GORE EXCLUDER AAA Endoprosthesis with C3 Delivery RLT281214Recalls & reported problems
W. L. Gore & Associates, Inc. · EXCLUDER AAA ENDO TRNK IPSI 28.5mmX12mmX14cm 18Fr · Class 3 · Implant
All models of GORE EXCLUDER AAA Endoprosthesis with C3 Delivery (51)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00733132622498), received between Aug 26, 2020 and Mar 19, 2024 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Patient Device Interaction Problem” (1 report).
2 other recalls cover devices cleared under the same FDA submission (P020004); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Patient Device Interaction Problem1
- Patient-Device Incompatibility1
Patient problems
- Aneurysm2
- Clinical Endoleak1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryMar 19, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
ON (B)(6) 2024, THE PATIENT UNDERWENT REINTERVENTION AND EMBOLIZATION OF THE ANEURYSM SAC WITH SHAPEMEDICAL PLUGS WAS PERFORMED. THE PHYSICIAN REPORTS SOME SAC ENLARGEMENT (AMOUNT UNKNOWN AND DATE OF ONSET OF TYPE II ENDOLEAK DETERMINED IS UNKNOWN) THE PATIENT TOLERATED THE PROCEDURE, ENDOLEAK RESOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18932175 ↗InjurySep 18, 2023Patient Device Interaction Problem; Patient Device Interaction Problem
ON (B)(6) 2014, THE PATIENT WAS TREATED FOR AN ABDOMINAL AORTIC ANEURYSM. THE PHYSICIAN IMPLANTED TWO GORE® EXCLUDER® AAA ENDOPROSTHESES. THE PATIENT TOLERATED THE PROCEDURE. ON (B)(6) 2023, IMAGING REVEALED A TYPE IA AND TYPE IB ENDOLEAK, AS WELL AS A TYPE II ENDOLEAK. THE FSA BELIEVES THAT THE TYPE II ENDOLEAK CAUSED NECK DILATATION, RESULTING IN THE TWO TYPE I ENDOLEAKS. REINTERVENTION IS PLANNED. IMAGES ARE AVAILABLE.
Patient problems: Clinical Endoleak
MAUDE report 17769649 ↗InjurySep 8, 2022Patient-Device Incompatibility; Patient-Device Incompatibility
ON (B)(6) 2018, THE PATIENT UNDERWENT ENDOVASCULAR TREATMENT OF AN ABDOMINAL AORTIC ANEURYSM USING GORE® EXCLUDER® AAA ENDOPROSTHESES. ON (B)(6) 2022, THE GORE REPRESENTATIVE BECAME AWARE OF A PROXIMAL TYPE I ENDOLEAK WITHOUT DEVICE MIGRATION. NECK DEGRADATION WAS REPORTED AS THE CAUSE OF THE ENDOLEAK. IT IS REPORTEDLY UNKNOWN IF ANEURYSM ENLARGEMENT WAS PRESENT. ON (B)(6) 2022, A REINTERVENTION WAS PERFORMED WHEREBY A GORE® EXCLUDER® AAA AORTIC EXTENDER COMPONENT WAS PLACED PROXIMALLY. THE ENDOLEAK WAS SUCCESSFULLY RESOLVED. THERE WERE NO FURTHER REPORTED ISSUES. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: Aneurysm
MAUDE report 15385490 ↗MalfunctionAug 26, 2020Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT DURING A PROCEDURE GORE® EXCLUDER® AAA ENDOPROSTHESIS (RLT281214/21448982) WAS ADVANCED OVER THE .035 MEIER¿ GUIDEWIRE THROUGH THE 18FR DRYSEAL SHEATH. AFTER A FULL DEPLOYMENT WHEN THE EXCLUDER® DEVICE WAS TAKEN OUT, THE OLIVE TIP REMAINED ON THE WIRE AT THE TIP OF THE SHEATH. THIS WAS AN AREA OF ACUTE ANGLE IN THE ILIAC ARTERY. A FOGARTY® BALLOON WAS ADVANCED ABOVE THE OLIVE TIP AND INFLATED. IT WAS THEN PULLED BACK AND THE OLIVE TIP WAS CAPTURED IN THE DRY SEAL SHEATH. ALL OF THE COMPONENTS WERE RETRIEVED. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: Aneurysm
MAUDE report 10457830 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under P020004. One submission often covers many models, so check the product and lot details in each record.
- Z-0996-2020 · Jan 6, 2020 · TerminatedGORE EXCLUDER Iliac Branch Endoprosthesis (IBE)
- Z-0995-2020 · Jan 6, 2020 · TerminatedGORE EXCLUDER AAA Endoprosthesis
Other System, Endovascular Graft, Aortic Aneurysm Treatment devices
- Endurant® IIs ESBF2814C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls34 deaths254 reports
- Endurant® IIs ESBF3214C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls19 deaths173 reports
- Endurant® IIs ESBF3614C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT1 recall37 deaths214 reports
- Endurant® IIs ESBF2514C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls18 deaths136 reports
- Endurant® IIs ESBF2314C103EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls9 deaths60 reports
- RelayPro Thoracic Stent-Graft System 28-N4-34-154-34UBolton Medical Inc. · System, Endovascular Graft, Aortic Aneurysm Treatment3 recalls7 reports
- Vela Proximal Endograft System A34-34/C100-O20 VEndologix LLC · System, Endovascular Graft, Aortic Aneurysm Treatment3 deaths152 reports
- Endurant® II ETCF2828C49EMedtronic, Inc. · SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT2 recalls7 deaths44 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page