo Liquid Chemical Sterilant Processing Systems. · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00724995162566), received between Jan 15, 2021 and Jan 15, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Device Damaged by Another Device” (1 report).
8 other recalls cover devices cleared under the same FDA submission (K090036, K101409, K102462, K113520, K131078, K161683, K170956, K173256, K180342, K182827, K190104, K191343, K192929, K210737, K211607, K222615, K230582, K232914, K251666); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Device Damaged by Another Device1
- Failure to Disconnect1
- Partial Blockage1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 15, 2021Partial Blockage; Failure to Disconnect; Device Damaged by Another Device
PENTAX MEDICAL WAS MADE AWARE OF A COMPLAINT ON (B)(6) 2020 UNDER RMA NOTIFICATION (B)(4) FOR AN "ACCESSORY STUCK IN SCOPE", INVOLVING PENTAX MEDICAL VIDEO GASTROSCOPE MODEL EG-3670URK, SERIAL NUMBER (B)(4) AT (B)(6) HOSPITAL. THE REPORTER ALSO STATED THAT THE EVENT OCCURRED DURING THE PROCEDURE ALTHOUGH THERE WAS NO PATIENT INVOLVEMENT OR INJURY REPORTED. THE USER FACILITY USED A TAPERED TIP CATHETER DURING THE PROCEDURE. THE USER FACILITY REPLIED VIA EMAIL TO A GOOD FAITH EFFORT EMAIL ON 22-DEC-2020, STATING THE CATHETER WAS A BOSTON SCIENTIFIC RAPID EXCHANGE TAPERED TIP XL CANNULA REF #M00545820. THE CUSTOMER OWNED ENDOSCOPE WAS RECEIVED BY PENTAX MEDICAL FOR EVALUATION ON 22-DEC-2020. DURING EVALUATION OF THE ENDOSCOPE UNDER SERVICE ORDER 3130641, THE PENTAX MEDICAL SERVICE REPAIR TECHNICIAN FOUND AN "ACCESSORY STUCK IN PRIMARY OPERATION CHANNEL" CONFIRMING THE CUSTOMER'S COMPLAINT AND DOCUMENTING THE FOLLOWING ADDITIONAL FINDINGS ON 29-DEC-2020: BIOPSY INLET T-PIECE STUCK ACCESSORY AT AFT TUBE, SUCTION TUBE RESISTANCE, ULTRASOUND IMAGE NOISE/ INTERFERENCE, RESIDUE ON AIR/WATER SOCKET, LEAK AT CONTROL BODY, CONTROL BODY CRACKED FROM SIDE BODY COVER AT PROXIMAL END, FAILED DRY L
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11179943 ↗
FDA clearance and classification
Cleared through 510(k) K251666 on Apr 8, 2026 (decision: Substantially Equivalent). Submitted by STERIS Corporation as “SYSTEM 1E Liquid Chemical Sterilant Processing System, Model P6700; SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6900”. The FDA review took 313 days. FDA record ↗
Cleared through Special 510(k) K232914 on Oct 3, 2023 (decision: Substantially Equivalent). Submitted by STERIS Corporation as “SYSTEM 1E Liquid Chemical Sterilant Processing System (P6500); SYSTEM 1 endo Liquid Chemical Sterilant Processing System (P6800); SYSTEM 1 endo Liquid Chemical Sterilant Processing System (P6900)”. The FDA review took 14 days. FDA record ↗
Cleared through Special 510(k) K230582 on Mar 28, 2023 (decision: Substantially Equivalent). Submitted by STERIS Corporation as “SYSTEM 1 endo Liquid Chemical Sterilant Processing System, Model P6900”. The FDA review took 26 days. FDA record ↗
…and 16 more FDA submissions listed for this device.
The FDA regulates product code MED (sterilant, medical devices) under 21 CFR 880.6885 as a Class II device. 21 CFR 880.6885 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K090036, K101409, K102462, K113520, K131078, K161683, K170956, K173256, K180342, K182827, K190104, K191343, K192929, K210737, K211607, K222615, K230582, K232914, K251666. One submission often covers many models, so check the product and lot details in each record.
