GORE TAG Conformable Thoracic Endoprosthesis TGU404015Recalls & reported problems
W. L. Gore & Associates, Inc. · TAG THORACIC ENDO CONFORM 40mmX15cm 24Fr · Class 3 · Implant
All models of GORE TAG Conformable Thoracic Endoprosthesis (20)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 00733132618163), received between Sep 13, 2018 and Nov 15, 2022 — classified as 6 injuries, 1 malfunction.
The most frequently reported specific device problem is “Insufficient Device Problem Information” (2 reports).
6 other recalls cover devices cleared under the same FDA submission (P040043); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Insufficient Device Problem Information2
- Activation, Positioning or Separation Problem1
- Compatibility Problem1
- Unintended Movement1
- Use of Device Problem1
Patient problems
- Aneurysm5
- Vascular Dissection2
- Aortic Dissection1
- Clinical Endoleak1
- No Code Available1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
InjuryNov 15, 2022Activation, Positioning or Separation Problem; Insufficient Device Problem Information; Activation, Positioning or Separation Problem; Insufficient Device Problem Information
ON (B)(6) 2018, THE PATIENT UNDERWENT ENDOVASCULAR TREATMENT OF A THORACIC AORTIC DISSECTION WITH CONFORMABLE GORE® TAG® THORACIC ENDOPROSTHESES (TGU404015/15928245)-MOST ADJACENT TO THE TBE AORTIC COMPONENT. ON (B)(6) 2018, THE PRE-TREATMENT COMPUTED TOMOGRAPHY (CT) IMAGING SHOWED THAT THE MAXIMUM DIAMETER OF AN AORTIC ANEURYSM/ LESION WAS 54MM. THE FOLLOW UP CT SCANS SHOWED THAT THE MAXIMUM DIAMETER OF AN AORTIC ANEURYSM/ LESION INCREASED IN SIZE FROM 55 MM ON NOVEMBER 4, 2019 TILL 63MM ON (B)(6) 2022. ALSO, A TYPE III ENDOLEAK WAS DETERMINED. THE EVENT IS RELATED TO THE DEVICE AND THE ENDOVASCULAR PROCEDURE. THE EVENT IS ONGOING. NO TREATMENT WAS REPORTED. THE PHYSICIAN IS MONITORING THE PATIENT.
Patient problems: Aneurysm; Clinical Endoleak; Vascular Dissection
MAUDE report 15802719 ↗InjuryJul 30, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2018, THE PATIENT WAS IMPLANTED WITH CONFORMABLE GORE® TAG® THORACIC ENDOPROSTHESES TO TREAT A TYPE B AORTIC DISSECTION. IT WAS REPORTED THAT COMPUTED TOMOGRAPHY IMAGES DATED (B)(6) 2019, REVEALED A DISTAL TYPE I ENDOLEAK WITH DEVICE MALPOSITION ON THE DISTAL END. IT IS UNKNOWN IF THERE IS ANEURYSM SAC ENLARGEMENT. ON (B)(6) 2019, THE PHYSICIAN REINTERVENED AND IMPLANTED A CONFORMABLE GORE® TAG® THORACIC ENDOPROSTHESES AND EXTENDED DISTALLY. THE ENDOLEAK WAS RESOLVED AND THE PATIENT TOLERATED THE REINTERVENTION.
Patient problems: Aortic Dissection
MAUDE report 8844304 ↗InjuryApr 24, 2019Use of Device Problem; Insufficient Device Problem Information; Use of Device Problem; Insufficient Device Problem Information
ON (B)(6) 2018, THE PATIENT UNDERWENT ENDOVASCULAR TREATMENT OF A THORACIC AORTIC DISSECTION WITH CONFORMABLE GORE® TAG® THORACIC ENDOPROSTHESES (TGU404015/15928245)-MOST ADJACENT TO THE TBE AORTIC COMPONENT. ON (B)(6) 2018, THE PRE-TREATMENT COMPUTED TOMOGRAPHY (CT) IMAGING SHOWED THAT THE MAXIMUM DIAMETER OF AN AORTIC ANEURYSM/ LESION WAS 54MM. ON (B)(6) 2018, DURING THE FOLLOW UP COMPUTED TOMOGRAPHY SCAN, A TYPE III ENDOLEAK ASSOCIATED WITH INSUFFICIENT OVERLAP PER IFU BETWEEN AN AORTIC EXTENDER (TGU404015/15928245) AND AN INVESTIGATIONAL AORTIC COMPONENT WAS DISCOVERED. THE MAXIMUM DIAMETER OF AN AORTIC ANEURYSM/ LESION WAS 51MM. THE PHYSICIAN WAS MONITORING THE PATIENT. ON (B)(6) 2019, THE 12 MONTHS FOLLOW UP CT CONFIRMED THE PRESENCE OF THE TYPE III ENDOLEAK BETWEEN AN AORTIC EXTENDER (TGU404015/15928245) AND AN INVESTIGATIONAL AORTIC COMPONENT. THE MAXIMUM DIAMETER OF AN AORTIC ANEURYSM/ LESION INCREASED IN SIZE FROM 50MM ON (B)(6) 2018 TILL 54MM ON (B)(6) 2019. ON (B)(6) 2019, THE PATIENT UNDERWENT THE ADDITIONAL PROCEDURE AND THE THORACIC AORTIC STENT GRAFT WAS PLACED. ON (B)(6) 2019, THE PATIENT DISCHARGED FROM THE HOSPITAL. NO FURTHER ADVERSE EVENTS HAVE BEEN REPORTED.
