GORE VIABAHN Endoprosthesis with Heparin VBH100502ARecalls & reported problems
W. L. Gore & Associates, Inc. · VIABAHN SX ENDO HEPARIN 35 10mmX5cm 11Fr120cmCath · Class 3 · Implant
All models of GORE VIABAHN Endoprosthesis with Heparin (62)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 4
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00733132623211), received between Sep 6, 2018 and Feb 16, 2021 — classified as 4 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Material Separation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Material Separation2
- Detachment of Device or Device Component1
- Difficult or Delayed Separation1
- Patient Device Interaction Problem1
- Patient-Device Incompatibility1
- Physical Resistance/Sticking1
- Use of Device Problem1
Patient problems
- Perforation of Vessels2
- Aneurysm1
- Device Embedded In Tissue or Plaque1
- Insufficient Information1
- No Known Impact Or Consequence To Patient1
- Occlusion1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryFeb 16, 2021Physical Resistance/Sticking; Difficult or Delayed Separation
DURING TREATMENT OF A SPLENIC ANEURYSM, A GORE® VIABAHN® ENDOPROSTHESIS (VIABAHN DEVICE) DID NOT ADVANCE TO THE SPLENIC ARTERY DUE TO TORTUOUS ANATOMY. ATTEMPTS WERE MADE TO WITHDRAW THE VIABAHN DEVICE THROUGH THE 12FR COOK SHEATH; HOWEVER, THE DEVICE COULD NOT BE WITHDRAWN. THE DEVICE WAS DEPLOYED IN THE GASTRODUODENAL ARTERY AND THE DEPLOYMENT LINE WAS PULLED OUT. HOWEVER, THE DELIVERY CATHETER BECAME STUCK AND COULD NOT BE REMOVED. DECISION WAS MADE TO HAVE THE TRANSPLANT SURGEON CUT AND REMOVE THE DELIVERY CATHETER DURING A SCHEDULED RENAL TRANSPLANT SURGERY ON (B)(6) 2020. (B)(4).
Patient problems: Aneurysm; Insufficient Information
MAUDE report 11335205 ↗InjuryFeb 28, 2020Patient-Device Incompatibility; Patient-Device Incompatibility
ON (B)(6) 2020, THE PATIENT UNDERWENT TREATMENT FOR A POPLITEAL ARTERY OCCLUSION USING GORE® VIABAHN® ENDOPROSTHESES. IT WAS REPORTED A GORE® VIABAHN® ENDOPROSTHESIS (VBH100502A/20466865) WAS ADVANCED INTO ITS INTENDED LOCATION. REPORTEDLY THE DEPLOYMENT LINE WAS PULLED, HOWEVER, THE DEVICE FAILED TO COMPLETELY EXPAND LEAVING A PORTION OF THE DEVICE CONSTRAINED. THE PHYSICIAN BALLOONED THE GORE® VIABAHN® ENDOPROSTHESIS AND THE DEVICE WAS FULLY EXPANDED. HOWEVER, AS THE DEPLOYMENT CATHETER WAS BEING REMOVED FROM THE PATIENT, IT WAS OBSERVED THAT THE POPLITEAL ARTERY HAD RUPTURED. THE PROCEDURE WAS COMPLETED WITH THE PHYSICIAN PERFORMING A FEM-POP BYPASS USING A VASCULAR GRAFT. THE PATIENT TOLERATED THE PROCEDURE. THE PHYSICIAN STATED IT IS UNKNOWN WHAT CAUSED THE PARTIAL DEVICE EXPANSION OR THE RUPTURE.
