Skip to content

GORE EXCLUDER AAA Endoprosthesis PXA230300Recalls & reported problems

W. L. Gore & Associates, Inc. · Aortic Extender · Class 3 · Implant

All models of GORE EXCLUDER AAA Endoprosthesis (81)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
1
Injuries reported
5
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00733132609871), received between Sep 17, 2018 and Feb 9, 2024 — classified as 1 death, 5 injuries.

The most frequently reported specific device problem is “Patient-Device Incompatibility” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (P020004); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 1 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Patient-Device Incompatibility2
  • Adverse Event Without Identified Device or Use Problem1
  • Improper or Incorrect Procedure or Method1
  • Insufficient Device Problem Information1
  • Migration1

Patient problems

  • Aneurysm3
  • Aneurysm expansion1
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Clinical Endoleak1
  • Death1
  • Fistula1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • InjuryFeb 9, 2024Patient-Device Incompatibility; Patient-Device Incompatibility

    THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON (B)(6) 2013, THIS PATIENT UNDERWENT AN ENDOVASCULAR TREATMENT FOR AN ABDOMINAL AORTIC ANEURYSM USING GORE® EXCLUDER® AAA ENDOPROSTHESES. AN AORTIC EXTENDER ENDOPROSTHESIS WAS IMPLANTED PROXIMALLY AND A CONTRALATERAL LEG ENDOPROSTHESIS WAS IMPLANTED ON THE RIGHT SIDE. ON AN UNKNOWN DATE, ENLARGEMENT OF THE ANEURYSM DUE TO A PROXIMAL TYPE I ENDOLEAK AND A RIGHT DISTAL TYPE I ENDOLEAK WAS OBSERVED. REPORTEDLY, THE TREND OF ANEURYSMAL ENLARGEMENT BECAME MORE PRONOUNCED ABOUT 3 YEARS AGO. ON (B)(6) 2024, A REINTERVENTION WAS PERFOMED. AN ADDITIONAL STENT GRAFT WAS IMPLANTED PROXIMALLY. ALSO, AN ADDITIONAL STENT GRAFT WAS IMPLANTED TO EXTEND TO THE RIGHT EXTERNAL ILIAC ARTERY AFTER EMBOLIZATION OF THE RIGHT INTERNAL ILIAC ARTERY. ANGIOGRAPHY CONFIRMED NO ENDOLEAK AND THE PROCEDURE WAS COMPLETED. THE PATIENT TOLERATED THE PROCEDURE.

    Patient problems: Aneurysm expansion; Clinical Endoleak

    MAUDE report 18674403 ↗
  • InjuryOct 4, 2021Migration; Migration

    THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON (B)(6) 2011, A PATIENT UNDERWENT ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM USING GORE® EXCLUDER® AAA ENDOPROSTHESIS. PXA230300 WAS IMPLANTED DISTALLY IN THE LEFT LEG. ON UNKNOWN DATE, AT A FOLLOW-UP, AN ENLARGEMENT OF A LEFT COMMON ILIAC ARTERY WAS OBSERVED, LEADING TO A MIGRATION OF THE STENT GRAFT TO DISTAL. ON (B)(6) 2021, REINTERVENTION WAS PERFORMED. LEFT INTERNAL ILIAC ARTERY WAS EMBOLIZED USING COIL. PLC141400J WAS IMPLANTED FROM THE FLOW DIVIDER AND PLA121400J WAS IMPLANTED TO EXTEND TO A LEFT INTERNAL ILIAC ARTERY. THE PATIENT TOLERATED THE PROCEDURE. THE PHYSICIAN STATED THAT IT HAD BEEN TEN YEARS SINCE THE INITIAL TREATMENT, THIS EVENT OCCURRED AS AN INEVITABLE.

    Patient problems: Aneurysm

    MAUDE report 12568061 ↗
  • InjurySep 1, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ON (B)(6) 2007, THE PATIENT WAS IMPLANTED WITH A GORE® EXCLUDER® AAA ENDOPROSTHESES TO TREAT AN ABDOMINAL AORTIC ANEURYSM. THE PATIENT TOLERATED THE PROCEDURE. IT WAS REPORTED THAT ANGIOGRAM IMAGES, DATED (B)(6) 2021, REVEALED A POSSIBLE PROXIMAL TYPE I ENDOLEAK. IT IS UNKNOWN IF THERE WAS ANEURYSM SAC ENLARGEMENT. ON (B)(6) 2021 THERE WAS A REINTERVENTION. THE PHYSICIAN IMPLANTED A GORE® EXCLUDER® AAA ENDOPROSTHESIS AORTIC EXTENDER IN THE PROXIMAL SEAL ZONE FROM THE ORIGINAL PROCEDURE. THE ENDOLEAK WAS RESOLVED. THE PATIENT TOLERATED THE REINTERVENTION PROCEDURE.

