GORE EXCLUDER AAA Endoprosthesis PXL161007Recalls & reported problems
W. L. Gore & Associates, Inc. · Iliac Extender · Class 3 · Implant
All models of GORE EXCLUDER AAA Endoprosthesis (81)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00733132610068), received between Oct 18, 2019 and Apr 29, 2025 — classified as 3 injuries.
The most frequently reported specific device problem is “Device Endoleak” (1 report).
2 other recalls cover devices cleared under the same FDA submission (P020004); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Device Endoleak1
Patient problems
- Aneurysm1
- Clinical Endoleak1
- Thrombosis/Thrombus1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryApr 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON (B)(6) 2012, THIS PATIENT UNDERWENT AN ENDOVASCULAR TREATMENT FOR ABDOMINAL AORTIC ANEURYSM USING GORE® EXCLUDER® AAA ENDOPROSTHESIS. DURING A FOLLOW-UP EXAMINATION, DISTAL MIGRATION (APPROX. 1.5 CM) OF THE TRUNK-IPSILATERAL LEG WAS NOTED. THEREFORE, A REINTERVENTION WAS INDICATED, THOUGH NO ENDOLEAK WAS SHOWN. ON (B)(6) 2025, AN ADDITIONAL PROCEDURE WAS PERFORMED. TWO AORTIC EXTENDERS WERE IMPLANTED PROXIMAL TO THE TRUNK-IPSILATERAL LEG. DURING THE PROCEDURE, IT WAS FOUND THAT THROMBUS FORMATION INSIDE THE CONTRALATERAL LEG AND ILIAC EXTENDER ON THE RIGHT SIDE. THROMBUS ASPIRATION WAS PERFORMED THROUGH THE INTRODUCER SHEATH AND AN ADDITIONAL CONTRALATERAL LEG WAS PLACED TO FIX THE REMAINING THROMBI. REPORTEDLY, THE PHYSICIAN FORGOT TO ADMINISTRATE HEPARIN BEFORE DEVICE INSERTION AND HE NOTICED IT AFTER INSERTING THE INTRODUCER SHEATH. THE PHYSICIAN THOUGHT IT WAS A CAUSE OF THROMBUS FORMATION ON THE RIGHT SIDE. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: Thrombosis/Thrombus
MAUDE report 21931831 ↗InjuryJul 2, 2024Device Endoleak; Device Endoleak
IT WAS REPORTED THAT ON (B)(6) 2015, A PATIENT PRESENTING WITH A RIGHT COMMON ILIAC ANEURYSM WAS TREATED WITH A GORE EXCLUDER AAA ENDOPROSTHESIS. ON (B)(6) 2018, THE PATIENT PRESENTED WITH A TYPE IIIA ENDOLEAK. ON (B)(6) 3018, AN ADDITIONAL CONTRALATERAL LEG WAS IMPLANTED TO TREAT THE ENDOLEAK.
Patient problems: Clinical Endoleak
MAUDE report 19654654 ↗InjuryOct 18, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2012, THIS PATIENT UNDERWENT AN ENDOVASCULAR REPAIR FOR ABDOMINAL AORTIC ANEURYSM USING A GORE® EXCLUDER® AAA ENDOPROSTHESIS. OP NOTES PROVIDED FOR CURRENT REINTERVENTION REVEALED THERE WAS A REINTERVENTION IN EARLY 2019. THE PHYSICIAN EXTENDED THE LEFT LIMB OF THE ENDOPROSTHESIS (IT IS UNKNOWN WHICH DEVICE WAS IN THE LEFT LIMB AND REINTERVENED ON) INTO THE EXTERNAL ILIAC ARTERY TO PROVIDE BETTER SEAL AND DID A TRANSCAVAL COILING WHERE FLOW CHANNEL WAS ACCESSED AND TREATED. THE PATIENT TOLERATED THIS REINTERVENTION. THERE WAS ANEURYSM SAC ENLARGEMENT WITH AN UNKNOWN CAUSE. IT WAS REPORTED THAT IMAGES (DATE AND MODALITY IS UNKNOWN) REVEALED A POSSIBLE TYPE II ENDOLEAK AND ANEURYSM SAC ENLARGEMENT. ON (B)(6) 2019, THERE WAS A REINTERVENTION. THE PHYSICIAN CHOSE TO RELINE THE GORE® EXCLUDER® AAA ENDOPROSTHESIS AND ANCHOR PROXIMALLY. THE PATIENT TOLERATED THE REINTERVENTION.
Patient problems: Aneurysm
MAUDE report 9207763 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P020004. One submission often covers many models, so check the product and lot details in each record.
- Z-0995-2020 · Jan 6, 2020 · TerminatedGORE EXCLUDER AAA Endoprosthesis
- Z-0996-2020 · Jan 6, 2020 · TerminatedGORE EXCLUDER Iliac Branch Endoprosthesis (IBE)
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page