Skip to content

GORE VIABAHN Endoprosthesis with Heparin VBHR070502ARecalls & reported problems

W. L. Gore & Associates, Inc. · VIABAHN SX ENDO HEPARIN 35 RO 7mmX5cm 8Fr 120cmCath · Class 3 · Implant

All models of GORE VIABAHN Endoprosthesis with Heparin (62)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 00733132622948), received between Nov 13, 2024 and Mar 24, 2025 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 24, 2025Break; Break

    ON (B)(6) 2025, A PATIENT WITH BILATERAL LOWER-LIMB BYPASS GRAFTS, UNDERWENT LEFT LEG BYPASS REVISION (BRAND OF GRAFTS AND DATE OF INITIAL IMPLANT UNKNOWN). IT WAS REPORTED THE PATIENT DID NOT TAKE ANTICOAGULANT MEDICATION FOR AT LEAST 12 DAYS. ON (B)(6) 2025, THE PATIENT PRESENTED WITH LOWER LIMB THREATENING ISCHEMIA OF UNKNOWN DURATION. IT WAS NOTED THERE WAS THROMBOSIS IN THE LEFT FEMORAL-TIBIAL BYPASS ARTERY BYPASS GRAFT, AND SEVERE STENOSIS IN THE LEFT EXTERNAL ILIAC ARTERY. THE DECISION WAS MADE TO USE A 7 MM X 5 CM GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE (VIABAHN DEVICE) IN THE SEVERELY STENOSED LEFT EXTERNAL ILIAC ARTERY. IT WAS REPORTED, THE LESION IN THE EXTERNAL ILIAC ARTERY WAS PRE-DILATED WITH A 6 MM X 60 MM ARMADA BALLOON. THE PHYSICIAN ACCESSED THE PATIENT USING A 10 F UNKNOWN BRAND INTRODUCER SHEATH OVER AN .035" TERUMO ADVANTAGE GUIDE WIRE (180 CM) TO THE TARGET TREATMENT SITE. REPORTEDLY, THE ENDOPROSTHESIS DEPLOYED. AFTER WITHDRAWAL OF THE DELIVERY SYSTEM CATHETER, IT WAS NOTICED THE DISTAL SECTION OF THE CATHETER WAS RETAINED INSIDE THE PATIENT. THE CASE WAS EXTENDED BY APPROXIMATELY 30 - 40 MINUTES AND ADDITIONAL HEPARIN WAS UTILIZED. THE B

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21678264 ↗
  • MalfunctionNov 13, 2024

    SCRUB TECH FLUSHED STENT AND SAW THAT IT WAS LEAKING AROUND SEAL THAT IS USED TO DEPLOY STENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20673652 ↗

Other System, Endovascular Graft, Arteriovenous (Av) Dialysis Access Circuit Stenosis Treatment devices

All System, Endovascular Graft, Arteriovenous (Av) Dialysis Access Circuit Stenosis Treatment devices →