Skip to content

Welch Allyn, Inc. 901029 (REF VS100)Recalls & reported problems

Welch Allyn, Inc. · VS100 UDI BASE UNIT · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
11
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00732094214154), received between May 31, 2023 and Aug 18, 2026 — classified as 11 malfunctions.

The most frequently reported specific device problem is “Material Split, Cut or Torn” (9 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 1 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Feb 18, 2026Class IIOpen, Classified

    Display screen may unintentionally flicker due to a software issue. Exposure to flickering lights may induce photosensitive seizure activity in susceptible individuals.

    Z-1573-2026 · Root cause: Device Design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Material Split, Cut or Torn9
  • Material Frayed2
  • Sparking2
  • Explosion1
  • Fire1

Patient problems

  • No Clinical Signs, Symptoms or Conditions11

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 18, 2026Material Split, Cut or Torn

    Baxter received a report stating that a VS100 Spot Vision Screener had a damaged power cord, it appeared to be peeling. This occurred before use. There was no patient injury or death reported with this event.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50291764 ↗
  • MalfunctionJul 17, 2025Fire; Explosion; Fire; Explosion

    CUSTOMER REPORTED THE BATTERY HAD EXPLODED SEVERAL TIMES AND THE DEVICE HAD CAUGHT ON FIRE. THERE WAS NO ADVERSE EVENT REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22529989 ↗
  • MalfunctionJul 2, 2025Sparking; Material Split, Cut or Torn; Sparking; Material Split, Cut or Torn

    CUSTOMER REPORTED POWER SUPPLY FOR THIS DEVICE HAD EXPOSED WIRING ON THE CORNER THAT WAS GOING FROM THE BANK TO THE DEVICE AND WAS SPARKING. THERE WAS NO INJURY REPORTED. THIS INCIDENT WAS CAPTURED UNDER BAXTER COMPLAINT REF # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22385401 ↗
  • MalfunctionJun 19, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    CUSTOMER REPORTED PHYSICAL DAMAGE WITH EXPOSED COPPER WIRINGS. THERE WAS NO INJURY REPORTED. THIS INCIDENT WAS CAPTURED UNDER BAXTER COMPLAINT REF #(B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22289205 ↗
  • MalfunctionMay 12, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    CUSTOMER REPORTED POWER CORD IS FRAYED AND SOME WIRES ARE GETTING EXPOSED. THERE WAS NO INJURY REPORTED. THIS INCIDENT WAS CAPTURED UNDER HILLROM COMPLAINT REF # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22002479 ↗
  • MalfunctionApr 3, 2025Sparking; Material Split, Cut or Torn; Sparking; Material Split, Cut or Torn

    CUSTOMER REPORTED PHYSICAL DAMAGE. WIRING ON THE POWER SUPPLY WAS EXPOSED BETWEEN THE ADAPTOR BOX AND THE PLUG INTO THE DEVICE AND WAS SPARKING. THERE WAS NO INJURY REPORTED. THIS INCIDENT WAS CAPTURED UNDER HILLROM COMPLAINT REF # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21761326 ↗
  • MalfunctionMar 28, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    CUSTOMER REPORTED POWER CORD HAS FRAYED, EXPOSED WIRES. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21717684 ↗
  • MalfunctionFeb 11, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    CUSTOMER REPORTED DEVICE WILL NOT HOLD CHARGE BECAUSE THE POWER CORD IS BROKEN AND HAS FRAYED WIRES. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21355114 ↗
  • MalfunctionJan 23, 2025Material Split, Cut or Torn; Material Split, Cut or Torn

    THE BAXTER TECHNICIAN FOUND EXPOSED WIRES ON THE POWER SUPPLY DURING VISUAL INSPECTION. THIS WAS A SECONDARY FINDING.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21216925 ↗
  • MalfunctionSep 30, 2024Material Frayed; Material Split, Cut or Torn; Material Frayed; Material Split, Cut or Torn

    THE CUSTOMER REPORTED THAT THE VS100 POWER CORD WAS FRAYED. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT # (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20333363 ↗
  • MalfunctionMay 31, 2023Material Frayed; Material Frayed

    THE CUSTOMER REPORTED THE VS100 WOULD NOT HOLD A CHARGE WHEN UNPLUGGED. THE DEVICE WAS RETURNED FOR BENCH REPAIR WHERE THE TECHNICIAN DETERMINED THAT THE POWER CORD WAS FRAYED WITH EXPOSED WIRES. THIS INCIDENT WAS CAPTURED UNDER HILLROM COMPLAINT REF (B)(4).

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17032078 ↗

Other Refractometer, ophthalmic devices

All Refractometer, ophthalmic devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page