Patient problems: Perforation of Vessels
MAUDE report 9769531 ↗MalfunctionDec 20, 2019Material Separation; Material Separation
ON (B)(6) 2019, THE PATIENT WAS IMPLANTED WITH GORE® VIABAHN® ENDOPROSTHESES TO TREAT A POPLITEAL ANEURYSM. IT WAS REPORTED A .035 GUIDEWIRE WAS USED DURING THE PROCEDURE. THE SHEATH BRAND AND SIZE ARE REPORTEDLY UNKNOWN. IT WAS REPORTED DURING DEPLOYMENT, THE DEPLOYMENT LINE BROKE AND THE DEVICE WAS PARTIALLY OPENED APPROXIMATELY 98 PERCENT. THE PHYSICIAN REPORTEDLY STATED IS UNKNOWN WHAT CAUSED THE DEPLOYMENT LINE TO BREAK. IT WAS REPORTED THE CATHETER WAS CUT IN ORDER TO GAIN ACCESS TO THE REMAINING DEPLOYMENT LINE. THE DEPLOYMENT LINE WAS PULLED AND THE DEVICE WAS REPORTEDLY COMPLETELY OPENED. HOWEVER, A PORTION OF THE DEPLOYMENT LINE REMAINED IN THE PATIENT¿S BODY AND COULD NOT BE RETRIEVED. THE PROCEDURE WAS COMPLETED. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 9500958 ↗MalfunctionJul 23, 2019Detachment of Device or Device Component; Detachment of Device or Device Component
THE FOLLOWING WAS REPORTED TO GORE: THE AV/DIALYSIS PATIENT PRESENTED FOR CENTRAL VENOUS AND CEPHALIC ARCH STENOSIS TREATMENT. THE GORE® VIABAHN® ENDOPROSTHESIS (VIABAHN) WAS ADVANCED TO THE TREATMENT SITE IN THE CEPHALIC ARCH AND DEPLOYED WITHOUT ISSUE. HOWEVER, THE DEPLOYMENT LINE REMAINED ATTACHED TO THE VIABAHN. THE DOCTOR, ON A SEPARATE WIRE, ADVANCED A 10X4 MUSTANG¿ BALLOON INTO THE VIABAHN AND INFLATED IT TO FULL SIZE. THIS FIXATED THE VIABAHN ALLOWING THE DOCTOR TO PULL THE DEPLOYMENT LINE WHILE HIS ASSISTANT PUSHED ON THE BALLOON. THE COMBINATION OF SIMULTANEOUSLY PUSHING THE BALLOON AND PULLING THE DEPLOYMENT LINE FREED THE DEPLOYMENT LINE. THE DEPLOYMENT LINE WAS REMOVED FROM THE PATIENT. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: Occlusion
MAUDE report 8817710 ↗InjuryDec 13, 2018Patient Device Interaction Problem; Patient Device Interaction Problem
THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON (B)(6) 2018 A PATIENT WAS UNDERGOING A PROCEDURE TO RELINE A PREVIOUSLY IMPLANTED STENT IN AN AV ACCESS APPLICATION IN THE SUBCLAVIAN VEIN WITH A GORE® VIABAHN® ENDOPROSTHESIS. THE DEVICE WAS ATTEMPTED TO BE ADVANCED WITHIN THE PREVIOUSLY IMPLANTED STENT, BUT IT WAS NOT POSSIBLE TO DO SO. THE UNDEPLOYED DEVICE WAS THEN WITHDRAWN AND AS IT WAS BEING WITHDRAWN IT BECAME STUCK ON THE EDGE OF THE SHEATH, CAUSING IT TO ACCORDION ON THE DELIVERY CATHETER. THE SHEATH AND DEVICE WERE THEN REMOVED. AS THEY WERE REMOVED, SOME VESSEL DAMAGE OCCURRED WHICH REQUIRED A CUT-DOWN AND SOME ADDITIONAL SUTURING. A NEW 10FR SHEATH WAS THEN INSERTED AND THE CASE WAS COMPLETED WITH A NON-GORE STENT. THE PATIENT DID WELL.
Patient problems: Perforation of Vessels
MAUDE report 8162024 ↗InjurySep 6, 2018Material Separation; Use of Device Problem; Material Separation; Use of Device Problem
THE FOLLOWING INFORMATION WAS REPORTED TO GORE: ON (B)(6) 2018 A PATIENT WAS UNDERGOING A PROCEDURE FOR ARTERIO-VENOUS ACCESS IN THE UPPER ARM WITH A GORE® VIABAHN® ENDOPROSTHESIS. THE ACCESS VESSEL WAS PRE-DILATED WITH AN 8FR SHEATH, THEN THE SHEATH WAS REMOVED. AN ATTEMPT WAS MADE TO ADVANCE THE DEVICE WITHOUT A SHEATH IN PLACE, BUT IT COULD NOT BE ADVANCED. THE DELIVERY SYSTEM WITH THE UNDEPLOYED STENT WAS REMOVED FROM THE PATIENT. UPON REMOVAL, IT WAS NOTED THE LEADING OLIVE TIP WAS MISSING FROM THE DELIVERY CATHETER. A CONVERSION TO OPEN REPAIR ENSUED. THE OLIVE TIP WAS NOT POSITIVELY IDENTIFIED, BUT THE SURGEON BELIEVES THE TIP WAS REMOVED IN THE PROCESS OF SUCTIONING AND CLEANING UP THE OPERATIVE SITE. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7852965 ↗
Other System, Endovascular Graft, Arteriovenous (Av) Dialysis Access Circuit Stenosis Treatment devices
- COVERA AVSM10060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall33 reports
- COVERA AVSM08060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall21 reports
- COVERA AVSM09060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall18 reports
- COVERA AVSM10080Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall16 reports
- COVERA AVSM10040Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall15 reports
- COVERA AVFM10060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall14 reports
- COVERA AVSM09080Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall1 death14 reports
- COVERA AVSM10100Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall13 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026