Skip to content

GORE EXCLUDER AAA Endoprosthesis with C3 Delivery RLT281214Recalls & reported problems

W. L. Gore & Associates, Inc. · EXCLUDER AAA ENDO TRNK IPSI 28.5mmX12mmX14cm 18Fr · Class 3 · Implant

All models of GORE EXCLUDER AAA Endoprosthesis with C3 Delivery (51)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00733132622498), received between Aug 26, 2020 and Mar 19, 2024 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Patient Device Interaction Problem” (1 report).

2 other recalls cover devices cleared under the same FDA submission (P020004); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem2
  • Patient Device Interaction Problem1
  • Patient-Device Incompatibility1

Patient problems

  • Aneurysm2
  • Clinical Endoleak1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 19, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ON (B)(6) 2024, THE PATIENT UNDERWENT REINTERVENTION AND EMBOLIZATION OF THE ANEURYSM SAC WITH SHAPEMEDICAL PLUGS WAS PERFORMED. THE PHYSICIAN REPORTS SOME SAC ENLARGEMENT (AMOUNT UNKNOWN AND DATE OF ONSET OF TYPE II ENDOLEAK DETERMINED IS UNKNOWN) THE PATIENT TOLERATED THE PROCEDURE, ENDOLEAK RESOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18932175 ↗
  • InjurySep 18, 2023Patient Device Interaction Problem; Patient Device Interaction Problem

    ON (B)(6) 2014, THE PATIENT WAS TREATED FOR AN ABDOMINAL AORTIC ANEURYSM. THE PHYSICIAN IMPLANTED TWO GORE® EXCLUDER® AAA ENDOPROSTHESES. THE PATIENT TOLERATED THE PROCEDURE. ON (B)(6) 2023, IMAGING REVEALED A TYPE IA AND TYPE IB ENDOLEAK, AS WELL AS A TYPE II ENDOLEAK. THE FSA BELIEVES THAT THE TYPE II ENDOLEAK CAUSED NECK DILATATION, RESULTING IN THE TWO TYPE I ENDOLEAKS. REINTERVENTION IS PLANNED. IMAGES ARE AVAILABLE.

    Patient problems: Clinical Endoleak

    MAUDE report 17769649 ↗
  • InjurySep 8, 2022Patient-Device Incompatibility; Patient-Device Incompatibility

    ON (B)(6) 2018, THE PATIENT UNDERWENT ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM USING GORE® EXCLUDER® AAA ENDOPROSTHESES. ON (B)(6) 2022, THE GORE REPRESENTATIVE BECAME AWARE OF A PROXIMAL TYPE I ENDOLEAK WITHOUT DEVICE MIGRATION. NECK DEGRADATION WAS REPORTED AS THE CAUSE OF THE ENDOLEAK. IT IS REPORTEDLY UNKNOWN IF ANEURYSM ENLARGEMENT WAS PRESENT. ON (B)(6) 2022, A REINTERVENTION WAS PERFORMED WHEREBY A GORE® EXCLUDER® AAA AORTIC EXTENDER COMPONENT WAS PLACED PROXIMALLY. THE ENDOLEAK WAS SUCCESSFULLY RESOLVED. THERE WERE NO FURTHER REPORTED ISSUES. THE PATIENT TOLERATED THE PROCEDURE.

    Patient problems: Aneurysm

    MAUDE report 15385490 ↗
  • MalfunctionAug 26, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT DURING A PROCEDURE GORE® EXCLUDER® AAA ENDOPROSTHESIS (RLT281214/21448982) WAS ADVANCED OVER THE .035 MEIER¿ GUIDEWIRE THROUGH THE 18FR DRYSEAL SHEATH. AFTER A FULL DEPLOYMENT WHEN THE EXCLUDER® DEVICE WAS TAKEN OUT, THE OLIVE TIP REMAINED ON THE WIRE AT THE TIP OF THE SHEATH. THIS WAS AN AREA OF ACUTE ANGLE IN THE ILIAC ARTERY. A FOGARTY® BALLOON WAS ADVANCED ABOVE THE OLIVE TIP AND INFLATED. IT WAS THEN PULLED BACK AND THE OLIVE TIP WAS CAPTURED IN THE DRY SEAL SHEATH. ALL OF THE COMPONENTS WERE RETRIEVED. THE PATIENT TOLERATED THE PROCEDURE.

    Patient problems: Aneurysm

    MAUDE report 10457830 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under P020004. One submission often covers many models, so check the product and lot details in each record.

  • Z-0996-2020 · Jan 6, 2020 · Terminated
    GORE EXCLUDER Iliac Branch Endoprosthesis (IBE)
  • Z-0995-2020 · Jan 6, 2020 · Terminated
    GORE EXCLUDER AAA Endoprosthesis

Other System, Endovascular Graft, Aortic Aneurysm Treatment devices

All System, Endovascular Graft, Aortic Aneurysm Treatment devices →

Sources and updates