GORE EXCLUDER AAA Endoprosthesis PXC231000Recalls & reported problems
W. L. Gore & Associates, Inc. · Contralateral Leg · Class 3 · Implant
All models of GORE EXCLUDER AAA Endoprosthesis (81)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00733132618248), received between Jul 29, 2022 and May 8, 2025 — classified as 2 injuries.
The most frequently reported specific device problem is “Fluid/Blood Leak” (1 report).
2 other recalls cover devices cleared under the same FDA submission (P020004); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak1
- Inadequacy of Device Shape and/or Size1
Patient problems
- Clinical Endoleak1
- Ruptured Aneurysm1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryMay 8, 2025Inadequacy of Device Shape and/or Size; Inadequacy of Device Shape and/or Size
THE FOLLOWING WAS REPORTED TO GORE: ON (B)(6) 2012, A PATIENT WAS TREATED FOR AN ABDOMINAL AORTIC ANEURYSM USING A GORE® EXCLUDER® AAA ENDOPROSTHESIS. THE PHYSICIAN USED A FLARED LIMB ON THE 23 MM CONTRALATERAL LEG. ON (B)(6) 2025, THE PATIENT WAS TREATED FOR A TYPE 1B ENDOLEAK ON THE CONTRALATERAL LEG. A GORE® EXCLUDER® ILIAC BRANCH ENDOPROSTHESIS (IBE) WAS PLACED INTO THE CONTRALATERAL LEG, AND A GORE® VIABAHN® VBX BALLOON EXPANDABLE ENDOPROSTHESIS WAS INSERTED FROM THE ARM AND DEPLOYED IN THE INTERNAL ILIAC ARTERY. THE PROCEDURE WAS SUCCESSFUL. THE PHYSICIAN REPORTED THAT HE BELIEVED THAT THE ILIAC ARTERY HAD WIDENED OVER TIME, LEADING TO THE ENDOLEAK.
Patient problems: Clinical Endoleak
MAUDE report 21986408 ↗InjuryJul 29, 2022Fluid/Blood Leak; Fluid/Blood Leak
THE FOLLOWING WAS REPORTED TO GORE: ON (B)(6) 2012, THE PATIENT PRESENTED WITH AN ABDOMINAL AORTIC ANEURYSM AND UNDERWENT TREATMENT UTILIZING THREE GORE® EXCLUDER® AAA ENDOPROSTHESES (AAA). THE PATIENT TOLERATED THE PROCEDURE. ON (B)(6) 2022, THE PATIENT PRESENTED EMERGENTLY WITH A RUPTURED RIGHT COMMON ILIAC ANEURYSM AND A DISTAL TYPE 1B ENDOLEAK AND UNDERWENT TREATMENT UTILIZING TWO ADDITIONAL GORE® EXCLUDER® AAA ENDOPROSTHESES AND COIL EMBOLIZATION OF THE RIGHT INTERNAL ILIAC ARTERY. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: Ruptured Aneurysm
MAUDE report 15133513 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P020004. One submission often covers many models, so check the product and lot details in each record.
- Z-0995-2020 · Jan 6, 2020 · TerminatedGORE EXCLUDER AAA Endoprosthesis
- Z-0996-2020 · Jan 6, 2020 · TerminatedGORE EXCLUDER Iliac Branch Endoprosthesis (IBE)
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.