GORE VIABAHN Endoprosthesis with Heparin VBH130502ARecalls & reported problems
W. L. Gore & Associates, Inc. · VIABAHN SX ENDO HEPARIN 35 13mmX5cm 12Fr120cmCath · Class 3 · Implant
All models of GORE VIABAHN Endoprosthesis with Heparin (62)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00733132623280), received between Jul 1, 2019 and Feb 11, 2021 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Activation Failure” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Activation Failure2
- Failure to Unfold or Unwrap1
- Material Split, Cut or Torn1
Patient problems
- Fistula1
- Insufficient Information1
- No Information1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionFeb 11, 2021Activation Failure; Activation Failure
THE FOLLOWING WAS REPORTED TO GORE: DURING TREATMENT IN THE SUPERIOR VENA CAVA, ACCESS WAS MADE FROM THE GROIN AREA. AFTER SHEATH PLACEMENT, A GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE (VSX DEVICE) WAS DELIVERED TO THE TREATMENT AREA. DEPLOYMENT WAS INITIATED AND THE VSX DEVICE EXPANDED ABOUT HALFWAY WHEN DEPLOYMENT STOPPED. AS REPORTED, THE PHYSICIAN MADE ATTEMPTS TO COMPLETE THE DEPLOYMENT BY PULLING STEADILY ON THE LINE. HOWEVER, THE DEPLOYMENT LINE BROKE. ACCESS WAS THEN MADE FROM THE AXILLARY ARTERY AND A BALLOON CATHETER WAS DELIVERED TO THE TREATMENT SITE. THE BALLOON WAS INFLATED AND THE VSX DEVICE EXPANDED COMPLETELY. THE PROCEDURE ENDED WITH GOOD RESULTS. THE PATIENT TOLERATED THE PROCEDURE.
Patient problems: Insufficient Information
MAUDE report 11314726 ↗MalfunctionJan 2, 2020Material Split, Cut or Torn; Material Split, Cut or Torn
THE FOLLOWING WAS REPORTED TO GORE: THE PATIENT PRESENTED FOR UPPER ARM FISTULA REPAIR. THE GORE® VIABAHN® ENDOPROSTHESIS WAS ADVANCED INTO POSITION AND THE DEPLOYMENT LINE PULLED. WHILE PULLING THE DEPLOYMENT LINE, THE DEPLOYMENT LINE BROKE. THE DOCTOR WAS ABLE TO GRASP THE PARTIALLY DEPLOYED DEVICE WITH A PAIR OF HEMOSTATS AND PULLED IT OUT. WHEN OUT OF THE PATIENT THE DOCTOR OBSERVED THE TIP OF THE CATHETER WAS STILL IN THE PATIENT ON THE WIRE. THE TIP WAS REMOVED WITH THE WIRE.
Patient problems: Fistula
MAUDE report 9541186 ↗MalfunctionJul 1, 2019Failure to Unfold or Unwrap; Activation Failure; Failure to Unfold or Unwrap; Activation Failure
THE FOLLOWING WAS REPORTED TO GORE BY THE DOCTOR: THE PATIENT PRESENTED FOR OPEN ARCH REPAIR. THE DOCTOR PLACED THE GORE® VIABAHN® ENDOPROSTHESIS IN THE INNOMINATE ARTERY. WHEN THE DEPLOYMENT LINE WAS PULLED, THE DOCTOR OBSERVED IT TO APPEAR TO UNRAVEL. THE GORE® VIABAHN® ENDOPROSTHESIS DID NOT COMPLETELY OPEN. THE DOCTOR CONTINUED THE REPAIR USING OPEN TECHNIQUE.
Patient problems: No Information
MAUDE report 8750353 ↗
Other System, Endovascular Graft, Arteriovenous (Av) Dialysis Access Circuit Stenosis Treatment devices
- COVERA AVSM10060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall33 reports
- COVERA AVSM08060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall21 reports
- COVERA AVSM09060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall18 reports
- COVERA AVSM10080Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall16 reports
- COVERA AVSM10040Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall15 reports
- COVERA AVFM10060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall14 reports
- COVERA AVSM09080Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall1 death14 reports
- COVERA AVSM10100Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall13 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page