The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00733132609895), received between Feb 18, 2019 and Dec 18, 2023 — classified as 6 injuries.
The most frequently reported specific device problem is “Human-Device Interface Problem” (1 report).
2 other recalls cover devices cleared under the same FDA submission (P020004); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem5
- Human-Device Interface Problem1
Patient problems
- Aneurysm5
- Aneurysm expansion1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryDec 18, 2023Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON (B)(6) 2013, THIS PATIENT UNDERWENT AN ENDOVASCULAR TREATMENT FOR ABDOMINAL AORTIC ANEURYSM AND COMMON ILIAC ARTERY ANEURYSMS USING GORE® EXCLUDER® AAA ENDOPROSTHESES. ALL THE PLANNED DEVICES WERE IMPLANTED SUCCESSFULLY. CONSIDERING THE DIAMETER OF COMMON ILIAC ARTERIES (RIGHT 16 ¿ 22 MM, LEFT 16 - 22 MM), AORTIC EXTENDER COMPONENTS (RIGHT PXA280300, LEFT PXA260300) WERE IMPLANTED IN THE DISTAL END OF BILATERAL SIDES. THE PATIENT TOLERATED THE PROCEDURE. ON (B)(6) 2023, A REINTERVENTION WAS PERFORMED AS LEFT COMMON ILIAC ANEURYSM ENLARGEMENT WAS FOUND. GORE® EXCLUDER® ILIAC BRANCH ENDOPROSTHESIS (IBE) WAS PLANNED TO BE PLACED IN THE LEFT ILIAC ARTERY. PRIOR TO STARTING THIS REINTERVENTION, THE FSA NOTICED EVIDENCE FROM THE CT IMAGES THAT ANOTHER REINTERVENTION HAD BEEN PERFORMED ON THE RIGHT DISTAL SIDE AS WELL. IT APPEARED THAT AN ADDITIONAL STENT GRAFT HAD BEEN PLACED IN THE RIGHT SIDE TOO AFTER THE INITIAL PROCEDURE (TIMING AND REASON UNKNOWN). DURING THE REINTERVENTION ON (B)(6) 2023, THE ILIAC BRANCH COMPONENT (IBC) WAS SUCCESSFULLY DEPLOYED; HOWEVER, DELIVERY OF THE INTERNAL ILIAC COMPONENT (IIC) TO THE INTERNAL ILIAC ARTERY
Patient problems: Aneurysm expansion
MAUDE report 18352425 ↗InjuryJul 14, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2011, THIS PATIENT WAS IMPLANTED WITH GORE®EXCLUDER®AAA ENDOPROSTHESES TO TREAT AN ABDOMINAL AORTIC ANEURYSM. ON (B)(6) 2011, COMPUTED TOMOGRAPHY (CT) SCAN REVEALED A PROXIMAL TYPE I ENDOLEAK. ON (B)(6) 2011, THE PATIENT UNDERWENT REINTERVENTION AND AN AORTIC EXTENDER COMPONENT WAS IMPLANTED TO TREAT THE ENDOLEAK. THE PATIENT TOLERATED THE PROCEDURE AND THE ENDOLEAK WAS RESOLVED. ON (B)(6) 2021, COMPUTED TOMOGRAPHY (CT) SCAN FOR A SUSPECTED PROXIMAL TYPE I ENDOLEAK. ON (B)(6) 2021, THE PATIENT UNDERWENT REINTERVENTION AND AN AORTIC EXTENDER COMPONENT WAS IMPLANTED TO TREAT THE ENDOLEAK. ANGIOGRAPHY AT THIS TIME DETERMINED THE ANEURYSM HAD ENLARGED TO 87MM HOWEVER, THE PROXIMAL TYPE I WAS UNABLE TO BE CONFIRMED. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: Aneurysm
MAUDE report 12170765 ↗InjuryMar 22, 2021Human-Device Interface Problem; Human-Device Interface Problem
ON (B)(6) 2021, THIS PATIENT UNDERWENT REINTERVENTION FOR TREATMENT OF A PROXIMAL TYPE I ENDOLEAK. A GORE® AORTIC EXTENDER COMPONENT WAS PLACED PROXIMALLY TO EXTEND COVERAGE OF A PREVIOUSLY IMPLANTED GORE® AORTIC EXTENDER COMPONENT. ADDITIONALLY ENDOANCHORS WERE SCREWED IN TO OBTAIN SECURE TRANSMURAL FIXATION. THE PATIENT TOLERATED THE PROCEDURE AND THE ENDOLEAK WAS REDUCED BUT NOT FULLY RESOLVED.
