Skip to content

Raptor 00711177 (UDI 00724995183653)Recalls & reported problems

US Endoscopy · The disposable Raptor grasping device is used to grasp tissue and/or retrieve foreign bodies, excised tissue and stents during endoscopic procedures. · Class 2

All models of Raptor (3)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Quick summary

This Raptor grasping device is a disposable tool used during endoscopy to grab tissue and retrieve foreign bodies or stents. The 28 reports mostly describe jaws that would not open or close, locked up, or broke during a procedure, plus a few cases where the device became stuck in the bowel and its wires had to be cut. There were no deaths and no injuries. Reports rose from 3 in the prior 12 months to 9 in the most recent 12 months. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
27
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 27 adverse event reports that list this exact device (DI 00724995183653), received between Jul 17, 2020 and Jun 5, 2026 — classified as 27 malfunctions.

The most frequently reported specific device problem is “Difficult to Open or Close” (10 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 1 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 2 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 3 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Difficult to Open or Close10
  • Insufficient Device Problem Information6
  • Mechanical Problem5
  • Break2
  • Unintended Movement2
  • Activation, Positioning or Separation Problem1
  • Defective Device1
  • Difficult to Remove1

Patient problems

  • No Clinical Signs, Symptoms or Conditions23
  • Insufficient Information3
  • Foreign Body In Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 25 most recent of 27 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 5, 2026

    A RETAINED ESOPHAGEAL STENT IN THE TERMINAL ILEUM WAS ATTEMPTED FOR RETRIEVAL USING A STERIS RAPTOR GRASPING DEVICE, THE DEVICE MALFUNCTIONED AND BECAME EMBEDDED IN THE TERMINAL ILEUM DISTAL TO THE STENT. MULTIPLE ENDOSCOPIC MANEUVERS WERE PERFORMED TO DISENGAGE AND RETRIEVE THE DEVICE; HOWEVER, THESE WERE UNSUCCESSFUL. THE WIRES OF THE RAPTOR GRASPING DEVICE WERE CUT, THE SCOPE WAS READVANCED INTO THE TERMINAL ILEUM WHERE ADDITIONAL ATTEMPTS AT REMOVING THE DISTAL END OF THE RAPTOR WERE UNSUCCESSFUL.

    Patient problems: Foreign Body In Patient

    MAUDE report 25408783 ↗
  • MalfunctionApr 30, 2026Difficult to Open or Close

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE, THE RAPTOR GRASPING JAWS WOULD NOT OPEN DURING USE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY FOLLOWING A SHORT DELAY TO OPEN THE JAWS OF THE DEVICE. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25043414 ↗
  • MalfunctionApr 29, 2026Difficult to Open or Close

    THE DISTRIBUTOR REPORTED THAT DURING A PATIENT PROCEDURE THEIR RAPTOR GRASPING DEVICE WOULD NOT CLOSE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25035786 ↗
  • MalfunctionApr 29, 2026Difficult to Open or Close

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE, THE RAPTOR GRASPING DEVICE LOCKED DURING USE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING ANOTHER DEVICE FOLLOWING A SHORT DELAY. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25035494 ↗
  • MalfunctionJan 25, 2026Mechanical Problem

    THE USER FACILITY REPORTED THE JAWS TO THEIR RAPTOR GRASPING DEVICE "BROKE AND WOULD NOT CLOSE" DURING A PATIENT PROCEDURE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24175832 ↗
  • MalfunctionJan 15, 2026Mechanical Problem

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE THE JAWS TO TWO OF THEIR RAPTOR GRASPING DEVICES WOULD NOT CLOSE. THE PROCEDURE WAS COMPLETED USING A THIRD RAPTOR GRASPING DEVICE. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24086927 ↗
  • MalfunctionDec 11, 2025Unintended Movement; Unintended Movement

