GORE VIABAHN Endoprosthesis VBCR081002ARecalls & reported problems
W. L. Gore & Associates, Inc. · VIABAHN SX ENDO 35 RO 8mmX10cm 8Fr 120cmCath · Class 3 · Implant
All models of GORE VIABAHN Endoprosthesis (26)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00733132623532), received between Sep 11, 2024 and Apr 8, 2026 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Activation Failure” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Activation Failure3
- Difficult or Delayed Positioning1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 8, 2026Difficult or Delayed Positioning; Activation Failure
THE FOLLOWING WAS REPORTED TO GORE: ON (B)(6) 2025, A PATIENT PRESENTED FOR TREATMENT IN THE LEFT AXILLARY VEIN DUE TO CHRONIC REOCCURRING STENOSIS. DURING THE PROCEDURE, THE GORE® VIABAHN® ENDOPROSTHESIS (VIABAHN® DEVICE) WAS ADVANCED OVER AN 0.035 ANGIODYNAMICS BENTSON GUIDEWIRE. AS REPORTED, THE VIABAHN® DEVICE WAS ADVANCED WITHOUT THE AID OF A SHEATH. DEPLOYMENT WAS STARTED WHEN THE VIABAHN® STENT STARTED MOVING DISTALLY FROM THE TREATMENT AREA. THE VIABAHN® STENT WAS REPOSITIONED, BUT THE VIABAHN® LINE WAS STUCK AND DEPLOYMENT COULD NOT BE CONTINUED. THE CONSTRAINED VIABAHN® DEVICE WAS REMOVED FROM THE PATIENT WITH NO ISSUES AND NO HARM TO PATIENT. AS REPORTED, THE PHYSICIAN SAID THE ISSUE LIKELY OCCURRED BECAUSE HE REPOSITIONED WHILE DEPLOYING THE STENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24824008 ↗MalfunctionSep 19, 2024Activation Failure; Activation Failure
THE FOLLOWING INFORMATION WAS REPORTED BY THE FSA: ON (B)(6) 2024, A GORE® VIABAHN® ENDOPROSTHESIS (VIABAHN DEVICE) WAS INTENDED FOR USE IN AN TO LINE A FISTULA FOR DIALYSIS AV ACCESS. THE PHYSICIAN REPORTED THERE WAS FAILURE TO DEPLOY THE DEVICE BECAUSE THE DEPLOYMENT LINE WAS STUCK. THE PROCEDURE WAS COMPLETED WITH ANOTHER VIABAHN DEVICE. THERE WERE NO CONSEQUENCES OR IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20260323 ↗MalfunctionSep 11, 2024Activation Failure; Activation Failure
ON (B)(6) 2024, A PATIENT UNDERWENT TREATMENT FOR AN UNKNOWN REASON WHERE AN 8 MM X 10 CM GORE® VIABAHN® ENDOPROSTHESIS WAS INTENDED TO BE USED IN AN UNKNOWN LOCATION. IT WAS REPORTED THE PHYSICIAN ADVANCED THE DELIVERY CATHETER USING AN 8F INTRODUCER SHEATH (BRAND UNKNOWN) OVER AN .035" ADVANTAGE GLIDEWIRE TO THE TARGET TREATMENT ZONE. REPORTEDLY, DURING DEPLOYMENT THE LINE WAS PULLED AND BECAME CAUGHT, CAUSING THE DEPLOYMENT LINE TO BECOME STRETCHED OUT NEAR THE TURN AROUND. REPORTEDLY, DEPLOYMENT WAS ABORTED, AND THE PHYSICIAN WITHDREW THE CATHETER, WITH THE ENDOPROSTHESIS FULLY CONSTRAINED. THERE WAS NO DISTAL EXPANSION INITIATED OF THE ENDOPROSTHESIS AND THE DEPLOYMENT LINE DID NOT BREAK. IT WAS REPORTED THERE WAS NO IMPACT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20191361 ↗
Other System, Endovascular Graft, Arteriovenous (Av) Dialysis Access Circuit Stenosis Treatment devices
- COVERA AVSM10060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall33 reports
- COVERA AVSM08060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall21 reports
- COVERA AVSM09060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall18 reports
- COVERA AVSM10080Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall16 reports
- COVERA AVSM10040Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall15 reports
- COVERA AVFM10060Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall14 reports
- COVERA AVSM09080Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall1 death14 reports
- COVERA AVSM10100Angiomed GmbH & Co. Medizintechnik KG · System, endovascular graft, arteriovenous (AV) dialysis access circuit stenosis treatment1 recall13 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page