GORE TAG Conformable Thoracic Endoprosthesis TGU404010Recalls & reported problems
W. L. Gore & Associates, Inc. · TAG THORACIC ENDO CONFORM 40mmX10cm 24Fr · Class 3 · Implant
All models of GORE TAG Conformable Thoracic Endoprosthesis (20)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00733132617739), received between Jul 24, 2018 and Dec 14, 2020 — classified as 7 injuries.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (1 report).
6 other recalls cover devices cleared under the same FDA submission (P040043); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem5
- Insufficient Device Problem Information1
- Patient-Device Incompatibility1
Patient problems
- Aneurysm3
- No Known Impact Or Consequence To Patient2
- Paraplegia1
- Vascular Dissection1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryDec 14, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2015, THE PATIENT WAS TREATED FOR A THORACIC AORTIC ANEURYSM. THREE CONFORMABLE GORE® TAG® THORACIC ENDOPROSTHESES WERE IMPLANTED. NO FURTHER INFORMATION WAS AVAILABLE. ON (B)(6) 2017, THE PATIENT WAS IMPLANTED WITH A CONFORMABLE GORE® TAG® THORACIC ENDOPROSTHESIS. NO INFORMATION IS AVAILABLE FOR THE CAUSE OF THE IMPLANT. ON (B)(6) 2020, THE PATIENT WAS IMPLANTED WITH A CONFORMABLE GORE® TAG® THORACIC ENDOPROSTHESIS. IT WAS USED TO PROXIMALLY EXTEND ONE OF THE PREVIOUSLY IMPLANTED DEVICES. THERE WAS NO INFORMATION AVAILABLE ON THE REASON THE DEVICE WAS IMPLANTED. THE PHYSICIAN STATED THAT THERE WAS DISEASE PROGRESSION. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 11004187 ↗InjuryOct 19, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON (B)(6) 2020 A PATIENT UNDERWENT ENDOVASCULAR TREATMENT OF AN AORTIC ANEURYSM IN THE TAMBE STUDY. THE CASE INCLUDED THE USE OF A CONFORMABLE GORE® TAG® THORACIC ENDOPROSTHESIS. THE PATIENT EXHIBITED SPINAL CORD ISCHEMIA THE SAME DAY WHICH WAS TREATED WITH PLACEMENT OF A SPINAL DRAIN. REHABILITATION SERVICES WERE PROVIDED. ON (B)(6) 2020 PARAPLEGIA WAS CONFIRMED AND PHYSICAL THERAPY IS IN PROGRESS.
Patient problems: Paraplegia
MAUDE report 10702038 ↗InjuryAug 5, 2020Patient-Device Incompatibility; Patient-Device Incompatibility
ON (B)(6) 2019, THIS PATIENT UNDERWENT ENDOVASCULAR TREATMENT OF A THORACIC AORTIC ANEURYSM USING CONFORMABLE GORE TAG THORACIC ENDOPROSTHESES. A TGU343410 WAS DEPLOYED PROXIMAL, AND A TGU404010 WAS THEN PLACED DISTALLY. THE OVERLAP OF THE DEVICES WAS GREATER THAN 3 CM. ON (B)(6) 2019, 5-MONTH FOLLOW-UP CTA SCAN SHOWED A TYPE II ENDOLEAK WITH FILLING OF THE INTERCOSTALS ARTERIES. THE ANEURYSM INCREASED IN SIZE FROM 66 MM AT BASELINE TO 68 MM. THE PHYSICIAN OPTED TO REPEAT A CTA IN A YEAR. ON (B)(6) 2020, THE MORE RECENT CTA SHOWED CONTINUED GROWTH TO 76 MM. THERE ALSO APPEARED TO BE A TYPE III ENDOLEAK ON THE INNER CURVE OF THE TWO DEVICES. THERE APPEARED TO BE SOME ¿BIRD BEAKING¿ OF THE TGU404010 DEVICE. THIS WAS CONFIRMED WITH AN ANGIOGRAM ON (B)(6) 2020, AND THE ENTIRE STENTED SEGMENT WAS RELINED USING A SINGLE TGMR404015 DEVICE. FINAL ANGIOGRAMS SHOWED COMPLETE RESOLUTION OF THE ENDOLEAK.