Patient problems: Aneurysm; No Code Available
MAUDE report 8547005 ↗InjuryApr 23, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2019, THE PATIENT WAS IMPLANTED WITH CONFORMABLE GORE® TAG® THORACIC ENDOPROSTHESES TO TREAT A DESCENDING THORACIC DISSECTION. IT WAS REPORTED THAT ON FOLLOW UP IMAGES COMPUTED TOMOGRAPHY IMAGES DATED (B)(6) 2019, REVEALED THE DISSECTION HAD EXTENDED TO THE CELIAC. THE TGU404015/18666716 DEVICE WAS THE MOST DISTALLY IMPLANTED. THERE WAS A REINTERVENTION ON (B)(6) 2019, TO EXTEND THE MOST DISTAL DEVICE. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: Vascular Dissection
MAUDE report 8541897 ↗InjuryJan 25, 2019Unintended Movement; Unintended Movement
ON (B)(6) 2019, THE PATIENT UNDERWENT TREATMENT OF A THORACIC AORTIC ANEURYSM WITH A CONFORMABLE GORE® TAG® THORACIC ENDOPROSTHESIS. IT WAS REPORTED THE PHYSICIAN INTENDED TO IMPLANT THE DEVICE JUST DISTAL TO THE BOVINE ARCH. REPORTEDLY, THE DEVICE ¿JUMPED¿ FORWARD APPROXIMATELY 2CM AND PARTIALLY COVERING THE BOVINE ARCH. THE CAUSE OF THE DEVICE MOVEMENT IS REPORTEDLY UNKNOWN. IT WAS REPORTED THE DEVICE WAS LEFT IN THE PATIENT AND A BYPASS WAS PERFORMED TO PERFUSE THE BOVINE ARCH. THE PROCEDURE CONCLUDED WITHOUT ANY FURTHER COMPLICATIONS BEING REPORTED AND THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: Aneurysm
MAUDE report 8279261 ↗InjuryJan 10, 2019Compatibility Problem; Compatibility Problem
ON (B)(6) 2014, THE PATIENT WAS TREATED FOR A THORACIC AORTIC ANEURYSM USING ENDOLOGIX STENT GRAFT AND CONFORMABLE GORE® TAG® THORACIC ENDOPROSTHESES. ON AN UNKNOWN DATE, IMAGING IDENTIFIED A TYPE III ENDOLEAK COMPONENT DISCONNECT BETWEEN THE ENDOLOGIX DEVICE AND A CTAG DEVICE (LOT NUMBER UNKNOWN). IT WAS REPORTED THE CAUSE OF TYPE III ENDOLEAK IS UNKNOWN. NO ANEURYSM ENLARGEMENT WAS REPORTED. ON (B)(6) 2018, A REINTERVENTION WAS PERFORMED BRIDGING THE TWO DEVICES. THE ENDOLEAK WAS REPORTEDLY RESOLVED AND THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: Aneurysm
MAUDE report 8238055 ↗MalfunctionSep 13, 2018Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2018, A CONFORMABLE GORE® TAG® THORACIC ENDOPROSTHESIS (ITEM/LOT NOT AVAILABLE) WAS TO BE IMPLANTED AS PART OF A THORACIC ENDOVASCULAR AORTIC REPAIR. ACCORDING TO THE REPORT, THE GORE® TAG® DEVICE COULD NOT BE ADVANCED TO ITS TARGET LOCATION. THE DECISION WAS MADE TO WITHDRAW THE DELIVERY CATHETER. UPON DEVICE WITHDRAWAL, THE GORE® DRYSEAL SHEATH (ITEM/LOT NOT AVAILABLE) VALVE WAS REPORTEDLY RUPTURED. AFTER THE DEVICE WAS WITHDRAWN, IT WAS REPORTED THE PHYSICIAN NOTICED A METAL PROTRUSION FROM THE LEADING OLIVE AT THE TIP OF THE DELIVERY CATHETER. THE SHEATH WAS REMOVED AND REPLACED, AND A NEW CTAG OF THE SAME CATALOG NUMBER WAS USED TO CONTINUE THE PROCEDURE. THE PROCEDURE WENT FORWARD TO COMPLETION WITHOUT ANY FURTHER REPORTED ISSUES. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: Aneurysm
MAUDE report 7875169 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P040043. One submission often covers many models, so check the product and lot details in each record.
- Z-0290-2021 · Sep 9, 2020 · Open, ClassifiedGORE TAG CONFORMABLE Thoracic Stent Graft with ACTIVE CONTROL System
- Z-0291-2021 · Sep 9, 2020 · Open, ClassifiedGORE TAG Conformable Thoracic Stent Graft with ACTIVE CONTROL System
- Z-2052-2020 · Dec 17, 2019 · TerminatedConformable Thoracic Stent Graft, Rx only, Sterile EO - Product Usage: This device is intended for endovascular repair of lesions of the descending thoracic aorta.
- Z-0048-2018 · Sep 25, 2017 · TerminatedGORE TAG, THORACIC ENDOPROSTHESIS, CONFORMABLE. The GORE TAG Thoracic Endoprosthesis is intended for endovascular repair of all lesions of the descending thoracic aorta.
- Z-1236-06 · Apr 28, 2006 · TerminatedGORE TAG THORACIC ENDOPROSTHESIS System, Endovascular Graft, Aortic Aneurysm Treatment, Catalogue Number TG3410, 34mm x 10cm
- Z-1237-06 · Apr 28, 2006 · TerminatedGORE TAG THORACIC ENDOPROSTHESIS System, Endovascular Graft, Aortic Aneurysm Treatment, Catalogue Number TG3415, 34mm x 15cm
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page