Skip to content

GORE VIABAHN Endoprosthesis with Heparin VBHR062502ARecalls & reported problems

W. L. Gore & Associates, Inc. · VIABAHN SX ENDO HEPARIN 35 RO 6mmX25cm 7Fr 120cmCath · Class 3 · Implant

All models of GORE VIABAHN Endoprosthesis with Heparin (62)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00733132622863), received between Dec 14, 2023 and Jun 26, 2025 — classified as 1 injury, 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Material Fragmentation1

Patient problems

  • Foreign Body Embolism1
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • InjuryJun 26, 2025Material Fragmentation; Material Fragmentation

    THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON (B)(6) 2025, A GORE® VIABAHN® ENDOPROSTHESIS (VSX DEVICE) WAS INTENDED FOR USE IN THE SUPERFICIAL FEMORAL ARTERY DURING TREATMENT OF STENOSIS. DURING THE ADVANCEMENT OF A 6MM X 25CM VSX DEVICE INTO A 6FR X 55CM COOK RAABE INTRODUCER SHEATH OVER AN 0.018" V18 GUIDE WIRE, THE DEVICE HAD GOT STUCK IN THE VALVE OF THE INTRODUCER SHEATH. THE VSX DEVICE AND INTRODUCER SHEATH WERE REMOVED. A 7FR X 55CM COOK RAABE INTRODUCER SHEATH WAS ADVANCED AND A NEW 6MM X 25CM VSX DEVICE WAS SUCCESSFULLY IMPLANTED ACROSS THE LESION. HOWEVER, AFTER THE DELIVERY CATHETER WAS REMOVED FROM THE PATIENT, THE DISTAL TIP WAS OBSERVED TO BE MISSING. THE PHYSICIAN PERFORMED AN ANGIOGRAM AND SAW THAT THE DISTAL TIP HAD MIGRATED DOWN TO THE POPLITEAL ARTERY. THE PHYSICIAN ATTEMPTED TO SNARE THE DISTAL TIP WITH NO SUCCESS. THE PATIENT WAS MOVED TO THE OPERATING ROOM, AND THE PHYSICIAN PERFORMED A CUT DOWN AND REMOVED THE DISTAL TIP. IT WAS NOTED THAT THE PHYSICIAN FELT RESISTANCE ADVANCING THE VSX DEVICE THROUGH THE STENOSIS.

    Patient problems: Foreign Body Embolism

    MAUDE report 22340794 ↗
  • MalfunctionDec 14, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    THE FOLLOWING WAS REPORTED TO GORE: DURING TREATMENT OF A STENOTIC OCCLUSION IN THE PATIENT'S LEFT SUPERFICIAL FEMORAL ARTERY, ACCESS WAS MADE FROM THE CONTRALATERAL SIDE. A 7FR X 45CM TERUMO SHEATH WAS USED TO ADVANCE THE FIRST GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE (VIABAHN DEVICE) OVER A .035 GUIDEWIRE (BRAND NOT SPECIFIED). AS REPORTED, THE PHYSICIAN DID NOT PRE-DILATE THE STENOTIC OCCLUSION AND WAS PUSHING THE DEVICE WITH SOME FORCE. HOWEVER, THE DEVICE DID NOT CROSS THE LESION. THE VIABAHN DEVICE WAS PULLED BACK THROUGH THE SHEATH. AFTER DILATING THE TARGET LESION, THE SAME VIABAHN DEVICE WAS ADVANCED AGAIN AND WAS DEPLOYED. WHEN PULLING THE CATHETER BACK THROUGH THE SHEATH, THE DISTAL TIP GOT CAUGHT AT THE EDGE OF THE SHEATH AND THE CATHETER BROKE. AS REPORTED, THE BROKEN PIECE WAS PULLED OUT THROUGH THE SHEATH AS IT WAS NEAR THE OPENING OF THE SHEATH. IT WAS REPORTED THE DELIVERY CATHETER SEPARATION WAS AT THE SHEATH OPENING AND IT WAS CLOSE ENOUGH TO PULL OUT MANUALLY. A SECOND VIABAHN DEVICE WAS IMPLANTED, OVERLAPPING THE FIRST DEVICE IN THE SAME ANATOMICAL LOCATION. THE PATIENT IS REPORTED TO BE DOING WELL FOLLOWING THE PROCEDURE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18333046 ↗

Other System, Endovascular Graft, Arteriovenous (Av) Dialysis Access Circuit Stenosis Treatment devices

All System, Endovascular Graft, Arteriovenous (Av) Dialysis Access Circuit Stenosis Treatment devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page