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10369156 ↗InjuryOct 22, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2019, THE PATIENT UNDERWENT ENDOVASCULAR REPAIR OF AN ABDOMINAL AORTIC ANEURYSM USING CONFORMABLE GORE® TAG® THORACIC (CTAG) ENDOPROSTHESES, GORE® EXCLUDER® AAA (EXCLUDER) ENDOPROSTHESES, AND GORE® VIABAHN® VBX BALLOON EXPANDABLE (VBX) ENDOPROSTHESES. IMAGES (DATE AND MODALITY) REVEAL A POSSIBLE UNKNOWN ENDOLEAK OR A ¿POSSIBLE GUTTER¿ (BETWEEN WHICH DEVICES IS UNKNOWN). THERE IS ANEURYSM SAC ENLARGEMENT, AMOUNT IS UNKNOWN. IT WAS REPORTED THAT ON (B)(6) 2019, THERE WAS A REINTERVENTION. THE PHYSICIAN EXTENDED THE SUPERIOR MESENTERIC ARTERY (SMA) WITH A GORE® VIABAHN® VBX BALLOON EXPANDABLE (VBX) ENDOPROSTHESES AND THEN PLACED A 40 MM CONFORMABLE GORE® TAG® THORACIC ENDOPROSTHESIS. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: Aneurysm
MAUDE report 9218828 ↗InjuryMar 12, 2019Insufficient Device Problem Information; Insufficient Device Problem Information
ON (B)(6) 2016, THE PATIENT WAS IMPLANTED WITH CONFORMABLE GORE® TAG® THORACIC ENDOPROSTHESES TO TREAT A THORACOABDOMINAL AORTIC ANEURYSM. THE DEVICES WERE PLACED IN THE DISTAL DESCENDING THORACIC AORTA. PATIENT TOLERATED THE PROCEDURE. REPORTEDLY, THE PRE-TREATMENT COMPUTED TOMOGRAPHY ANGIOGRAPHY (CTA) SHOWED THAT THE MAXIMUM DIAMETER OF THE AORTIC ANEURYSM/LESION WAS 65MM. IT WAS REPORTED THAT A DISSECTION ORIGINATED IMMEDIATELY ADJACENT BELOW THE DISTAL END OF THE IMPLANTED THORACIC ENDOGRAFT EXTENDING THROUGHOUT THE ABDOMINAL AORTA INTO THE COMMON ILIAC ARTERIES BILATERALLY. THE FIRST DEVICE (TGU404015/14222568) WAS DEPLOYED AT THE AREA OF SEVERE TORTUOSITY COVERING THE PROXIMAL TEAR BUT APPROXIMATELY 5 CM SUPERIOR TO THE CELIAC AXIS. THIS DEPLOYED NICELY. HOWEVER, THE PHYSICIAN WANTED TO EXTEND THIS DOWN THE CELIAC ACCESS TO EXCLUDE THE DISSECTION, THEREFORE A TGU404010/13263733 WAS USED AS A DISTAL EXTENSION. THE DEVICE WAS PLACED INTO THE AORTA AND SECOND ANGIOGRAM WAS PERFORMED DEMONSTRATING THE LOCATION OF THE CELIAC AXIS AGAIN AND THIS WAS CONFIRMED, AND THE SECOND DEVICE WAS DELIVERED 1 OR 2 CM SUPERIOR TO THE CELIAC ACCESS. FROM (B)(6) 2016 TO (B)(6) 2017, THE ANEURYSM
Patient problems: Vascular Dissection
