GORE TAG Conformable Thoracic Endoprosthesis TGU212110Recalls & reported problems
W. L. Gore & Associates, Inc. · TAG THORACIC ENDO CONFORM 21mmX10cm 18Fr · Class 3 · Implant
All models of GORE TAG Conformable Thoracic Endoprosthesis (20)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00733132617777), received between Jul 23, 2018 and Nov 17, 2025 — classified as 4 injuries, 1 malfunction.
The most frequently reported specific device problem is “Improper or Incorrect Procedure or Method” (1 report).
6 other recalls cover devices cleared under the same FDA submission (P040043); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Improper or Incorrect Procedure or Method1
- Insufficient Device Problem Information1
- Material Separation1
- Off-Label Use1
Patient problems
- Vascular Dissection2
- No Clinical Signs, Symptoms or Conditions1
- No Code Available1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryNov 17, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE FOLLOWING WAS REPORTED TO GORE: ON (B)(6) 2011, PATIENT UNDERWENT AN THORACIC ENDOVASCULAR AORTIC REPAIR (TEVAR) PROCEDURE UTILIZING A GORE® TAG® CONFORMABLE THORACIC ENDOPROSTHESIS. ON (B)(6) 2025, PATIENT CALLED STATING THE DEVICE WAS DEFECTIVE ACCORDING TO HIS CURRENT PHYSICIANS AND THAT IT NEEDS TO BE EXPLANTED. NO FURTHER INFORMATION WAS PROVIDED BY THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23568411 ↗MalfunctionMar 5, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2020, THE PATIENT WAS IMPLANTED WITH A CONFORMABLE GORE® TAG® THORACIC ENDOPROSTHESIS (CTAG) TO TREAT AN AORTIC TRANSECTION. IT WAS REPORTED THAT THE DEPLOYMENT OF THE CTAG WAS COMPLETED SUCCESSFULLY AND THERE WAS NO HARM DONE TO THE PATIENT. UPON REMOVAL OF THE CATHETER, THE OLIVE TIP WAS CAUGHT IN THE VALVE OF THE COOK 18F INTRODUCER SHEATH. THE SURGEON USED FORCE TO ATTEMPT TO PULL THE CATHETER FROM THE INTRODUCER SHEATH AND THE OLIVE TIP BECAME DETACHED FROM THE CATHETER AND REMAINED IN THE SHEATH. THE PROCEDURE WAS COMPLETED WITH THE CTAG CATHETER STILL ON THE WIRE AND THE TIP STILL STUCK IN THE SHEATH. AFTER A COMPLETION ANGIOGRAM WAS PERFORMED, THE PHYSICIAN REMOVED THE COOK 18F SHEATH WITH THE CTAG CATHETHER AND AMPLATZ SUPER STIFF GUIDEWIRE ALL AS ONE UNIT. THE PHYSICIAN THEN SUCCESSFULLY CLOSED THE ACCESS SITE. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: No Code Available
MAUDE report 9793203 ↗InjuryOct 4, 2019Material Separation; Improper or Incorrect Procedure or Method; Material Separation; Improper or Incorrect Procedure or Method
ON (B)(6) 2019 THE PATIENT WAS BEING TREATED FOR A PENETRATING AORTIC ULCER (PAU) USING A CONFORMABLE GORE® TAG® THORACIC ENDOPROSTHESIS. IT WAS REPORTED THAT THE PHYSICIAN USED A NON-GORE (COOK) SHEATH AND DURING REMOVAL OF THE DELIVERY SYSTEM, THE DISTAL OLIVE TIP CAUGHT THE SHEATH AND DISLODGED. THE PHYSICIAN REPORTED RESISTANCE IN TRYING TO GET PAST THE VALVE IN THE COOK SHEATH. WHEN THE PHYSICIAN PULLED AGAINST THE RESISTANCE, THE OLIVE TIP CAME OFF IN THE SHEATH. THE PHYSICIAN REPORTEDLY PUSHED THE WIRE FORWARD AND THAT DISLODGED THE OLIVE FROM THE SHEATH AND DOWN INTO THE PATIENT¿S GROIN. REPORTEDLY A CUT DOWN INSTEAD OF PERCUTANEOUS CLOSURE WAS PERFORMED TO ENSURE THE OLIVE DID NOT GET CAUGHT ON PRE-CLOSE SUTURES. THE PATIENT REPORTEDLY TOLERATED THE PROCEDURE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9159387 ↗InjuryFeb 27, 2019Off-Label Use; Off-Label Use