    THE USER FACILITY REPORTED VIA MEDWATCH REPORT (B)(4) THAT THE JAWS TO THEIR RAPTOR GRASPING DEVICE WOULD NOT MOVE TO THE CLOSED POSITION DURING THEIR PRE-PROCEDURE CHECK. THE PROCEDURE WAS COMPLETED WITHOUT DELAY UTILIZING ANOTHER DEVICE. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23783240 ↗
  • MalfunctionNov 28, 2025Mechanical Problem; Mechanical Problem

    THE USER FACILITY REPORTED VIA MEDWATCH MW5176915 THAT THEIR RAPTOR GRASPING DEVICE FAILED TO OPEN AND CLOSE DURING A PATIENT PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23672169 ↗
  • MalfunctionMar 27, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    THE USER FACILITY REPORTED VIA MEDWATCH MW5165926 THAT, "STASIS DISPOSABLE RAPTOR GRASPING DEVICE NEEDED FOR A CASE. PACKAGE OPENED AND UPON REVIEW, THE CLEAVER DID NOT WORK. NOT SURE IF BROKEN. DEVICE NOT USED AND ANOTHER RAPTOR GRASPING DEVICE TAKEN OFF SHELF AND FUNCTIONING AND THEN USED IN PATIENT CASE." NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21712982 ↗
  • MalfunctionMar 7, 2025Mechanical Problem; Mechanical Problem

    THE USER FACILITY REPORTED DURING A PATIENT PROCEDURE THAT THE JAWS TO TWO OF THEIR RAPTOR GRASPING DEVICES WERE NOT OPENING PROPERLY AND RESULTED IN A SHORT PROCEDURE DELAY. THE PROCEDURE WAS COMPLETED USING A THIRD RAPTOR GRASPING DEVICE. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21548051 ↗
  • MalfunctionFeb 21, 2025Insufficient Device Problem Information; Insufficient Device Problem Information

    USER FACILITY PERSONNEL STATED THAT THE PATIENT PROCEDURE WAS COMPLETED SUCCESSFULLY WITH THE RAPTOR GRASPING DEVICE. NO REPORT OF INJURY.

    Patient problems: Insufficient Information; No Clinical Signs, Symptoms or Conditions

    MAUDE report 21435870 ↗
  • MalfunctionOct 10, 2024Difficult to Open or Close; Insufficient Device Problem Information; Difficult to Open or Close; Insufficient Device Problem Information

    THE USER FACILITY REPORTED VIA MEDWATCH (B)(4) THAT THEIR RAPTOR GRASPING DEVICE WAS DEFECTIVE/NOT FUNCTIONING PROPERLY DURING A STENT REMOVAL. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20415802 ↗
  • MalfunctionJul 3, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    THE DISTRIBUTOR REPORTED THAT DURING A PATIENT PROCEDURE, THE USER FACILITY'S RAPTOR GRASPING DEVICE WOULD NOT CLOSE AND WAS REMOVED IN THE OPEN POSITION. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19664465 ↗
  • MalfunctionJun 17, 2024Insufficient Device Problem Information; Insufficient Device Problem Information

    THE USER FACILITY REPORTED THE JAWS TO THEIR RAPTOR GRASPING DEVICE WOULD NOT REOPEN DURING A PATIENT PROCEDURE. ANOTHER MODALITY WAS UTILIZED TO COMPLETE THE PROCEDURE WITH REPORT OF DELAY. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19554915 ↗
  • MalfunctionApr 17, 2024Unintended Movement; Unintended Movement

    THE USER FACILITY REPORTED THAT THEIR RAPTOR GRASPING DEVICE PARTIALLY DETACHED INSIDE OF THE PATIENT DURING PROCEDURAL USE. THE DEVICE WAS WITHDRAWN FROM THE PATIENT AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19129736 ↗
  • MalfunctionJul 14, 2023Break; Break

    THE DISTRIBUTOR REPORTED THAT DURING A STENT REMOVAL THEIR RAPTOR GRASPING DEVICES WERE BROKEN. THE EXAMINATION WAS STOPPED. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17327662 ↗
  • MalfunctionJun 29, 2023Difficult to Open or Close; Difficult to Open or Close