- Z-1180-2016 · Dec 4, 2015 · TerminatedQuick Connects. Steris Corporation. Model numbers: QKC1736E, QKC1736INT, QKC1737E, and QKC1737INT. Each Quick Connect contains a flow unit consisting of tubing, tethers, and adapt
- Z-0616-2014 · Nov 21, 2013 · TerminatedSYSTEM 1E Liquid Chemical Sterilant Processing System, Distributed by STERIS Corporation, Mentor, OH The SYSTEM 1E Liquid Chemical Sterilant Processing System is intended for the
- Z-1596-2013 · Mar 4, 2013 · TerminatedQuick Connect QLC1677E, QRC1699E, QPC1721E and QFC1683E. STERIS Corporation, Mentor, OH 44060 QLC1677E, QRC1699E, QPC1721E and QFC1683E contain processing instructions, a Quick Re
- Z-2284-2012 · Aug 10, 2012 · TerminatedQKC1692E The QKC1692E Quick Connect device connects endoscopes to the sterilization tray for C1220E liquid chemical sterilization processor.
- Z-2335-2012 · Aug 1, 2012 · TerminatedThe SYSTEM 1E Liquid Sterilant Processing System. The SYSTEM 1E Liquid Sterilant Processing System is intended for the liquid chemical sterilization of manually cleaned immersib
- Z-0869-2012 · Dec 7, 2011 · TerminatedSTERIS SYSTEM 1E Liquid Chemical Sterilant Processing System. Model P6500, 510 (k) Numbers: K090036 Cleared 4/05/2010; K101409 cleared 8/2/2010; K102462, cleared 9/2/2010; add to
- Z-3124-2011 · Aug 11, 2011 · TerminatedSYSTEM 1E Liquid Chemical Sterilant Processing System, Model # P6500, S/N: 400001 through 405005; Mfr. By: STERIS Corporation, Mentor, OH 44060. Product Usage: The SYSTEM 1E Liqu
- Z-1960-2011 · Mar 14, 2011 · TerminatedSTERIS SYSTEM 1E Liquid Chemical Sterilant Processing System The SYSTEM 1 E Liquid Chemical Sterilant Processing System is intended for the liquid chemical sterilization of manual
Common questions about the NA Quick Connect
Has the NA Quick Connect been recalled?
No FDA recall record in our data names this exact device identifier (DI 00724995162566), based on the FDA recall file dated Nov 27, 2026. However, 8 recalls involve devices cleared under the same FDA submission. They may or may not include this model, so check the product, model and lot details in those records.
What problems have been reported with the NA Quick Connect?
The FDA's MAUDE database lists 1 adverse event report for this device, received on Jan 15, 2021. That report lists device problems “Device Damaged by Another Device” (1), “Failure to Disconnect” (1) and “Partial Blockage” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the NA Quick Connect?
None of the reports are classified as deaths, 0 as injuries and 1 as a malfunction. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the NA Quick Connect FDA approved?
It was cleared through 510(k) K090036 on Apr 5, 2010; its record lists 19 FDA submissions in total. Strictly, the FDA "clears" 510(k) devices as substantially equivalent to devices already on the market rather than "approving" them; approval is the PMA route.
Who makes the NA Quick Connect, and is it still sold?
The FDA's device database lists Steris Corporation as the company responsible for it (brand NA, model Quick Connect). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Sterilant, medical devices devices
- Revital-Ox Resert 4455AWSteris Corporation · Sterilant, medical devices1 recall3 reports
- McKesson 341Mckesson Medical-Surgical Inc. · Sterilant, Medical Devices1 recall0 reports
- MetriCide 10-1400Metrex Research, LLC · Sterilant, medical devices1 recall0 reports
- McKesson 68-102800Mckesson Medical-Surgical Inc. · Sterilant, Medical Devices1 recall0 reports
- MetriCide28 10-2800Metrex Research, LLC · Sterilant, medical devices1 recall0 reports
- McKesson 68-101400Mckesson Medical-Surgical Inc. · Sterilant, Medical Devices1 recall0 reports
- CIDEX™ 20390Advanced Sterilization Products Services Inc. · Sterilant, medical devices33 reports
- Rapicide High Level Disinfectant and Sterilant ML02-0059Medivators Inc. · Sterilant, Medical Devices6 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page