Skip to content

GORE VIABAHN Endoprosthesis with Heparin VBHR070501ARecalls & reported problems

W. L. Gore & Associates, Inc. · VIABAHN SX ENDO HEPARIN 35 RO 7mmX5cm 8Fr 75cmCath · Class 3 · Implant

All models of GORE VIABAHN Endoprosthesis with Heparin (62)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00733132622924), received between Nov 13, 2024 and Jul 3, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Activation Failure” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Activation Failure2
  • Break1
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 3, 2025Break; Activation Failure; Break; Activation Failure

    ON (B)(6) 2025, A PATIENT UNDERWENT TREATMENT FOR A STENOSED FISTULA WHERE A GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE (VIABAHN DEVICE) WAS ATTEMPTED TO BE USED IN THE FEMORAL ARTERY. IT WAS REPORTED DURING THE PROCEDURE, WHILE ADVANCING THE VIABAHN DEVICE OVER A TERUMO GLIDEWIRE ADVANTAGE .035" GUIDEWIRE THROUGH AN 8FR CORDIS INTRODUCER SHEATH, THE PHYSICIAN-INITIATED DEPLOYMENT BY PULLING THE DEPLOYMENT LINE. AFTER PULLING APPROXIMATELY 18 INCHES, THE DEPLOYMENT LINE BROKE. IT IS UNKNOWN WHETHER EXCESSIVE SPEED OR FORCE WAS USED WHEN PULLING THE DEPLOYMENT LINE. REPORTEDLY, THE TARGET TREATMENT AREA WAS REACHED PRIOR TO INITIATING DEPLOYMENT. THERE WAS NO CALCIFICATION NOTED AND THERE WAS NO RESISTANCE DURING ADVANCEMENT. IT WAS REPORTED THERE WAS NO DISTAL EXPANSION OR OPENING OF THE STENT INITIATED. THE PHYSICIAN REMOVED THE DEVICE AND SUCCESSFULLY COMPLETED THE PROCEDURE USING A NEW 7MM X 10CM VIABAHN DEVICE. THE PHYSICIAN REMAINS UNSURE ABOUT THE CAUSE OF THE ISSUE. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE CONSEQUENCES.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22395229 ↗
  • MalfunctionNov 13, 2024Activation Failure; Physical Resistance/Sticking; Activation Failure; Physical Resistance/Sticking

    ON (B)(6) 2024, A PATIENT UNDERWENT TREATMENT FOR AN AVD CLOT WHERE A 7 MM X 5 CM GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE (GORE® VIABAHN® DEVICE) WAS USED IN AN UPPER ARM SYNTHETIC LOOP GRAFT (BRAND UNKNOWN, AND DATE OF PLACEMENT UNKNOWN). IT WAS REPORTED THE PHYSICIAN ACCESSED THE PATIENT USING AN 8F TERUMO INTRODUCER SHEATH OVER AN UNKNOWN BRAND 0.035" GUIDEWIRE TO THE TARGET TREATMENT ZONE. IT IS UNKNOWN IF THE LESION WAS PRE-DILATED. AT THE TARGET TREATMENT SITE, THE DEPLOYMENT LINE WAS PULLED AND BECAME STUCK. REPORTEDLY, THE DEPLOYMENT LINE DID NOT BREAK. THE DECISION WAS MADE TO WITHDRAW THE DELIVERY CATHETER. AN ATTEMPT TO PULL THE DELIVERY CATHETER BACK INTO THE SHEATH WAS UNSUCCESSFUL. THE INTRODUCER SHEATH AND DELIVERY CATHETER WERE WITHDRAWN TOGETHER. THE PROCEDURE WAS COMPLETED USING ANOTHER SHEATH AND GORE® VIABAHN® DEVICE (SAME SIZE). IT WAS REPORTED THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20675673 ↗

Other Stent, Superficial Femoral Artery devices

All Stent, Superficial Femoral Artery devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page