    Patient problems: Aneurysm

    MAUDE report 12411554 ↗
  • InjuryJul 21, 2021Insufficient Device Problem Information; Insufficient Device Problem Information

    ON (B)(6) 2009, THE PATIENT UNDERWENT ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM USING A GORE¿ EXCLUDER¿ AAA ENDOPROSTHESIS. ON AN UNKNOWN DATE, FOLLOW-UP EXAMINATION REPORTEDLY REVEALED A SUSPECTED TYPE IA ENDOLEAK. THE CAUSE OF THE SUSPECTED ENDOLEAK WAS NOT REPORTED. IT WAS UNKNOWN WHETHER ANEURYSM ENLARGEMENT WAS PRESENT. ON (B)(6) 2021, THE PATIENT UNDERWENT A REINTERVENTION PROCEDURE WHEREBY AN ADDITIONAL STENT GRAFT WAS DEPLOYED TO EXTEND THE DEVICE PROXIMALLY. NO ENDOLEAK WAS OBSERVED ON FINAL PROCEDURAL ANGIOGRAPH. THE PATIENT TOLERATED THE PROCEDURE.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 12203357 ↗
  • InjuryFeb 28, 2020Patient-Device Incompatibility; Patient-Device Incompatibility

    FOLLOWING WAS REPORTED TO GORE. ON (B)(6) 2011, A PATIENT UNDERWENT A TREATMENT OF DISSECTION OF THE DESCENDING AORTA WITH GORE® EXCLUDER® AAA ENDOPROSTHESIS AND GORE® TAG® THORACIC ENDOPROSTHESIS. TWO AORTIC EXTENDER COMPONENT (PXA230300) WERE PLACED FOLLOWED BY A TAG (TGT2815). THE PROCEDURE WAS COMPLETED WITHOUT ANY ISSUES. THE PATIENT TOLERATED THE PROCEDURE. ON (B)(6) 2016, A FOLLOW UP IMAGE REVEALED THE ANEURYSM DIAMETER WAS 50MM. ON AN UNKNOWN DATE, A FOLLOW UP IMAGE REVEALED DISTAL TYPE I ENDOLEAK AT THE DISTAL END OF ONE OF THE TWO AORTIC EXTENDER ENDOPROSTHESIS (EXACT LOCATION IS UNKNOWN), AND THE ANEURYSM DIAMETER MEASURED TO BE 55MM. ON (B)(6) 2020, A REINTERVENTION WAS PERFORMED TO TREAT THE ENDOLEAK. A TGM343415 WAS PLACED AT THE DISTAL SIDE OF ONE OF THE TWO AORTIC EXTENDER ENDOPROSTHESIS, TO RIGHT ABOVE THE CELIAC ARTERY. FINAL ANGIOGRAPHY REVEALED NO ENDOLEAK, AND THE PROCEDURE WAS COMPLETED. THE PATIENT TOLERATED THE PROCEDURE.

    Patient problems: Aneurysm

    MAUDE report 9770999 ↗
  • DeathSep 17, 2018Improper or Incorrect Procedure or Method; Improper or Incorrect Procedure or Method

    ON AN UNKNOWN DATE, THE PATIENT UNDERWENT TREATMENT OF A RUPTURED ABDOMINAL AORTIC ANEURYSM WITH GORE® EXCLUDER® AAA ENDOPROSTHESES. THREE 23 MM AORTIC EXTENDER COMPONENTS (PXA230300J/ LOT NUMBERS UNKNOWN) WERE IMPLANTED JUST ABOVE THE AORTIC BIFURCATION. THE DIAMETER OF THE INFRARENAL AORTA WAS A REPORTED 17-18 MM. THE PATIENT TOLERATED THE PROCEDURE. APPROXIMATELY FOUR MONTHS LATER, ON UNKNOWN DATE, THE PATIENT WAS ADMITTED TO THE HOSPITAL FOR ABDOMINAL PAIN, BLACK STOOL AND ANEMIA. AN ENDOSCOPY SHOWED AN IRREGULAR FORMED TUMOR MASS WITH BLEEDING IN THE DUODENUM. A BIOPSY WAS PERFORMED, AND IGG4-RELATED DISEASE WAS SUSPECTED. A COMPUTED TOMOGRAPHY SCAN INDICATED THE AORTA HAD BEEN ENLARGED (SIZE UNKNOWN) AT THE REGION WHERE THE DEVICES WERE IMPLANTED. NO ENDOLEAK WAS OBSERVED. THE PATIENT WAS ADMINISTERED MEDICATION (PROTON PUMP INHIBITOR AND IRON) AS A CONSERVATIVE MEDICAL TREATMENT, AND DISCHARGED FROM THE HOSPITAL REPORTEDLY HAVING SHOWN IMPROVEMENT. ONE MONTH LATER, ON UNKNOWN DATE, THE PATIENT WAS ADMITTED TO THE HOSPITAL FOR BLEEDING. THE PATIENT REPORTEDLY EXPIRED OF HEMORRHAGIC SHOCK DUE TO SEVERE MELENA. AN AUTOPSY WAS PERFORMED AND AN AORTO-DUODENAL FISTULA WAS CONFIRME

    Patient problems: Death; Fistula

    MAUDE report 7881579 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P020004. One submission often covers many models, so check the product and lot details in each record.

  • Z-0996-2020 · Jan 6, 2020 · Terminated
    GORE EXCLUDER Iliac Branch Endoprosthesis (IBE)
  • Z-0995-2020 · Jan 6, 2020 · Terminated
    GORE EXCLUDER AAA Endoprosthesis

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices

All System, Endovascular Graft, Aortic Aneurysm Treatment devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page