GORE EXCLUDER AAA Endoprosthesis PXT311413Recalls & reported problems
W. L. Gore & Associates, Inc. · Class 3 · Implant
All models of GORE EXCLUDER AAA Endoprosthesis (81)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00733132610334), received between Sep 4, 2018 and Jun 9, 2020 — classified as 4 injuries.
The most frequently reported specific device problem is “Patient-Device Incompatibility” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (P020004); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Patient-Device Incompatibility2
- Adverse Event Without Identified Device or Use Problem1
- Insufficient Device Problem Information1
Patient problems
- Aneurysm3
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryJun 9, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2010, THIS PATIENT UNDERWENT TREATMENT FOR AN ABDOMINAL AORTIC ANEURYSM AND WAS IMPLANTED WITH GORE® EXCLUDER® AAA ENDOPROSTHESES. IT WAS REPORTED THAT THE PROXIMAL PLACEMENT OF THE TRUNK IPSILATERAL LEG COMPONENT WAS LANDED APPROXIMATELY 2 CM BELOW THE LOWEST (ACCESSORY) RENAL ARTERY. ON (B)(6) 2020, THE PATIENT UNDERWENT REINTERVENTION DUE TO EXPANSION OF THE ANEURYSM SAC AND A GORE® AORTIC EXTENDER COMPONENT WAS IMPLANTED TO RELINE THE PROXIMAL END. THE CAUSE OF THE ENLARGEMENT WAS NOT DETERMINED; A TYPE II ENDOLEAK WAS SUSPECTED HOWEVER NO VISIBLE ENDOLEAK WAS DETERMINED BY CONTRAST IMAGING AT THE TIME. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: Aneurysm
MAUDE report 10136295 ↗InjuryMar 12, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
ON (B)(6) 2013 A TEVAR PROCEDURE WAS PERFORMED WHEREBY A GORE® TAG® THORACIC ENDOPROSTHESIS WAS IMPLANTED TO BRIDGE A PREVIOUSLY PERFORMED ELEPHANT TRUNK PROCEDURE AND A PREVIOUSLY IMPLANTED GORE® TAG® THORACIC ENDOPROSTHESIS. FOLLOWING THE TEVAR PROCEDURE, THE PATIENT UNDERWENT ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM USING GORE® EXCLUDER® AAA ENDOPROSTHESES. ON (B)(6) 2020 A THORACOABDOMINAL ANEURYSMORRHAPHY WAS PERFORMED TO PREVENT AN IMPENDING RUPTURE OF THE ABDOMINAL AORTIC ANEURYSM DUE TO ANEURYSM ENLARGEMENT. REPORTEDLY THE AMOUNT OF ANEURYSM ENLARGEMENT IS UNAVAILABLE. NO SUSPECTED CAUSE OF ANEURYSM ENLARGEMENT WAS REPORTED. THERE WERE NO FURTHER REPORTED ISSUES IN RELATION TO THE GORE® EXCLUDER® AAA ENDOPROSTHESES. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: Aneurysm
MAUDE report 9826176 ↗InjuryNov 9, 2018Patient-Device Incompatibility; Patient-Device Incompatibility
ON (B)(6) 2012, THE PATIENT UNDERWENT ENDOVASCULAR REPAIR OF AN 44MM ABDOMINAL AORTIC ANEURYSM USING GORE® EXCLUDER® AAA ENDOPROSTHESES. THE PATIENT TOLERATED THE PROCEDURE. ON (B)(6) 2018, A FOLLOW-UP STUDY REVEALED THAT THE DIAMETER OF THE ANEURYSM SAC WAS 57MM. HOWEVER, NO OBVIOUS ENDOLEAK WAS OBSERVED. THE PHYSICIAN CONSIDERED A POSSIBLE TYPE III ENDOLEAK ORIGINATING FROM THE GRAFT TEAR AND PLACED ADDITIONAL STENT GRAFTS IN THE ENTIRE ORIGINAL STENT GRAFTS. THE PHYSICIAN REPORTEDLY SUSPECTS THE ISSUE MAY HAVE BEEN A TYPE II ENDOLEAK AS OPPOSED TO A TYPE III ENDOLEAK. HOWEVER, TO REMOVE THE POSSIBLE TYPE III ENDOLEAK, THE PHYSICIAN DECIDED TO PLACE ADDITIONAL STENT GRAFTS TO RELINE THE PRE-EXISTING SYSTEM. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: Aneurysm
MAUDE report 8055845 ↗InjurySep 4, 2018Patient-Device Incompatibility; Patient-Device Incompatibility
ON (B)(6) 2010, THIS PATIENT UNDERWENT ENDOVASCULAR REPAIR OF AN ABDOMINAL AORTIC ANEURYSM USING GORE® EXCLUDER® AAA ENDOPROSTHESES. ON (B)(6) 2018, DURING FOLLOW UP IT WAS DISCOVERED THAT, DUE TO DISEASE PROGRESSION, THE PROXIMAL NECK OF THE TRUCK-IPSILATERAL LEG COMPONENT HAD LOST WALL APPOSITION AND WAS NOW IN THE ANEURYSM SAC. ANEURYSM ENLARGEMENT WAS NOT REPORTED. ON THE SAME DAY, A REINTERVENTION OCCURRED WHEREBY THE PREVIOUSLY IMPLANTED DEVICES WERE RELINED USING GORE® EXCLUDER® AAA ENDOPROSTHESES. IN ADDITION TWO GORE® VIABAHN ENDOPROSTHESES WERE IMPLANTED IN THE RENAL ARTERIES. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7842856 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P020004. One submission often covers many models, so check the product and lot details in each record.
- Z-0996-2020 · Jan 6, 2020 · TerminatedGORE EXCLUDER Iliac Branch Endoprosthesis (IBE)
- Z-0995-2020 · Jan 6, 2020 · TerminatedGORE EXCLUDER AAA Endoprosthesis
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page