Assurance 00711885Recalls & reported problems

US Endoscopy · The Assurance Clip - 18 mm is compatible with endoscopes and is indicated for clip placement within digestive tract for the purpose of mechanical pressure treatment of bleeding of small arteries and pulsation. · Class 2 · Implant

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00724995219154), received between Oct 4, 2024 and Apr 14, 2026 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Detachment of Device or Device Component1
  • Device Fell1
  • Failure to Fire1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 14, 2026Failure to Fire; Detachment of Device or Device Component

    THE USER FACILITY REPORTED WHILE USING THEIR ASSURANCE CLIP DURING A PATIENT PROCEDURE THAT THE CLIP WAS ONLY ABLE TO PARTIALLY DEPLOY. WHILE ATTEMPTING TO FULLY DEPLOY THE CLIP, THE CLIP'S INSIDE METAL HOOK DETACHED AND FELL INTO THE PATIENT. THE PIECES OF THE METAL HOOK WERE RETRIEVED AND THE PATIENT RECEIVED X-RAYS THAT CONFIRMED NO FURTHER ISSUES. THE PROCEDURE WAS COMPLETED SUCCESSFULLY. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24875516 ↗
  • MalfunctionMar 5, 2026Adverse Event Without Identified Device or Use Problem

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE A PIECE OF WIRE FROM THEIR ASSURANCE CLIP WAS PROTRUDING FOLLOWING DEPLOYMENT. THE WIRE WAS REMOVED FROM THE PATIENT TO COMPLETE THE PROCEDURE SUCCESSFULLY. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24523663 ↗
  • MalfunctionOct 4, 2024Device Fell; Device Fell

    THE USER FACILITY REPORTED THAT DURING A PATIENT PROCEDURE TWO ASSURANCE CLIPS FELL OFF THE DEVICE INTO THE PATIENT. THE CLIPS HAD TO BE RETRIEVED USING A NET. THE PROCEDURE WAS COMPLETED SUCCESSFULLY USING ANOTHER ASSURANCE CLIP. NO REPORT OF INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20368173 ↗

Other Hemostatic metal clip for the GI tract devices