MAUDE report 8414705 ↗InjuryMar 12, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2016, THE PATIENT WAS IMPLANTED WITH CONFORMABLE GORE® TAG® THORACIC ENDOPROSTHESES TO TREAT A THORACOABDOMINAL AORTIC ANEURYSM. THE DEVICES WERE PLACED IN THE DISTAL DESCENDING THORACIC AORTA. PATIENT TOLERATED THE PROCEDURE. REPORTEDLY, THE PRE-TREATMENT COMPUTED TOMOGRAPHY ANGIOGRAPHY (CTA) SHOWED THAT THE MAXIMUM DIAMETER OF THE AORTIC ANEURYSM/LESION WAS 65MM. FROM (B)(6) 2016 TO (B)(6) 2017, THE ANEURYSM DIAMETER DECREASED IN SIZE FROM 59MM TO 43MM. HOWEVER, ON (B)(6) 2018, THE FOLLOW UP CTA SHOWED THAT THE MAXIMUM DIAMETER OF THE AORTIC ANEURYSM/LESION WAS 48.6MM. ON (B)(6) 2019, THE FOLLOW UP CTA SHOWED THAT THE MAXIMUM DIAMETER OF THE AORTIC ANEURYSM/LESION WAS 43MM. ADDITIONALLY, THE AORTIC AND LEFT ILIAC DISSECTION WAS IDENTIFIED. REPORTEDLY, THE ENDOVASCULAR TREATMENT AND AN OPEN PROCEDURE WERE NOT POSSIBLE DUE TO THE SIGNIFICANT RISK OF MORBIDITY AND MORTALITY. THE PHYSICIAN IS MONITORING THE PATIENT.
Patient problems: Aneurysm
MAUDE report 8414695 ↗InjuryJul 24, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON (B)(6) 2016, THE PATIENT UNDERWENT TREATMENT OF A THORACOABDOMINAL AORTIC ANEURYSM WHEREBY CONFORMABLE GORE® TAG® THORACIC ENDOPROSTHESES AND GORE® VIABAHN® ENDOPROSTHESES WITH HEPARIN BIOACTIVE SURFACE WERE IMPLANTED. ON (B)(6) 2018 THE PATIENT EXHIBITED A GUTTER LEAK AROUND THE LEFT RENAL ARTERY. ON (B)(6) 2018 THE PATIENT'S LEFT RENAL ARTERY WAS STENTED AND A LEFT RENAL ARTERY BRANCH IN THE GUTTER SPACE WAS EMBOLIZED. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: Aneurysm
MAUDE report 7715012 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P040043. One submission often covers many models, so check the product and lot details in each record.
- Z-0291-2021 · Sep 9, 2020 · Open, ClassifiedGORE TAG Conformable Thoracic Stent Graft with ACTIVE CONTROL System
- Z-0290-2021 · Sep 9, 2020 · Open, ClassifiedGORE TAG CONFORMABLE Thoracic Stent Graft with ACTIVE CONTROL System
- Z-2052-2020 · Dec 17, 2019 · TerminatedConformable Thoracic Stent Graft, Rx only, Sterile EO - Product Usage: This device is intended for endovascular repair of lesions of the descending thoracic aorta.
- Z-0048-2018 · Sep 25, 2017 · TerminatedGORE TAG, THORACIC ENDOPROSTHESIS, CONFORMABLE. The GORE TAG Thoracic Endoprosthesis is intended for endovascular repair of all lesions of the descending thoracic aorta.
- Z-1236-06 · Apr 28, 2006 · TerminatedGORE TAG THORACIC ENDOPROSTHESIS System, Endovascular Graft, Aortic Aneurysm Treatment, Catalogue Number TG3410, 34mm x 10cm
- Z-1237-06 · Apr 28, 2006 · TerminatedGORE TAG THORACIC ENDOPROSTHESIS System, Endovascular Graft, Aortic Aneurysm Treatment, Catalogue Number TG3415, 34mm x 15cm
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page