GORE VIABAHN Endoprosthesis with Heparin VBHR060502ARecalls & reported problems

W. L. Gore & Associates, Inc. · VIABAHN SX ENDO HEPARIN 35 RO 6mmX5cm 7Fr 120cmCath · Class 3 · Implant

All models of GORE VIABAHN Endoprosthesis with Heparin (62)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00733132622764), received between Mar 4, 2022 and Oct 25, 2024 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Break” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break2
  • Detachment of Device or Device Component2
  • Adverse Event Without Identified Device or Use Problem1
  • Entrapment of Device1
  • Loss of or Failure to Bond1
  • Material Separation1
  • Physical Resistance/Sticking1
  • Separation Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Device Embedded In Tissue or Plaque1
  • Foreign Body In Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 25, 2024Break; Break

    THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON (B)(6) 2024, A GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE (VIABAHN DEVICE) WAS INTENDED OR FOR USE TO TREAT A POPLITEAL ANEURYSM. REPORTEDLY, THE VIABAHN DEVICE WAS ADVANCED OVER A 0.035" ROSENWIRE THROUGH AN UNKNOWN SIZE AND BRAND INTRODUCER SHEATH. IT WAS NOTED WHEN THE DEPLOYMENT LINE WAS PULLED IT BROKE AND THE VIABAHN DEVICE FAILED TO DEPLOY. THE VIABAHN DEVICE WAS REMOVED WITHOUT DIFFICULTY. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER VIABAHN DEVICE. THERE WERE NO CONSEQUENCES OR IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20537964 ↗
  • MalfunctionApr 15, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    ON (B)(6) 2024, A PATIENT UNDERWENT A PHYSICIAN-MODIFIED ENDOVASCULAR GRAFT (PMEG) PROCEDURE FOR AN ANEURYSM WHERE A 5 MM X 6 CM GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE (VIABAHN® DEVICE) WAS INTENDED TO BE USED IN THE LEFT RENAL ARTERY. IT WAS REPORTED THE PHYSICIAN ACCESSED THE PATIENT VIA THE GROIN USING A 7F MEDTRONIC TOURGUIDE SHEATH OVER AN UNKNOWN GUIDEWIRE AND THROUGH A FENESTRATION ON A COOK® ZENITH FLEX® AAA ENDOVASCULAR GRAFT. REPORTEDLY, THERE WERE NO MODIFICATIONS MADE TO THE VIABAHN DEVICE OR DELIVERY SYSTEM AS PART OF THE PROCEDURE. THE DELIVERY CATHETER WAS ADVANCED BUT WOULD NOT COME OUT OF THE SHEATH INTO THE ARTERY DUE TO THE ANGLE OF THE ANATOMY AND SHEATH. THE PHYSICIAN ADJUSTED THE BEND IN THE SHEATH, BUT THE DELIVERY CATHETER STILL WOULD NOT ADVANCE. REPORTEDLY, THE PHYSICIAN DECIDED TO WITHDRAW THE DELIVERY CATHETER IN ORDER TO PREDILATE. THERE WAS A LOT OF FRICTION DURING WITHDRAWAL DUE TO THE COMPLEXITY OF THE CASE. WHEN THE DELIVERY CATHETER WAS REMOVED, A RADIOPAQUE (RO) OBJECT WAS OBSERVED ON IMAGING IN THE LEFT KIDNEY. INITIALLY THE PHYSICIAN SUSPECTED ONE OF THE THREE RO MARKER BANDS FROM A PHILLIPS QUICK-CROSS SUPPORT CATHETER (AL

    Patient problems: Foreign Body In Patient

    MAUDE report 19105673 ↗
  • InjuryJun 6, 2022Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED TO GORE THAT DURING A PROCEDURE TO TREAT THE SUPERFICIAL FEMORAL ARTERY THE GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE GOT STUCK AND THE PROCEDURE NEEDED TO BE CONVERTED TO A CUT DOWN PROCEDURE TO REMOVE THE DEVICE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14607194 ↗
  • MalfunctionMar 4, 2022Loss of or Failure to Bond; Break; Entrapment of Device; Material Separation; Detachment of Device or Device Component; Physical Resistance/Sticking; Separation Problem; Loss of or Failure to Bond; Material Separation; Detachment of Device or Device Component; Separation Problem

    THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON (B)(6) 2022, THE PATIENT WAS BEING TREATED FOR A STENOTIC LESION IN A ILIAC ARTERY USING A GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE. THE VESSEL WAS PRE-DILATED PRIOR TO THE ADVANCEMENT OF THE DEVICE. THE DEVICE WAS ADVANCE AND DEPLOYED IN ITS INTENDED LOCATION WITH NO REPORTED ISSUE. RESISTANCE WAS NOTED AS THE DELIVERY CATHETER WAS BEING REMOVED FROM THE PATIENT. AS THE DELIVERY CATHETER WAS REMOVED FROM THE GUIDEWIRE IT WAS NOTED THAT THE DISTAL SHAFT WAS NOT ATTACHED. THE DISTAL SHAFT WAS SEPARATED FROM THE DELIVERY CATHETER AND WAS LOCATED IN THE PATIENTS COMMON ILIAC ARTERY. THE PHYSICIANS SUSPECTED CAUSE OF THE DISTAL SHAFT SEPARATION WAS EXCESSIVE CALCIFICATION IN THE VESSEL. ANOTHER GORE DEVICE WAS IMPLANTED TO PIN THE DISTAL SHAFT TO THE VESSEL WALL. THE PATIENT TOLERATED THE PROCEDURE.

    Patient problems: Device Embedded In Tissue or Plaque; No Clinical Signs, Symptoms or Conditions

    MAUDE report 13675883 ↗

Other Stent, Superficial Femoral Artery devices

All Stent, Superficial Femoral Artery devices →