GORE EXCLUDER AAA Endoprosthesis with C3 Delivery RMT231214Recalls & reported problems
W. L. Gore & Associates, Inc. · Class 3 · Implant
All models of GORE EXCLUDER AAA Endoprosthesis with C3 Delivery (51)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00733132610518), received between Jan 15, 2019 and Sep 18, 2024 — classified as 7 injuries.
The most frequently reported specific device problem is “Patient-Device Incompatibility” (3 reports).
2 other recalls cover devices cleared under the same FDA submission (P020004); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Patient-Device Incompatibility3
- Adverse Event Without Identified Device or Use Problem2
- Insufficient Device Problem Information2
- Fluid/Blood Leak1
- Microbial Contamination of Device1
- Migration1
Patient problems
- Aneurysm expansion2
- Aneurysm1
- Clinical Endoleak1
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
- No Code Available1
- Pain1
- Pseudoaneurysm1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjurySep 18, 2024Fluid/Blood Leak; Fluid/Blood Leak
ON (B)(6) 2014, THIS PATIENT UNDERWENT ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM MEASURING 53MM IN DIAMETER AND WAS IMPLANTED WITH GORE® EXCLUDER® AAA ENDOPROSTHESIS FEATURING C3® DEPLOYMENT SYSTEM. THE PATIENT TOLERATED THE PROCEDURE AND WAS DISCHARGED FROM THE HOSPITAL ON (B)(6) 2014. ON (B)(6) 2014, FOLLOW-UP COMPUTED TOMOGRAPHY DETERMINED THE MAXIMUM DIAMETER OF THE ANEURYSM MEASURED 55MM. ON (B)(6) 2024, THE PATIENT WAS REPORTED TO HAVE A TYPE II ENDOLEAK RELATED TO THE AORTIC DEVICE OR PROCEDURE. FOLLOW-UP ULTRASOUND DETERMINED THE MAXIMUM DIAMETER OF THE ANEURYSM MEASURED 68 MM. THE CAUSE IS BELIEVED TO HAVE BEEN DUE TO DEVICE MIGRATION, COUPLED WITH THE PATIENT BEING REPORTED TO HAVE A SHORT AORTIC NECK. ON (B)(6) 2024, THE PATIENT IS SCHEDULED TO UNDERGO REINTERVENTION WHEREIN IR EMBOLIZATION OF THE TYPE II ENDOLEAK WITH ANESTHESIA IS TO BE PERFORMED.
Patient problems: Aneurysm expansion
MAUDE report 20251924 ↗InjuryJul 10, 2023Patient-Device Incompatibility; Insufficient Device Problem Information; Patient-Device Incompatibility; Insufficient Device Problem Information
IT WAS REPORTED THAT ON (B)(6) 2011, A PATIENT PRESENTING WITH AN ABDOMINAL AORTIC ANEURYSM WAS TREATED WITH A GORE® EXCLUDER® AAA ENDOPROSTHESIS. ON (B)(6) 2011, THE MAXIMUM DIAMETER OF THE AORTIC ANEURYSM MEASURED 47 MM VIA COMPUTED TOMOGRAPHY ANGIOGRAPHY (CTA) IMAGING. ON (B)(6) 2016, THE MAXIMUM DIAMETER OF THE AORTIC ANEURYSM MEASURED 44 MM VIA CTA IMAGING. ON (B)(6) 2016, THE MAXIMUM DIAMETER OF THE AORTIC ANEURYSM MEASURED 59 MM VIA CTA IMAGING. ON (B)(6) 2022, THE PATIENT WAS TREATED WITH THE IMPLANTATION OF AN ADDITIONAL CONTRALATERAL LEG ENDOPROSTHESIS IN THE RIGHT COMMON ILIAC ARTERY IN RESPONSE TO A RISK OF A TYPE IB ENDOLEAK. THE PHYSICIAN BELIEVES THAT A SHORT LANDING ZONE LESS THAN 1 CM CONTRIBUTED TO THE ENDOLEAK. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: Aneurysm expansion; Clinical Endoleak; Insufficient Information; No Clinical Signs, Symptoms or Conditions
MAUDE report 17285642 ↗InjuryApr 7, 2023Patient-Device Incompatibility; Migration; Patient-Device Incompatibility; Migration
InjuryOct 5, 2020Microbial Contamination of Device; Adverse Event Without Identified Device or Use Problem; Microbial Contamination of Device; Adverse Event Without Identified Device or Use Problem
THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON (B)(6) 2013 A PATIENT WAS UNDERGOING TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM WITH GORE® EXCLUDER® AAA ENDOPROSTHESES. ON (B)(6) 2018 THE PATIENT REPORTEDLY HAD A RIGHT GROIN PSEUDOANEURYSM WITH INFECTED PTFE GRAFTS. ON (B)(6) 2018 THE PATIENT UNDERWENT EXTERNAL ILIAC TO MID PTFE GRAFT BYPASS; TRIPLE LIGATION OF FEMORAL VESSELS AND CLOSURE OF THE PSEUDOANEURYSM. REPORTED SEPARATELY SEE PSTS EVENT# (B)(4) MEDWATCH#3007284313-2019-00095 ON JANUARY 21, 2020 ANOTHER PSEUDOANEURYSM WAS REPORTED, OCCURRING IN THE RIGHT GROIN TREATMENT ZONE. THIS WAS TREATED BY MEANS OF A GORE® VIABAHN® ENDOPROSTHESIS 8MM X 100MM. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: Pseudoaneurysm
MAUDE report 10631563 ↗InjuryApr 8, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON (B)(6) 2013 A PATIENT WAS UNDERGOING TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM WITH GORE® EXCLUDER® AAA ENDOPROSTHESES. ON (B)(6) 2017 THE PATIENT HAD A DEEP STERNAL WOUND INFECTION. ON (B)(6) 2017 THE PATIENT WAS TREATED WITH IV ANTIBIOTICS, STERNAL WOUND DEBRIDEMENT AND BILATERAL PECTORALIS MAJOR MUSCLE FLAP RECONSTRUCTION. ON (B)(6) 2018 THE PATIENT REPORTEDLY HAD A RIGHT GROIN PSEUDOANEURYSM WITH INFECTED PTFE GRAFTS. ON (B)(6) 2018 THE PATIENT UNDERWENT EXTERNAL ILIAC TO MID PTFE GRAFT BYPASS; TRIPLE LIGATION OF FEMORAL VESSELS AND CLOSURE OF THE PSEUDOANEURYSM. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: Unspecified Infection
MAUDE report 8492761 ↗InjuryFeb 26, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2015, THE PATIENT WAS IMPLANTED WITH A GORE® EXCLUDER® AAA ENDOPROSTHESES TO TREAT AN ABDOMINAL AORTIC ANEURYSM. IT WAS REPORTED THAT PATIENT HAD A PERSISTENT TYPE I ENDOLEAK. THERE WAS ANEURYSM ENLARGEMENT (AMOUNT IS UNKNOWN). THE PHYSICIAN REINTERVENED ON (B)(6) 2019, USED APTUS ANCHORS AND IMPLANTED A GORE® EXCLUDER® AAA AORTIC EXTENDER ENDOPROSTHESIS (B)(4). THE ENDOLEAK WAS NOT RESOLVED AND THE PHYSICIAN IS TAKING A WAIT AND WATCH APPROACH. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: Aneurysm
MAUDE report 8371480 ↗InjuryJan 15, 2019Patient-Device Incompatibility; Patient-Device Incompatibility
THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON (B)(6) 2014, THIS PATIENT UNDERWENT ENDOVASCULAR TREATMENT FOR UNCOMPLICATED ABDOMINAL AORTIC ANEURYSM AND WAS THEREFORE TREATED WITH GORE® EXCLUDER® AAA ENDOPROSTHESES. ON AN UNKNOWN DATE IN MARCH 2016, FOLLOW-UP IMAGING REVEALED A TYPE II ENDOLEAK. ON AN UNKNOWN DATE IN 2017, FOLLOW-UP IMAGING IDENTIFIED AN ANEURYSM ENLARGEMENT WITH A DIAMETER MEASURING 54X44 MM. REPORTEDLY, THE AMOUNT OF THE ANEURYSM INCREASE WAS UNKNOWN. ON (B)(6) 2018, THE ENDOPROTHESIS WERE EXPLANTED DUE TO THE TYPE II ENDOLEAK.
Patient problems: No Code Available
MAUDE report 8247563 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P020004. One submission often covers many models, so check the product and lot details in each record.
- Z-0995-2020 · Jan 6, 2020 · TerminatedGORE EXCLUDER AAA Endoprosthesis
- Z-0996-2020 · Jan 6, 2020 · TerminatedGORE EXCLUDER Iliac Branch Endoprosthesis (IBE)
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page