GORE EXCLUDER AAA Endoprosthesis with C3 Delivery RMT311417Recalls & reported problems
W. L. Gore & Associates, Inc. · Class 3 · Implant
All models of GORE EXCLUDER AAA Endoprosthesis with C3 Delivery (51)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00733132610761), received between Sep 27, 2018 and Jul 31, 2025 — classified as 5 injuries.
The most frequently reported specific device problem is “Incomplete or Inadequate Connection” (1 report).
2 other recalls cover devices cleared under the same FDA submission (P020004); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Incomplete or Inadequate Connection1
- Patient-Device Incompatibility1
Patient problems
- Aneurysm3
- No Clinical Signs, Symptoms or Conditions2
- Aneurysm expansion1
- No Code Available1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryJul 31, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE FOLLOWING WAS REPORTED TO GORE: ON (B)(6) 2011, PATIENT UNDERWENT AN UNKNOWN ENDOVASCULAR PROCEDURE UTILIZING GORE® EXCLUDER® AAA ENDOPROSTHESIS. ON (B)(6) 2025, PATIENT UNDERWENT AN ENDOVASCULAR PROCEDURE TO TREAT A DISTAL ANEURYSM IN THE RIGHT LIMB UTILIZING A GORE® EXCLUDER® AAA ENDOPROSTHESIS. REPORTEDLY, PHYSICIAN EMBOLIZED THE RIGHT HYPOGASTRIC ARTERY AND EXTENDED INTO THE RIGHT EXTERNAL ILIAC ARTERY WITH A CONTRALATERAL LEG. THERE WAS NO ENDOLEAK SEEN ON THE CT SCANS PRIOR TO THE REINTERVENTION AND CAUSE OF THE ANEURYSM SIZE AND GROWTH IS UNKNOWN. PATIENT IS DOING OKAY AFTER THE PROCEDURE.
Patient problems: Aneurysm expansion; No Clinical Signs, Symptoms or Conditions
MAUDE report 22675295 ↗InjuryDec 14, 2021Incomplete or Inadequate Connection; Incomplete or Inadequate Connection
THE FOLLOWING WAS REPORTED TO GORE: ON (B)(6) 2013, THE PATIENT PRESENTED WITH AN ANEURYSM IN THE AORTA AN UNDERWENT TREATMENT UTILIZING TWO GORE® EXCLUDER® AAA ENDOPROSTHESES (AAA) (RMT311417/(B)(4) & PLC161400/(B)(4)). THE PATIENT TOLERATED THE PROCEDURE. ON (B)(6) 2021, THE PATIENT PRESENTED WITH A SUSPECTED TYPE 3 ENDOLEAK OF THE PREVIOUSLY IMPLANTED AAA DEVICES AND UNDERWENT REINTERVENTION UTILIZING A GORE® EXCLUDER® CONFORMABLE AAA ENDOPROSTHESES (CXT). UPON DEPLOYMENT OF THE CXT DEVICE, THE DISTAL END OF THE DEVICE (MOST INFERIOR STENT ROW) WAS HUNG UP ON THE INNER WALL OF THE EXISTING 31MM AAA DEVICE, KEEPING IT FROM LAYING PARALLEL AND FULLY APPOSING THE WALL. AFTER BALLOONING, IT APPEARED THAT THE CTX DEVICE'S DISTAL STENT WAS BETTER APPOSED TO THE INNER WALL OF THE PREVIOUSLY PLACED AAA DEVICE. THE PATIENT TOLERATED THE PROCEDURE. REPORTEDLY. THERE IS SOME DOUBT AS TO HOW STABLE THE DEVICE CATHETER WAS KEPT DURING DEPLOYMENT BY THE FELLOW.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13005951 ↗InjuryApr 30, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2021, THIS PATIENT UNDERWENT FOLLOW-UP COMPUTED TOMOGRAPHY ANGIOGRAPHY WHICH DETERMINED A PROXIMAL TYPE I ENDOLEAK REPORTED TO BE CAUSED BY A LACK OF SEAL AND APPOSITION OF THE GRAFT. THE PATIENT IS GOING TO BE REFERRED TO ANOTHER PHYSICIAN FOR A POSSIBLE OPEN SURGICAL PROCEDURE, AS THE PATIENT HAS A SHORT PROXIMAL NECK AND THE PHYSICIAN DOES NOT THINK EXTENDING PROXIMALLY WILL RESOLVE THE ENDOLEAK.
Patient problems: Aneurysm
MAUDE report 11758223 ↗InjuryJun 25, 2020Patient-Device Incompatibility; Patient-Device Incompatibility
IT WAS REPORTED THAT ON (B)(6) 2011, THE PATIENT PRESENTED WITH A 55 MM ABDOMINAL AORTIC ANEURYSM THAT WAS TREATED WITH GORE® EXCLUDER® AAA ENDOPROSTHESES. FOLLOW-UP IMAGING SHOWED THAT THE ANEURYSM SIZE APPEARED TO BE STABLE UNTIL (B)(6) 2017 (52 MM). ANEURYSM GROWTH TO 60 MM WAS VISIBLE ON (B)(6) 2018. CONTINUOUS ANEURYSM GROWTH TO 65 MM WAS IDENTIFIED ON (B)(6) 2019. ON (B)(6) 2019, A TYPE II ENDOLEAK OF UNSPECIFIED ORIGIN WAS IDENTIFIED AND SUCCESSFULLY TREATED WITH EMBOLIZATION ON (B)(6) 2019. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: Aneurysm; No Code Available
MAUDE report 10194069 ↗InjurySep 27, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2011, THE PATIENT WAS IMPLANTED WITH A GORE® EXCLUDER® AAA ENDOPROSTHESES TO TREAT AN ABDOMINAL AORTIC ANEURYSM. IT WAS REPORTED COMPUTED TOMOGRAPHY IMAGES (DATE UNKNOWN), REVEALED THERE IS A PROXIMAL TYPE I ENDOLEAK WITH ANEURYSM SAC ENLARGEMENT. THERE IS ALSO ENLARGEMENT OF THE LEFT COMMON ILIAC ARTERY ANEURYSM (TREATED AREA) AND A DISTAL TYPE I ENDOLEAK (PREVIOUSLY IMPLANTED PXC201400/9437286). POSSIBLE REINTERVENTION ON (B)(6) 2018.
Patient problems: Aneurysm
MAUDE report 7917169 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P020004. One submission often covers many models, so check the product and lot details in each record.
- Z-0995-2020 · Jan 6, 2020 · TerminatedGORE EXCLUDER AAA Endoprosthesis
- Z-0996-2020 · Jan 6, 2020 · TerminatedGORE EXCLUDER Iliac Branch Endoprosthesis (IBE)
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page