ON (B)(6) 2019, THIS PATIENT UNDERWENT ENDOVASCULAR TREATMENT FOR INFRARENAL AORTIC DISSECTION AND WAS IMPLANTED WITH A CONFORMABLE GORE TAG® THORACIC ENDOPROSTHESES (CTAG). IT WAS REPORTED THAT THERE WAS PARTIAL UNINTENTIONAL COVERAGE OF THE RIGHT RENAL ARTERY. THERE WAS NO DEFICIENCY OF THE DEVICE REPORTED. THE PATIENT TOLERATED THE PROCEDURE. ON (B)(6) 2019, THE PATIENT WAS REPORTED TO BE EXPERIENCING RENAL FUNCTION. COMPUTED TOMOGRAPHY WAS PERFORMED THIS DATE AND SHOWED COMPLETE COVERAGE OF THE RIGHT RENAL ARTERY. THE PATIENT UNDERWENT REINTERVENTION AND THE WAS CONVERTED TO OPEN REPAIR AND THE CONFORMABLE GORE TAG® THORACIC ENDOPROSTHESES WAS EXPLANTED. THE PATIENT TOLERATED THE REINTERVENTION.
Patient problems: Vascular Dissection
MAUDE report 8376343 ↗InjuryJul 23, 2018Insufficient Device Problem Information; Insufficient Device Problem Information
ON (B)(6) 2018, THE PATIENT UNDERWENT ENDOVASCULAR TREATMENT OF A COARCTATION OF THE THORACIC AORTA AND A DESCENDING THORACIC AORTIC DISSECTION WITH GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS (TGU212110/16460122). IT WAS REPORTED THE PATIENT HAD A FEVER PRIOR TO THE IMPLANT PROCEDURE. THE PATIENT TOLERATED THE PROCEDURE. ON (B)(6) 2018, THE PATIENT UNDERWENT TREATMENT FOR A PSEUDOANEURYSM DISTAL TO THE IMPLANTED GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS. THERE WAS A POSSIBLE INFECTION STILL PRESENT. THE PHYSICIAN IMPLANTED AN ADDITIONAL GORE® TAG® THORACIC BRANCH ENDOPROSTHESIS (THU212110/16576646A). THE PATIENT TOLERATED THE PROCEDURE. THE PHYSICIAN IS TAKING A WAIT AND WATCH APPROACH FOR THE POSSIBLE INFECTION.
Patient problems: Vascular Dissection
MAUDE report 7713157 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P040043. One submission often covers many models, so check the product and lot details in each record.
- Z-0290-2021 · Sep 9, 2020 · Open, ClassifiedGORE TAG CONFORMABLE Thoracic Stent Graft with ACTIVE CONTROL System
- Z-0291-2021 · Sep 9, 2020 · Open, ClassifiedGORE TAG Conformable Thoracic Stent Graft with ACTIVE CONTROL System
- Z-2052-2020 · Dec 17, 2019 · TerminatedConformable Thoracic Stent Graft, Rx only, Sterile EO - Product Usage: This device is intended for endovascular repair of lesions of the descending thoracic aorta.
- Z-0048-2018 · Sep 25, 2017 · TerminatedGORE TAG, THORACIC ENDOPROSTHESIS, CONFORMABLE. The GORE TAG Thoracic Endoprosthesis is intended for endovascular repair of all lesions of the descending thoracic aorta.
- Z-1236-06 · Apr 28, 2006 · TerminatedGORE TAG THORACIC ENDOPROSTHESIS System, Endovascular Graft, Aortic Aneurysm Treatment, Catalogue Number TG3410, 34mm x 10cm
- Z-1237-06 · Apr 28, 2006 · TerminatedGORE TAG THORACIC ENDOPROSTHESIS System, Endovascular Graft, Aortic Aneurysm Treatment, Catalogue Number TG3415, 34mm x 15cm
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page