Skip to content

GORE VIABAHN Endoprosthesis with Heparin VBHR080502ARecalls & reported problems

W. L. Gore & Associates, Inc. · VIABAHN SX ENDO HEPARIN 35 RO 8mmX5cm 8Fr 120cmCath · Class 3 · Implant

All models of GORE VIABAHN Endoprosthesis with Heparin (62)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00733132623303), received between Oct 22, 2019 and Aug 14, 2026 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Material Separation” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 1 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Separation2
  • Activation Failure1
  • Complete Blockage1
  • Detachment of Device or Device Component1
  • Loss of or Failure to Bond1
  • Physical Resistance/Sticking1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient1
  • Obstruction/Occlusion1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 14, 2026Activation Failure; Physical Resistance/Sticking

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 50255607 ↗
  • InjuryAug 14, 2025Complete Blockage; Complete Blockage

    THE FOLLOWING WAS REPORTED TO GORE ON NOVEMBER 27, 2024, FROM THE IMEDIDATA STUDY DATABASE FOR THE TGR23-02AA TOGETHER AORTIC REGISTRY. ON (B)(6) 2024, THE PATIENT UNDERWENT ENDOVASCULAR TREATMENT AS AN ELECTIVE PROCEDURE FOR A RIGHT ILIAC ARTERY ANEURYSM USING A GORE® EXCLUDER® ILIAC BRANCH ENDOPROSTHESIS DEVICE SYSTEM, A GORE® EXCLUDER® CONFORMABLE AAA AORTIC EXTENDER COMPONENT, AND A GORE® VIABAHN® ENDOPROSTHESIS. THE SITES WERE ACCESSED PERCUTANEOUSLY BILATERALLY. THE DEVICES WERE SUCCESSFULLY NAVIGATED TO THE INTENDED LOCATION AND DEPLOYED WITH NO REPORTED ISSUES. DEVICE CATHETERS WERE SUCCESSFULLY REMOVED. THERE WERE NO REPORTED ISSUES. THE PATIENT WAS DISCHARGED TO HOME ON (B)(6) 2024. ON (B)(6) 2024, IT WAS REPORTED THAT THE RIGHT EXTERNAL ILIAC ARTERY HAD THROMBOSED LEADING TO OCCLUSION AND ISCHEMIA. THE GORE® EXCLUDER® ILIAC BRANCH ENDOPROSTHESIS ILIAC BRANCH COMPONENT (CEB) AND GORE® EXCLUDER® CONFORMABLE AAA AORTIC EXTENDER COMPONENT (PLA) WERE IMPLANTED IN THE PATIENT'S RIGHT EXTERNAL ILIAC ARTERY. AFTER ADDITIONAL REVIEW OF THE FOLLOW-UP IMAGING DATED (B)(6) 2024, COMPRESSION OF THE CEB DEVICE WAS OBSERVED. LOSS OF DEVICE PATENCY WAS CONFIRMED IN THE CEB AND PLA AND W

    Patient problems: Obstruction/Occlusion; Thrombosis/Thrombus

    MAUDE report 22808669 ↗
  • MalfunctionApr 8, 2024Loss of or Failure to Bond; Material Separation; Detachment of Device or Device Component; Loss of or Failure to Bond; Material Separation; Detachment of Device or Device Component

    ON (B)(6) 2024, A PATIENT UNDERWENT TREATMENT FOR AN UNKNOWN ETIOLOGY USING 8 MM X 5 CM GORE® VIABAHN® ENDOPROSTHESIS WITH HEPARIN BIOACTIVE SURFACE (VIABAHN® DEVICE) IN AN UNKNOWN ANATOMICAL LOCATION. IT WAS REPORTED THE TIP OF THE DELIVERY SYSTEM CATHETER CAME OFF AS IT WAS BEING WITHDRAWN FROM THE PATIENT. THE BROKEN PIECE WAS ACCOUNTED FOR. IT WAS REPORTED THERE WAS NO IMPACT TO THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19063303 ↗
  • MalfunctionOct 22, 2019Material Separation; Material Separation

    ON (B)(6) 2019, THE PATIENT WAS IMPLANTED WITH GORE® VIABAHN® ENDOPROSTHESES. IT WAS REPORTED THE PHYSICIAN HAD DUAL ACCESS WITH TWO 12FR DRYSEAL SHEATHS ON THE PATIENTS LEFT SIDE. IT WAS REPORTED THE GORE® VIABAHN® ENDOPROSTHESES WERE IMPLANTED IN THE PATIENT¿S LEFT COMMON ILIAC ARTERY USING A KISSING TECHNIQUE. ONE (11X5) VIABAHN DEVICE WAS IMPLANTED RETROGRADE FROM THE GROIN. THIS DEVICE EXTENDED INTO THE EXTERNAL ILIAC. A SECOND (8X10) VIABAHN DEVICE WAS IMPLANTED ANTEGRADE FROM THE AXILLARY. THIS DEVICE REPORTEDLY EXTENDED INTO THE INTERNAL ILIAC. THE PHYSICIAN THEN PLACED A (9X15) VIABAHN DEVICE THROUGH THE (8X10) VIABAHN DEVICE EXTENDING FURTHER INTO THE INTERNAL ARTERY. IT WAS REPORTED DURING EXTENSION OF THE (8X10) VIABAHN DEVICE INTO THE COMMON ILIAC ARTERY A (8X5) VIABAHN DEVICE WAS ADVANCED FROM THE AXILLARY SHEATH WITHOUT ISSUE INTO THE (8X10) VIABAHN DEVICE IN THE COMMON ILIAC. THE PHYSICIAN DEPLOYED THE STENT FULLY. IT WAS REPORTED THE DEPLOYMENT LINE BROKE AT THE HUB OF THE CATHETER. THE CATHETER WAS THEN REMOVED WITHOUT INCIDENT. THERE WAS REPORTEDLY NO ADVERSE EVENTS TO THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9218950 ↗

Other System, Endovascular Graft, Arteriovenous (Av) Dialysis Access Circuit Stenosis Treatment devices

All System, Endovascular Graft, Arteriovenous (Av) Dialysis Access Circuit Stenosis Treatment devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page