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE ONE JAW OF THEIR RAPTOR GRASPING DEVICE BECAME STUCK TO THE AXIOS STENT AND DID NOT FUNCTION PROPERLY. THERE WAS NO REPORT OF PROCEDURE DELAY OR INJURY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17227976 ↗
  • MalfunctionMay 4, 2023Difficult to Open or Close; Difficult to Open or Close

    THE USER FACILITY REPORTED THAT THE USER TRIED TO USE THE RAPTOR GRASPING DEVICE HOWEVER, IT DID NOT FUNCTION PROPERLY. THE JAWS ON THE DEVICE WOULD NOT CLOSE AND GOT STUCK ON A STENT AS IT WAS BEING REMOVED. ANOTHER RAPTOR GRASPING DEVICE WAS UTILIZED, AND THE PATIENT PROCEDURE WAS COMPLETED SUCCESSFULLY. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16869866 ↗
  • MalfunctionApr 20, 2023Difficult to Open or Close; Difficult to Open or Close

    THE USER FACILITY REPORTED VIA MEDWATCH REPORT (MW5115832) THAT THEIR RAPTOR GRASPING DEVICE WORKED ON THE FIRST PASS BUT WOULD NOT OPEN DURING THE SECOND PASS. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16780459 ↗
  • MalfunctionSep 9, 2022Difficult to Open or Close; Difficult to Open or Close

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE THAT INCLUDED USE OF THE RAPTOR GRASPING DEVICE, THE WIRES WITHIN THE HANDLE ACTUATOR BECAME DISLODGED, AND THE DEVICE COULD NOT BE CLOSED. THE ENDOSCOPE WAS WITHDRAWN, AND THE PROCEDURE WAS RESCHEDULED FOR A LATER DATE. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15393138 ↗
  • MalfunctionMay 26, 2022Difficult to Open or Close; Difficult to Open or Close

    THE USER FACILITY REPORTED VIA MEDWATCH # 5201030000-2022-8002 THAT DURING A PATIENT PROCEDURE THEIR RAPTOR GRASPING DEVICE WOULD NOT OPEN. USER FACILITY PERSONNEL UTILIZED AN ADDITIONAL RAPTOR GRASPING DEVICE AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO REPORT OF INJURY OR PROCEDURE DELAYS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14514849 ↗
  • MalfunctionMay 16, 2022Difficult to Open or Close; Difficult to Open or Close

    THE DISTRIBUTOR REPORTED THAT DURING A PATIENT PROCEDURE THEIR RAPTOR GRASPING DEVICE WOULD NOT CLOSE. USER FACILITY PERSONNEL UTILIZED AN ADDITIONAL RAPTOR GRASPING DEVICE AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO INJURY OR PROCEDURE DELAYS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14408366 ↗
  • MalfunctionMay 5, 2022Insufficient Device Problem Information; Insufficient Device Problem Information

    THE USER FACILITY REPORTED VIA USER FACILITY MEDWATCH REPORT MW5108030 THAT THEIR RAPTOR GRASPING DEVICE WOULD NOT WORK WHEN TAKEN OUT OF THE PACKAGE. NO REPORT OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14288355 ↗
  • MalfunctionMar 24, 2022Mechanical Problem; Mechanical Problem

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE THAT INCLUDED USE OF A RAPTOR GRASPING DEVICE THEIR FORCEPS WOULD NOT GRASP A FEEDING TUBE WITHOUT SLIDING RESULTING IN A PROCEDURE DELAY. NO REPORT OF PATIENT HARM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13889139 ↗
  • MalfunctionMar 10, 2022Defective Device

    RAPTOR GRASPING DEVICE #00711177 LOT#2200981 X2. FAULTY GARBAGE. FAULTY RIGHT OUT OF THE BAG. FDA SAFETY REPORT ID #(B)(4).

    Patient problems: Insufficient Information

    MAUDE report 13734237 ↗

Other Endoscopic grasping/cutting instrument, non-powered, exempt devices

All Endoscopic grasping/cutting instrument, non-powered, exempt devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page