Patient problems: Aneurysm
MAUDE report 11548072 ↗InjuryAug 2, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2013, A PATIENT UNDERWENT ENDOVASCULAR TREATMENT OF ABDOMINAL AORTIC ANEURYSMS USING GORE® EXCLUDER® AAA ENDOPROSTHESES. ON AN UNKNOWN DATE, ENLARGEMENT OF THE ANEURYSM DIAMETER (AMOUNT UNKNOWN) AND A DISTAL TYPE I ENDOLEAK FROM THE LEFT LEG WERE NOTED. IT WAS CONFIRMED THAT THERE WAS LACK OF APPOSITION OF PXA280300/10067062 AT THE INNER CURVATURE OF THE PROXIMAL AORTIC NECK. ON (B)(6) 2020, RE-INTERVENTION WAS PERFORMED. DURING THE OPERATION, A PROXIMAL TYPE I ENDOLEAK WAS NOT IDENTIFIED, HOWEVER AN ADDITIONAL STENT GRAFT WAS IMPLANTED IN THE PROXIMAL NECK TO SOLVE THE LACK OF APPOSITION. THE PHYSICIAN COMMENTED THAT THE PROXIMAL NECK OF THE PATIENT WAS ANGLED, SO THE LACK OF APPOSITION IN THE PROXIMAL SIDE WAS CONSIDERED TO BE DUE TO THE PATIENT¿S ANATOMY.
Patient problems: Aneurysm
MAUDE report 10354254 ↗InjuryJun 29, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2012, THIS PATIENT UNDERWENT ENDOVASCULAR TREATMENT FOR AN ABDOMINAL AORTIC ANEURYSM AND WAS IMPLANTED WITH GORE® EXCLUDER® AAA ENDOPROSTHESIS FEATURING C3® DELIVERY SYSTEM. A GORE® AORTIC EXTENDER WAS PLACED AS THE MOST PROXIMAL DEVICE TO EXCLUDE COVERAGE OF THE TRUNK IPSILATERAL LEG COMPONENT. THE PATIENT TOLERATED THE PROCEDURE. ON (B)(6) 2020, FOLLOW-UP IMAGING SHOWS WHAT APPEARS TO BE A PROXIMAL TYPE I ENDOLEAK. THE PHYSICIAN PLANS TO REINTERVENE TO PLACE AN ADDITIONAL AORTIC EXTENDER COMPONENT PROXIMALLY; BUT HAS NOT YET SCHEDULED A DATE. THERE WAS NO ANEURYSM ENLARGEMENT OR DEVICE MIGRATION REPORTED. THE CAUSE OF THE ENDOLEAK IS UNKNOWN.
Patient problems: Aneurysm
MAUDE report 10210059 ↗InjuryFeb 18, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE FOLLOWING INFORMATION WAS REPORTED TO GORE THROUGH THE GLOBAL REGISTRY FOR ENDOVASCULAR AORTIC TREATMENT (GREAT 10-11): ON (B)(6) 2014 A PATIENT WAS IMPLANTED WITH A GORE EXCLUDER AAA ENDOPROSTHESIS AORTIC EXTENDER COMPONENT AS A REINTERVENTION TO A PRIOR ENDOVASCULAR PROCEDURE. ON (B)(6) 2018, A REINTERVENTION OCCURRED WHEREBY A TRANSLUMBAR EMBOLIZATION WITH ONYX WAS PERFORMED TO TREAT A TYPE II ENDOLEAK. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: Aneurysm
MAUDE report 8348617 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P020004. One submission often covers many models, so check the product and lot details in each record.
- Z-0996-2020 · Jan 6, 2020 · TerminatedGORE EXCLUDER Iliac Branch Endoprosthesis (IBE)
- Z-0995-2020 · Jan 6, 2020 · TerminatedGORE EXCLUDER AAA Endoprosthesis
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page